Elcometer 138 User Manual page 6

Bresle salt kits
Hide thumbs Also See for 138:
Table of Contents

Advertisement

R
3
TEST PROCEDURE ISO
6
During a suitable period of time , without removing the needle, suck
and re-inject the solution at least four times .
7
At the end of the period, extract as much solution as possible and
remove the syringe from the patch .
8
Analyse the solution using the Elcometer 138 Meter, see Section 3.4
on page en- . Inject the sample directly into the sensor cell. Rinse the
6
sensor cell several times with the solution to be measured before
taking the reading.
4
See Appendix A on page en-15 for further instructions on using the Elcometer 138
Meters.
3.3 AFTER TEST
1
Record the temperature of the solution.
2
Remove the patch from the surface and clean the surface. If
required, any adhesive residue from the patch left on the test surface
can be removed by wiping with a cloth moistened with a suitable
solvent. Ensure that the solvent will not damage the surface before
use.
3
Rinse all components of the test kit, other than the patch, in fresh,
pure water. The components can then be used again.
c
On un-pitted blast-cleaned areas, a period of 10 minutes has been found to be satisfactory, though this time should be agreed by the interested parties.
d
During steps 6 and 7, it is essential that no solution is lost. If any solution is lost, the test shall be rejected.
en-5
:
8502-6 /
ISO
8502-9 (continued)
c
d
d
www.elcometer.com