Safety Precautions The following are general safety precautions that are not necessarily related to any specific part or procedure, and do not necessarily appear elsewhere in this publication. These precautions must be thoroughly understood and apply to all phases of operation and maintenance. WARNING Keep Away From Live Circuits Operating Personnel must at all times observe general safety precautions.
Safety Precautions Safety Symbols WARNING Warning notes call attention to a procedure, which if not correctly performed, could result in personal injury. CAUTION Caution notes call attention to a procedure, which if not correctly performed, could result in damage to the instrument. This symbol indicates that a shock hazard exists if the precautions in the instruction manual are not followed.
Do not contaminate the coolant with pipe sealant. On page 4. CAUTION If installed, connect optional interlock before applying RF power. On page 5. CAUTION Use only Bird coolant, P/N 5-1070, to prevent damage to the load. On page 10.
Safety Precautions Safety Statements USAGE ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO. BENUTZUNG WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN.
Page 6
Termaline® Semiconductor Load Resistor SERVICE SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE - TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO SERVICIO LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO.
Page 7
Safety Precautions CONNECT INTERLOCK TO TRANSMITTER BEFORE OPERATING. BRANCHER LE VERROUILLAGE À L'ÉMETTEUR AVANT EMPLOI. CONECTE EL INTERBLOQUEO AL TRANSMISOR ANTES DE LA OPERACION. VOR INBETRIEBNAHME VERRIEGELUNG AM SENDER ANSCHLIESSEN. PRIMA DI METTERE IN FUNZIONE L'APPARECCHIO, COLLEGARE IL DISPOSITIVO DI BLOCCO AL TRASMETTITORE.
When inquiring about updates to this manual refer to the part number and revision on the title page. Chapter Layout Introduction — Describes the features of the Bird Termaline, Semiconductor, RF Load Resistor, lists equipment supplied and optional equipment, and provides power-up instructions. Theory of Operation —...
Chapter 1 Introduction Bird 8860 Series Semiconductor Loads are portable 50 ohm coaxial RF transmission line terminations. They are designed for frequency ranges of DC – 28 MHz, specially calibrated for greater stability at 13.56 MHz. They provide accurate, dependable, and low reflection line terminations. Up to 1000 watts of RF power can be dissipated.
Theory of Operation Load Resistor Bird 8860 series loads consist of a thin-film-on-ceramic resistor immersed in a dielectric coolant. The resistor, individually selected for its accuracy, is enclosed in a special housing. When surrounded by the coolant, this produces a uniform, practically reflection-less line termination over the specified frequencies.
If the shipping container is not damaged, unpack the unit. Save shipping materials for repackaging. 2. Inspect unit for visual signs of damage. Note: If there is damage, immediately notify the shipping carrier and Bird Technologies. Setting Up Note: Before first using the load, get a resistance baseline for future maintenance. Refer to "RF...
Termaline® Semiconductor Load Resistor Installing the Thermoswitch Bird 8860 series loads can be equipped with an optional interlock thermoswitch, P/N 8890-008. It is normally closed, opening at 236 °C (457 °F), with a rating of 10A @ 120VAC and 5A @ 230VAC.
Installation Installing the Interlock Connection CAUTION If installed, connect optional interlock before applying RF power. 1. Unscrew the large knurled ring-nut (A) at the lower end of the coupling jack assembly. 2. Pull the large knurled ring-nut off the thermoswitch jack (B). 3.
Termaline® Semiconductor Load Resistor Connecting the RF Power After installing the load, the RF transmission line can be attached using standard coaxial line coupling kits. WARNING Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being applied. Leaking RF energy is a potential health hazard.
Chapter 4 Operating Instructions Normal Operation Bird 8860 series loads have no indicators or operating controls. They require no special operating procedures or surveillance when their performance limits are not exceeded. Follow the instructions for the specific generator equipment. Operation Under Abnormal Conditions The load can be subjected to moderate overloads for short periods.
Chapter 5 Maintenance WARNING Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being applied. Leaking RF energy is a potential health hazard. WARNING Disconnect the unit from all power sources before servicing. The unit may be energized from multiple sources.
Maintenance RF Assembly Resistance Test Note: These tests are by no means a necessity to the operation of the load but merely guidelines for the users information. Accurate measurement of the DC resistance between the inner and outer conductors of the RF input connector will provide a good check of the condition of the load resistor.
4. Install the load resistor. Repair Replace RF Connector The following procedure may be used to remove and replace Bird QC connectors. Note: If not using the LC connector normally supplied, the frequency and power must be limited to the capabilities of the connector.
Any maintenance or service procedure beyond the scope of those in this chapter should be referred to a qualified service center. If the unit needs to be returned for any reason, request an Return Material Authorization (RMA) through the Bird Technologies website. All instruments returned must be shipped prepaid and to the attention of the RMA number.
Termaline® Semiconductor Load Resistor Specifications Frequency Range DC – 28 MHz Power Rating 1000 W continuous duty Impedance, Nominal 50 ohms VSWR 1.1 max VSWR Stability < 0.1 dB, from 0 to 100% rated power at stability frequency Stability Frequency 13.56 MHz ±...
Limited Warranty All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year, unless otherwise specified, from date of shipment and to conform to applicable specifica- tions, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted;...