1. General description The following instructions for safety must be strictly observed to guarantee a safe operation of the instrument. The same precautions considered when using any technical equipment have to be observed. Failure to follow the recommended procedure as outlined in this manual or misuse of the instrument will void the manufacturer’s warranty.
If transport damages are detected, do not connect the instrument to the power supply. The bath should be cleaned thoroughly before starting operation. The USE 33 is equipped with cooling slots on both sides. In order to provide sufficient heat removal the instrument should be positioned in such a way that there is a minimum distance of 30 cm from the wall.
4. Instrument structure Bath Fig. 1: Front view USE 33 Temperature control cooling Water outlet Timer Ultrasound Fig. 2: Rear view USE 33 Power connection with fuse holder Fig. 3: right side USE 33 Air outlet Rev. 05 (01/2006) - SE - USE33E_Druck05.DOC...
Fig. 4: internal view USE 33 Water outlet. Minimum water level: must be above the water outlet 5. Operation The instrument can be switched to standby mode using the switch on the rear of the instrument. 5.1 Controller Cooling / Timer After switching on the instrument the actual temperature of the liquid is indicated in °C on the display.
Page 8
This key is used to switch the instrument on and off, respectively (standby mode) Indication of temperatures above 45°C No function in the operation of the USE 33 Indication: Cooling active Indication: Fan and Ultrasound active Rev. 05 (01/2006) - SE - USE33E_Druck05.DOC...
Page 9
This does not interfere with a possibly activated running timer. Set the duration time The duration time for the Ultrasonic Decalcifying Unit USE 33 is set by pressing the key SET twice and then pressing the keys Up or Down.
5.2 Ultrasonic operation / Programmable timer Features The timer is used to preset the duration time and to switch on or off the ultrasound, respectively. Key 1: UP This key is used to increase the set time interval which is indicated when the key SET TIME is pressed.
Page 11
Key 5: ALL ON This key is used to stop all control cycles. The timer settings are ignored and the ultrasonic module is switched on. This mode is shown in the display by the indication "ALL". Pressing this key again, the mode returns to the programmed settings, whereas those lines which are actually on remain on.
Page 12
Wednesday OFF1 OFF2 OFF3 Thursday OFF1 OFF2 OFF3 Friday OFF1 OFF2 OFF3 Saturday OFF1 OFF2 OFF3 Sunday OFF1 OFF2 OFF3 Monday OFF1... After having reached the desired program position, by pressing the key SET TIME the set time is indicated and can be changed by using the key UP or DOWN. When all keys are released, the new value is stored.
Adjustment advice The programming is made similar to that of an analogue timer, where one tag per day can be set each for start and stop time , but it is not a must. The endless loop in the programming cycle enables a quick access to the weekend days Saturday and Sunday by pressing the DOWN key.
First pull out the mains plug ! Using a screw driver it is easy to open the fuse socket and to change the fuse. Fuse socket 7. Technical specifications Ultrasonic decalcifying unit USE 33 Width : 450 mm Depth : 600 mm...
Page 15
Description: Cat. No.: 28-8252-00 Holder for glass jars for USE 33 Holder for specimen, no. 1 Qty. of specimen: 4 Specimen container size: d = 93 mm Height: 110 mm Container volume: approx. 800 ml. Max. size of specimen : 250 cm...
USE Application USE 33 2003-2004 Collection of USE 33 procedures. These procedures are compiled results of our users. Recommended common temperature range for decalcification of all samples is 16°C to 17°C. This temperature range can be adjusted in the USE 33.
Page 17
With other methods these samples min using USEDECALC optional (followed by running water or washing 5-10 min with fresh are not accessable. water in the USE 33) Lower jaw/ Weight: ? Insert No. 2 (46-1027-00, beaker made of glass) HE, Giemsa, Immunhistochemie Time is not optimised.
Page 18
Additional treatment: neutralize within approx. 3 h ® using USEDECALC optional (followed by with EDTA, only. running water or washing 5-10 min with fresh water in the USE 33) Missing: The proof of RNA where appropriate „Rapid Block Post conservation was not investigated Decalcification Procedure“ systematically.
Page 19
Fixation und identischer Entkalkung im USE 33, Dr. Stain Knolle, Frau Dr. Hege, Herr Balthasar, Klinikum Dessau. Prüfung der Eignung des USE 33 für den Einsatz an der Poliklinik für Kieferorthopädie Bonn, Prof. Götz, Frau Beil. CD34 Entkalkung mit dem Ultraschall für immunhistologische Nachweise, MHH, Dr.
Page 20
295 x 235 mm Weight : 15 kg Voltage : 230V / 50Hz / 750VA Wollenweberstr. 12 - D-31303 Burgdorf - Tel: +49 5136 888 4 0 Fax: + 49 5136 888 4 55 sales@medite.de www.medite.de Rev. 05 (01/2006) - SE - USE33E_Druck05.DOC...