Press [CFG], then [F1] BCKLTE.
Use the [F2] ▼and
Press [F1] Set to save the setting.
3.3.2

CONTRAST

The display contrast can be adjusted as needed. Temperature may affect the
contrast. The contrast may need to be turned down in hot conditions or turned up
in cold conditions.
Press [CFG], then [F2] CONTRST.
Use [F2] ▼and
Press [F1] Set to save the setting.
3.3.3

VOLUME

The key push beep and probe volumes are adjustable as needed.
is adjustable in the probe function.
Press [CFG], then [F3] VOLUME
Use [F2] ▼and
Press [F1] Set to save the setting
4

SENDING TONE

The TriTone Transceiver provides tone sending options for identification of cables
or cable pairs.
4.1

FREQUENCY SELECTION

The TriTone Transceiver generates 577 Hz or 987 Hz tone. Select a tone that is
compatible with the tone probe(s) that will be used. The TriTone Transceiver
retains the previous selection when the power is turned on.
set from the Main screen or any tone operating screen.
Press the [FREQ] key
Use [F2] ▼and [F3] ▲ keys to select the desired frequency
Press [F1] SET.
4.2

LEVEL SELECTION

The TriTone Transceiver offers transmit tone levels at Normal, High, and
Maximum. The lowest level that allows identification of the cable pairs should
normally be used. Higher tone levels may increase inductive coupling to adjacent
pairs and make it more difficult to identify the correct pair.
Allied Analogic, Inc.
▲ keys to adjust the backlight
[
]
F3
▲ keys to adjust the contrast
[
]
F3
▲ keys to adjust the KeyPush volume
[
]
F3
5
The probe volume
The frequency can be
Copyright © 2019