Contents Safety 1.1 Introduction 1.2 Intended Use 1.3 Authorized Operator 1.4 Safety measures on site Transport/ Installation 2.1 Technical data 2.2 Operation conditions 2.3 Transport / Storage 2.4 Unpacking 2.5 Set up and connection Operation Maintenance Euipment Putting out of Operation Wiring diagram Rev.
1 Safety 1.1 Introduction With the purchase of the Tissue Bath TFB 55 you decided for a high quality product of Medite GmbH. These instructions shall help you to use the instrument. Please read it thoroughly and follow the advices.
2 Installation and operation 2.1 Technical data Dimensions: Bath well: Ø 210mm, H 55 mm Overall: Ø 340mm, H 90 mm 30°C - 90º C Temperature: 230 V / 50 Hz / 200 VA Power supply: 115 V / 60 Hz / 200 VA T 1,6 Fuse: approx.
2.3 Transport/Storage The TFB 55 is delivered in a protecting box. Please move it only upright and avoid shock. 2.4 Unpacking Put the instrument on an even, solid surface and remove the protecting box and plastic wrap. If possible, keep the packing material to avoid damages in case of a later transport of the unit.
3. Operation Assembly of the instrument Place thermometer support with glass thermometer onto the rim of the bath opposite the control panel, making sure that the thermometer does not touch the floor of the bath. (The fixing and height adjustment of the thermometer on the thermometer support is done by the enclosed rubber ring being rolled over the thermometer from above) Fill with distilled or fully desalinated water up to 1-2 cm below the rim of the...
Page 8
Switch on the green main switch. Select temperature by turning the temperature regulator (knob). Always turn clockwise. Pilot lamp lights up and extinguishes when the temperature is achieved. Before the required temperature is achieved, heating will be switched off by the temperature regulator several times for short periods.
4 Maintenance Use gentle cleaning agents only for cleaning the bath unit since scouring powder would damage the black epoxy coating. In case of a non-functioning of the instrument, e.g. caused by voltage oscillations, please check whether the fuses in the back of the instrument are defect. CAUTION: Before dismounting the fuses pull out the mains plug! Should the defect not be eliminated by changing the fuses please call the service department.
6 Putting out of operation After end of the total period of use please deliver the equipment to an authorized disposal facility. The company Medite as manufacturer is also prepared to take back the equipment for disposal against payment for the disposal fee.