  • Page 1 Coating Thickness Tester User Manual Group4 (44) Group4 (44)
  • Page 2: Table Of Contents

    Coating Thickness Tester User Manual Contents Page 1. General Information ............1-1. Application ..............1-2. Description of the Gauge..........1-3. Supply Schedule............. 1-4. Probe................1-5. Specifications..............1-6. Front View..............1-7. LCD Display..............2. Ready to start............... 2-1. Power Supply..............2-2. Replacing the Battery............. 2-3.
  • Page 3: General Information

    Coating Thickness Tester User Manual 1. General Information The coating thickness gauges work either on the magnetic induction principle or on the eddy current principle, depending on the type of probe used. You can select the type of probe via MENU system, or it will work automatically.
  • Page 4: Application

    Coating Thickness Tester User Manual 1-1. Application • This compact and handy gauge is designed for non-destructive, fast and precise coating thickness measurements. The principal applications lie in the field of corrosion protection. It is ideal for manufacturers and their customers, for offices and specialist advisers, for paint shops and electroplaters, for the chemical, automobile, shipbuilding and aircraft industries and for light and heavy engineering.
  • Page 5: Specifications

    Coating Thickness Tester User Manual 1-5. Specifications Sensor probe Working principle Magnetic induction Eddy current principle 0~2000µm Measuring range 0~2000µm 0~78.7mils 0~78.7mils Guaranteed tolerance 0 ~ 1000 um 0 ~ 1000 um (of reading) (±2%±2um) (±2%±2um) 1000~ 2000 um 1000~ 2000 um (±3.5%) (±3.5%) 0~39.3mils...
  • Page 6: Front View

    Coating Thickness Tester User Manual 6. Front View 1-LCD display 2- Power ON/OFF key 3-Left Key: For choose and OK 4-Up key: For upward selection and working mode switch 5-Right Key: For return or delete the last measured data 6-Down key: For downward selection or working mode switch 7-Probe 8- Lanyard hole 9-Battery cover...
  • Page 7: Ready To Start

    Coating Thickness Tester User Manual 2. Ready to start 2-1. Power Supply For power on, please press and hold button. If no LCD display: please check if the battery is correctly installed or it is in low battery. Display: Please replace battery immediately. Note that the gauge will make faulty measurements if the voltage is very low.
  • Page 8: Settings

    Coating Thickness Tester User Manual 2-3-1-2. select measure mode (Probe) The probe can work in three modes: AUTO: The probe can automatically select the working mode. When placed on steel (magnetic substrates), it will work in magnetic induction principle. When placed on no-ferrous metals, it will work in eddy current principle.
  • Page 9: Auto Power Off

    In the case, enter the code 0000 into your phone/PC. The coating thickness gauges includes Bluetooth® wireless technology allowing wireless data transfer to properly equipped smartphones or tablets running iOS or Android operating systems. Download the app to your smartphone or tablet by going to https://play.google.com/store/apps/details?id=com.cem.ictt.activities&hl=zh_EN...
  • Page 10: Adjust Gray Scale (Contrast)

    Coating Thickness Tester User Manual • Press left button to enter the menu • Press up/down button to select the setting • Press left button to enter the setting • Press up/down button to select Bluetooth • Press left button to enter •...
  • Page 11: Calibration And Measurement

    Coating Thickness Tester User Manual • Press up/down button to view each data • Press right button to return or press left button to delete this group • data. 5. Calibration and Measurement The gauge can calibrate Zero-point of Fe or Non-Fe, or delete wrong calibrated zero-point of Fe or Non-Fe.
  • Page 12: Delete Functions

    Coating Thickness Tester User Manual 8. Delete Functions In MENU system, you can find following function: • Delete current data: in measurement interface, you can delete the • last data by pressing right button. • Delete all data: Enter menu-view; you can delete all data and statistics •...
  • Page 13 Coating Thickness Tester User Manual Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. This device complies with Part 15 of the FCC Rules.

Table of Contents