Ergosana SanaBike 500 easy User Manual

Bike ergometer

Advertisement

Quick Links

Bike ergometer SanaBike 500 - easy
User guide
Article-Nr.: 2.520132b
Revision: 10/2018
Page 1 of 21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SanaBike 500 easy and is the answer not in the manual?

Questions and answers

Summary of Contents for Ergosana SanaBike 500 easy

  • Page 1 Bike ergometer SanaBike 500 - easy User guide Article-Nr.: 2.520132b Revision: 10/2018 Page 1 of 21...
  • Page 2 SanaBike 500 easy CE Conformity Declaration ergosana GmbH herein declare that the medical products (Class IIa) of the ergometer system SanaBike 500 "easy" comply with the relevant requirements of the EC medical devices directive 93/42/EEC, Annex I. This declaration loses its validity if the above devices are modified without ergosana’s consent.
  • Page 3: Table Of Contents

    SanaBike 500 easy Tabel of contents ALLGEMEINES ............................5 Intended use ..............................5 Indications ............................... 6 1.2.1 Indications for cardiac rehabilitation ........................6 Contra-indications ............................6 1.3.1 Contra-indications for cardiac rehabilitation ......................7 Abort criteria ..............................7 Instruction ............................... 7 Maintenance ..............................
  • Page 4 SanaBike 500 easy MAINTENANCE AND TROUBLE SHOOTING ..................18 Messtechnische Kontrollen (MTK) und Sicherheitstechnische Kontrollen (STK) ..........18 Checking and setting the supply voltage ......................18 Changing a mains fuse ............................ 18 Eliminating electromagnetic interferences ..................... 18 Disposal ................................19 CHARACTERISTICS FOR THE BRAKING MOMENT CONTROL ...............
  • Page 5: Allgemeines

    SanaBike 500 easy Allgemeines The bike ergometer SanaBike 500 „easy“ is a modern high-performance ergometer and is intended to be used for remote operation (training programs provided by PC, ECG unit or similar). Alternatively, it may also be operated in manual mode, in which the load can be changed by the user in steps of 5 Watt on the display.
  • Page 6: Indications

    SanaBike 500 easy Indications Possible indications for exercise ECGs: • Diagnostic clarification of chest pain (angina pectoris including vasospastic angina) for myocardial • ischaemia (insufficient blood supply) or for coronary heart disease (coronary artery disease) • For patients with cardiac risk factors such as suspected coronary artery disease and arterial hypertonia (high blood pressure) •...
  • Page 7: Contra-Indications For Cardiac Rehabilitation

    SanaBike 500 easy Relative contra-indications: • serious valvular heart defect • manifest cardiovascular diseases • left main stenosis • known electrolyte imbalance • arterial hypertonia (RR >200 mmHg syst. / >110 mmHg diast.) • tachyarrhythmia or bradyarrhythmia • hypertrophic cardiomyopathy and other obstructions of the outflow tract •...
  • Page 8: Product Description

    SanaBike 500 easy Product description The ergometer SanaBike 500 easy is intended for remote operation together with a master device (PC, ECG unit or similar). It may also be operated in manual mode, where the load can be defined in steps of 5 watts by the user.
  • Page 9: Technical Specifications

    SanaBike 500 easy Technical specifications Drive mechanism Almost noiseless and maintenance-free, with Poly-V belt (no chain) Computer-controlled brakes with permanent measurement of torque and Braking principle braking performance which is independent of revolutions per minute. Performance/ - 1 to 20 watts (range is dependent of revolutions per minute)
  • Page 10: Installation

    SanaBike 500 easy Installation Location • Install the device in a suitable position (refer to safety instructions in section 0). • The unit must not be stored or operated in wet, moist or dusty surroundings. • The unit must not be exposed to direct sunlight or other sources of heat.
  • Page 11: Unit Components

    SanaBike 500 easy Unit components The console mounted on the handle bar is rotatable by 180°. Depending on the application, it can be facing either the examiner or the user. At the front, a high-resolution touchscreen provides all important information at a glance as well as all ele- ments to control and program the ergometer.
  • Page 12: Adjusting The Saddle Height And Handle Bars

    SanaBike 500 easy Adjusting the saddle height and handle bars The saddle can be adjusted upwards or downwards for optimal exercise position. It can be adjusted for heights between 120 and 210 cm. T-handles that protrude from under both the handlebar and saddle pipe are used for adjustment (for manual saddle height adjustment).
  • Page 13: Safety Notes

    SanaBike 500 easy Safety notes Precautions during operation/use • Before using the unit, make sure that the medical product consultant has conducted an introduction with regard to function and safety precautions. • The unit must not be used if there are any doubts as to it being isolated from earth or the suitability of the mains cable.
  • Page 14: Initial Operation

    SanaBike 500 easy Initial operation Once the ergometer has been prepared as detailed in chapter Fehler! Verweisquelle konnte nicht ge- funden werden., it can be connected to a master device (ECG, PC or similar). Ergometer adjustment Before the start of each training session, the sitting position needs to be individually adjusted to the patient.
  • Page 15: Operation Modes

    SanaBike 500 easy Operation modes Remote operation When the load is automatically controlled by an exercise/training program, this is done by the ergometry control program of an external master device such as PC, ECG unit or similar, depending on the application.
  • Page 16: Starting, Adjusting And Terminating Manual Operation Mode

    SanaBike 500 easy When pressed, the screen changes to manual mode to manually control the load with the default start value of 25 watts or the last defined load value. This value can be changed by the user by means of the two arrow symbol buttons "down"...
  • Page 17: Cleaning

    SanaBike 500 easy Cleaning Cleaning the device The surface of the casing can be cleaned with a soft cloth that is dry or moist. Commercially available clean- ing agents for household appliances can be used. Disinfection The saddle and the handlebar can be treated with the disinfectant Incidin Plus fluid as needed.
  • Page 18 SanaBike 500 easy Maintenance and trouble shooting Messtechnische Kontrollen (MTK) und Sicherheitstechnische Kontrollen (STK) The unit’s measuring technology should be checked every 24 months. The following verification must be performed: 1. Check of the overall mechanical condition of the ergometer 2.
  • Page 19 802.11 a/n 5100-5800 Disposal Devices that are no longer usable can be returned to ergosana for disposal. Alternatively, the device can be taken to an approved disposal location. • The control unit of the device contains a buffer battery, which must be disposed of separately.
  • Page 20 SanaBike 500 easy Characteristics for the braking moment control Family of characteristics for the area of the braking moment control Load [Watt] Article-Nr.: 2.520132b Revision: 10/2018 Page 20 of 21...
  • Page 21 SanaBike 500 easy Technical Customer Service and Sales Locations ergosana products are also sold by exclusively authorised Schiller agents with SCHILLER brand name. The agents are trained in the service of our devices. Please contact one of these special- ist dealers if your device requires servicing.

Table of Contents