Program
Problem
The CB-2000 not detected by computer
Message "No CB-2000 present"
is displayed.
Error messages appear while trying to upload data to the CB-2000.
"Illegal date and time stamp detected"
"Illegal capacitance value detected"
"Illegal temperature value detected"
The first capacitor is missing in the
CB-2000 program.
Warning message appears while trying to upload data to the CB-2000.
"Upload is limited to 1000 rows,
upload is cancelled"
Cause
Drivers for CB-2000 are not
installed.
USB cable is not
connected.
CB-2000 is not
switched on.
Date and time format should
be:
YYYY-MM-DD hh:mm:ss
Only digits and a decimal
separator are accepted.
Only digits are accepted.
The first line is expected to be
a header line and is ignored
while opening a data file.
A maximum of 1000 capacitor
units can be stored in the
CB-2000.
ABB Power Capacitors | Portable Capacitance Meter CB-2000 — User´s Guide 45
9. Trouble shooting
Remedy
Install drivers.
Connect USB cable and
switch on CB-2000.
Switch on CB-2000.
Correct date and time
stamp to the required for-
mat or leave input blank.
Correct the capacitance
values to the required for-
mat or leave input blank.
Correct temperature value
to the required format or
leave input blank.
Add an extra line at the
beginning of the data file.
Reduce the number of
rows to 1000 or less.
Section
2.1
5
5
5.3
5.3
5.3
5.3
4.3