Specifications Summary
See Users Manual on CD for full specifications.
Feature
566
-40 °C to 650 °C
IR Temperature
(-40 °F to 1202 °F)
Range
> 0 °C: ±1 % or ±1.0 °C
Accuracy above
0 °C (32 °F)
(> 32 °F: ±1 % or ±2 °F), whichever is greater
K T/C
-270 °C to 1372 °C (-454 °F to 2501 °F)
Temperature
Range
-270 °C to -40 °C: ±(1 °C + 0.2 °/1 °C)
(-454 °F to -40 °F: ±(2 °F + 0.2 °/1 °F))
K T/C Input
-40 °C to 1372 °C: ±1 % or 1 °C (-40 °F to
Accuracy
2501 °F: ±1 % or 2 °F), whichever is greater
Distance:Spot
30:1
(90 % energy)
Laser sighting
Offset single laser <1 mW
Minimum spot
19 mm
size
Emissivity
Digitally adjustable from 0.10 to 1.00 by 0.01
Data storage
20 points
Communication
None
Operating Altitude
3000 meters above mean sea level
Storage Altitude
12,000 meters above mean sea level
10 % to 90 % RH non-condensing up to 30 °C
Relative Humidity
(86 °F)
Operating
0 °C to 50 °C (32 °F to 122 °F)
Temperature
Storage
-20 °C to 60 °C (-4 °F to 140 °F)
Temperature
Power
2 AA /LR6 Batteries
12 hours with laser and backlight on; 100 hours
Battery Life
with laser and backlight off, at 100 % duty cycle
(continuously on)
Infrared Thermometers
Specifications Summary
568
-40 °C to 800 °C
(-40 °F to 1472 °F)
50:1
19 mm
99 points
USB 2.0
2 AA /LR6 Batteries or
USB connection when
used with a PC
5