GPS
®
III System
Processing Instructions

Advertisement

loading

Summary of Contents for Zimmer Biomet GPS III

  • Page 1 ® III System Processing Instructions...
  • Page 2 Step 1: Blood Draw 54 ml of blood is drawn from the patient according Attach supplied 18 gauge needle to 60 ml syringe. to facility protocol. Withdraw 6 ml of ACD-A (citrate anticoagulant). *Mini GPS® III System: 27 ml of blood is drawn from *Mini GPS®...
  • Page 3 Step 2: Load Blood Unscrew cap on center port No.1 and discard green packaging post. Slowly load blood into center port.
  • Page 4 Step 3: Balance Push Open/Stop button on control panel. The Fill GPS® III counterbalance (800-0508) with 60 “unlocked” indicator will illuminate. Turn latch ml of sterile saline and place into opposite side of counterclockwise to open lid. Place tube into centrifuge.
  • Page 5 Step 4: Spin Close the lid by rotating the lid latch clockwise. “Latched” indicator will illuminate. Set speed to 3200 RPM and time to 15 minutes. Press green button to start spin. Once spin is complete, press red button to illuminate the “Unlocked” indicator. Twist latch counterclockwise to open lid.
  • Page 6 Step 5: PPP Extraction To extract the platelet poor plasma (PPP), remove yellow cap on side port No. 2 and connect the 30 ml syringe. Slowly tilt the tube while withdrawing PPP. Replace yellow cap.
  • Page 7 Step 6: PRP Suspension and Extraction Remove red cap on side port No. With 10 ml syringe attached, Extract the remaining platelet 3 and connect 10 ml syringe. suspend the platelets by shaking rich plasma (PRP) contents into the tube for 30 seconds. the attached 10 ml syringe.
  • Page 8 Plasmax Plasma ® Concentration System Processing Instructions...
  • Page 9 Step 1: Load Plasma Plasmax® Plus: Unscrew yellow cap on port No. Spin paddle by hand until plasma is fully mixed into 1 of plasma concentrator and fill with 25 ml of beads. Mixing is complete when the white beads are PPP.
  • Page 10 Step 2: Spin and Extract Plasmax® Plus: Place concentrator into centrifuge Plasmax® Plus: Unscrew red cap on port No. 2. with the empty Plasmax® Plus counterbalance Connect sterile 10 ml syringe on port No. 2 and extract (800–0512). They should be placed on opposite approximately 10 ml of concentrated plasma.
  • Page 11 It is not to be redistributed, duplicated or disclosed without the express written consent of Biomet. For product information, including indications, contraindications, warnings, precautions and potential adverse effects, see the Biomet Biologics website at www.biometbiologics.com. ©2018 Zimmer Biomet 1792.1-US-en-REV0118...