Emissivity for OMEGA TI-120 series includes full screen setting and regional setting.
1. Click on Emissivity icon
2. For customized material emissivity setting, click
and
or scroll
to adjust (0.01-1.0)
3. For known material, scroll menu to select emissivity value accordingly
4. For unknown material, refer to "Emissivity List" in this manual for detail
5. For unknown material and which not included in this manual, refer to 13. for instruction
!Note: Regional Emissivity Setting refer to 4.1.3.3 for detail
4.2.4.2 Reflection Tempreature
Reflection temperature is used to compensate or correct the thermal radiation reflected from the measured
target
When there is an interference source near the measured target which is higher than the temperature of
the measured target, it is very important to set the reflection temperature parameters correctly for accurate
temperature measure
Click to enter appopriate value to save
!Note: Always avoid reflection interference if possible to improve measurement accuracy