OPERATING INFORMATION
The TDI Nitro-Save® Humidity Control Monitor is designed to control Nitrogen/CDA flowrate to accurately lower
humidity in desiccator storage cabinets. This unit displays temperature as well, but is not calibrated to monitor it.
CAUTION
The NS16 Humidity Control Monitor is ONLY to be used with the solenoid flow control valve that is supplied. This
valve is designed and specially wired to operate this unit. Use of any other device will void the 6 month warranty
and may cause serious damage. This unit is designed and calibrated to be used with nitrogen or CDA.
Stored parts should be a minimum of 1 ½" away from the sensor probe.
START UP
To start the Nitro-Save®, plug the AC adapter into an electrical outlet, and then connect it to the port located on the
back of the monitor. The monitor will automatically start to measure and display the temperature and humidity
readings.
NOTES:
Use only with the supplied AC adapter.
FOR INDOOR USE ONLY. Contact with water or use in dusty conditions may cause damage or affect
calibration.
Must Operate at 80 PSI (min) to 100 PSI (max) for Optimum Performance
To begin and display the main menu press
3-31-2017
TDI Humidity Control Monitor
#NS16 NITRO-SAVE®
Page 1
3351 East. Hemisphere Loop
Tucson, AZ 85706
Phone: (520) 799-8000
Fax: (520) 799-8002
E-mail: sales@tdiinternational.com
www.tdiinternational.com

Advertisement

loading

Summary of Contents for TDI Nitro-Save NS16

  • Page 1 TDI Humidity Control Monitor #NS16 NITRO-SAVE® OPERATING INFORMATION The TDI Nitro-Save® Humidity Control Monitor is designed to control Nitrogen/CDA flowrate to accurately lower humidity in desiccator storage cabinets. This unit displays temperature as well, but is not calibrated to monitor it. CAUTION The NS16 Humidity Control Monitor is ONLY to be used with the solenoid flow control valve that is supplied.
  • Page 2 MAIN MENU The Nitro-Save® has a simple menu system that is easy to navigate.  Use the up and down arrows ( ) to scroll through the main menu options.  The current option will be highlighted.  Press to select the highlighted option. ...
  • Page 3 CHANGE SETTINGS To change the high humidity setpoint, use the up and down arrows ( ) to highlight Change Settings and press to proceed. This display will show: NOTE: The high humidity setpoint is factory programmed to 20%RH, which represents general electronics requirements.
  • Page 4 CUSTOMER SERVICE Warranty TDI International, Inc. warrants this product to be free from defects in materials and workmanship for a period of six (6) months from the date of purchase. TDI International, Inc. will repair problems that occur during the warranty period, and that have not been caused by abuse or misuse, at no charge to the customer.