Page 1
Ergolift 2 Instructions for Use 001-06109-EN rev.0 February 2010...
Page 2
Printed in Canada 2010 BHM medical products are patented or patent pending. Our policy is one of continuous development, and we therefore reserve the right to make technical alterations without notice. The content of this publication may not be copied...
Foreword General Information Thank you for purchasing the Ergolift 2. Your Ergolift 2 is part of a series of quality products Please read this manual in its entirety before designed especially for hospitals, nursing using your Ergolift 2. The information in this homes and other health care uses.
Huntleigh Healthcare Ltd. carried out. 310-312 Dallow Road Luton, UK The Ergolift 2 must always be handled by a LU1 1TD trained caregiver, who must attend the patient Definitions Used in this Manual during lift application and operation, and in...
Required for Patient Transfer WARNING: The manufacturer cannot ensure full safety for a floor lift or an accessory of BHM Medical’s passive and active series of which the life span has been exceeded. lifts are designed for safe usage with one caregiver.
General Information Symbols Used Symbols Key to Symbols This symbol is accompanied by the name and the address of the authorized representative in the European Community. This symbol is accompanied by a date to indicate the date of manufacture and by the address of the manufacturer. This symbol indicates the products comply with the medical device directive 93/42/EEC.
Ergolift 2, in order to avoid BHM Medical’s warranty. Use only BHM injuries attributable to the use of inadequate Medical slings and accessories to maintain parts.
Do not splash or expose electric parts of • Maintenance” section of this manual. the device to water or moisture. Check nameplate for input voltage and WARNING: Do not overload the Ergolift 2 • beyond the approved lifting capacity of the frequency requirements.
Safety Instructions Battery Safety Practices Fire and Explosion Prevention Be careful not to drop the batteries. • Do not charge the lift in an unventilated • WARNING: Do not place or store the battery area. under direct sunlight or near a heat source. Do not expose the batteries.
How to Use the Ergolift 2 How to Use the Ergolift 2 Move the lift over to the patient, so the end of the boom is approximately 15 cm (6 in) Transferring a Patient away from the patient. For a patient in a...
Directional Lock being transferred, the emergency manual lowering device provides a safe way of The Ergolift 2 is equipped with a directional lowering the patient. lock which allows a better manoeuvrability. There must be a load applied on the actuator...
How to Use the Ergolift 2 Anti-crush Feature CAUTION: To prolong battery life, plug the unit in an outlet WHENEVER THE LIFT IS To prevent damage or injury, the motor on NOT IN USE. the lift has a limit switch that recognizes The battery will not overcharge.
How to Use the Ergolift 2 WARNING: Leave the unit plugged in. WARNING: Always use the same color loop At minimum, charge the battery until the on the rear shoulder straps for both sides light is green before using it again. This of the sling.
To ensure a better rolling surface for the Note all services or repairs to the wheels, clean THEM every 4 months. Ergolift 2 or its accessories in the log Remove any hair or debris from the wheel. book of this manual. Have the document signed by a qualified technician.
Page 17
Disinfection, Cleaning and Maintenance To replace the batteries: 1. Press the emergency stop button to ensure the lift will not accidentally operate while you are working on it. 2. Flip lift onto the mast to access under the base. 3. Unscrew the 7 bolts which hold the base cover (use an 8 mm socket).
Disinfection, Cleaning and Maintenance 10. Replace actuator, clevis pins and cotter 4. If the light does not illuminate, contact pins. Follow proper direction (see your local dealer or representative for drawing). Make sure the actuator wire is assistance. placed down next to the battery. Replace the fuse with the following: Fuse in North America: GMC 0.5A 250 V (when...
Disinfection, Cleaning and Maintenance 4. Remove the two clevis pins on each end of the damaged/inoperable motor. 5. Unplug the actuator and replace it by the new one. 3. Replace the actuator with a new one (see replacement parts). Plug the actuator into 6.
Annual Inspection 2. Remove the bolt and locknut that secure the existing front caster assembly to the The Ergolift 2 and the accessories must be fork. inspected annually by a qualified technician. 3. Remove all debris and clean any dust.
Page 21
Preventive Maintenance Schedule Preventive Maintenance SchedulePreventive Maintenance Schedule The Ergolift 2 is subject to wear and tear, and the following actions must be performed when specified to ensure that the product remains within its original manufacturing specifications. WARNING: The points on this checklist are the minimum the manufacturer recommends. In some cases more frequent inspections should be carried out.
Page 22
Preventive Maintenance Schedule Preventive Maintenance SchedulePreventive Maintenance Schedule INITIALLY BEFORE EVERY 2 EVERY 6 EVERY ITEM EVERY MONTHS MONTHS YEAR THE BOOM Check all hardware and swivel bar supports. Inspect for cracks or deflections. Inspect bolted joints of boom for wear. Inspect to ensure that the boom is centered between the base legs.
Page 23
Preventive Maintenance Schedule Preventive Maintenance SchedulePreventive Maintenance Schedule INITIALLY BEFORE EVERY 2 EVERY 6 EVERY ITEM EVERY MONTHS MONTHS YEAR Check that caster brakes operate efficiently. Overhaul casters and repack with all purpose grease. Inspect surfaces for scatches/chips. Clean surfaces according to the “Lift Cleaning and Care” section.
Troubleshooting Troubleshooting WARNING: Only a qualified technician is authorized to open the Ergolift 2. Repairs made to the lift by someone other than a qualified technician may cause serious ingjury. PROBLEMS TO CHECK • Make sure the handset is plugged correctly into the unit.
Technical Specifications Technical Specifications PRODUCT INFORMATION ERGOLIFT 2 Weight of floor lift 54.5 kg (120 lb) Weight of assembled base 147.4 kg (67 lb) Weight of Two-point spreader bar 1.4 kg (3 lb) Weight of scale 1.1 kg (2.5 lb)
• Ensure the lift is serviced regularly as the maintenance inspection checklist on page 20. • Contact an authorized BHM Medical Distributor / Service Agent immediatly if there are any problems with the operation of the device. • Ensure the log book is completed and signed.
Page 28
Log Book Service Type Pre-delivery Periodic inspection Minor Major Condition report : Action taken : Date : Inspected by: _________________ ____________________ ____________________ Printed Signature Service Type Pre-delivery Periodic inspection Minor Major Condition report : Action taken : Date : Inspected by: _________________ ____________________ ____________________ Printed...
Page 29
Log Book Service Type Pre-delivery Periodic inspection Minor Major Condition report : Action taken : Date : Inspected by: _________________ ____________________ ____________________ Printed Signature Service Type Pre-delivery Periodic inspection Minor Major Condition report : Action taken : Date : Inspected by: _________________ ____________________ ____________________ Printed...
(other than installation performed by BHM Medical) operation or care (including but not limited to abuse, misuse, failure to provide reasonable or necessary maintenance, unauthorized repairs or any alterations to the products). BHM Medical Inc. slings are void of warranty if not laundered as per instructions on the Sling Label.
Page 31
In no event shall BHM Medical’s liability for breach of warranty, or for any delay in the performance of this warranty due to cause beyond the control of BHM Medical, breach of contract, negligence or strict liability exceed the cost of the product covered hereby.
Need help?
Do you have a question about the Ergolift 2 and is the answer not in the manual?
Questions and answers
What size of casters are on a ergo lift 2
What size is the casters on a Ergo-2 lift