User Guide
ProCair 3000

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ProCair 3000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Medistore ProCair 3000

  • Page 1 User Guide ProCair 3000...
  • Page 2 Important Notice Before operating this medical equipment, it is important to read this User Guide and understand the operating instructions and safety precautions. Failure to do so could result in patient injury and/or damage to the product.
  • Page 3: Table Of Contents

    Page Safety Precautions Product Overview Installation Unpacking & Inspection System Installation Control Unit Installation Operation Control Unit Panel Removal and Transport Function Troubleshooting Maintenance and Cleaning 12-13 Technical Specifications Warranty Information...
  • Page 4: Safety Precautions

    120kg. are isolated from the internal metal motor components with protecting insulation. Only the Control Unit and Mattress combination as recommended by Medistore should be used, otherwise the correct function of this product cannot be guaranteed.
  • Page 5: Safety Precautions

    Control Unit. If parts need to be between the system and other devices repaired, circuit diagrams and parts may occur. If interference is suspected, lists can be provided by Medistore to move equipment from sensitive devices or suitably qualified service personnel. contact Medistore.
  • Page 6: Product Overview

    Product Overview ProCair 3000 Alternating Mattress Overlay System The ProCair 3000 is an Alternating Mattress Overlay System providing pressure application and relief for patients with, or vulnerable to, pressure ulcers. It is designed to be used on top of an existing single bed mattress and can be used on both standard and profiling bed frames.
  • Page 7: Installation

    Overlay and accessories from the shipping cartons. Inspect all items for any damage that may have occurred during shipping. Any damage or missing components should be reported to a Medistore Service Centre as soon as possible. The box should contain the following items: •...
  • Page 8: System Installation

    Installation System Installation Control Unit Installation The following describes the procedures to a) Position Control Unit by its hanging follow upon setting up the system for the hooks over foot board of the bed or on first time. the floor under the bed. a) Remove all covers and sheets from the existing mattress.
  • Page 9: Operation

    Basic Features Control Panel Establishing Pressure (inclined patient) The ProCair 3000 Analogue Control Unit front panel is shown in the diagram below. When moving the patient to a more upright position, pressure may need to be increased (approximately 20%) to provide added support and to avoid ‘bottoming out’.
  • Page 10: Removal And Transport Function

    Operation System Removal Transport Function Overlay Removal Overlay Transport Function 1. Switch off the Control Unit. (If 1. When preparing for transport mode, it applicable reset the alarm indication is best if the patient is not inclined but on the power-down alarm. Switch off supine.
  • Page 11: Troubleshooting

    10-12 minutes (one full cycle) before checking the increase was sufficient. Check effective system performance by conducting a ‘bottoming out’ test as described on page 10. If the problem cannot be resolved, please contact a Medistore Service Centre for further support.
  • Page 12: Maintenance And Cleaning

    Maintenance & Cleaning Infection Control and routine cleaning 1. Apply an intermediate level disinfectant must be carried out in accordance with (or a solution of Sodium Hypochlorite your local Infection Control Policy. or similar up to 10,000 ppm available It is suggested that all disinfection chlorine) to the Top Cover upper be done with a high grade surface either by spraying or by...
  • Page 13: Maintenance And Cleaning

    Air filters are available from a Medistore with oxygen or nitrous oxide. Service Centre. Preventative inspection and calibration is not required.
  • Page 14: Technical Specifications

    Technical Specifications Cycle Control: Purpose designed distributor valve supplying operating air to the inflatable cells Cycle Time 10-12 minutes Supply Voltage 230V, 50Hz 120V, 60Hz 0.2A for Control Unit 0.01 - 6A for power down alarm (availability of supply monitored by in-line power-down alarm which provides failure indication) Fuse Rating Mains Plug - 3A, Control Unit - 1A...
  • Page 15: Warranty Information

    Medistore Service Centre by prepaid You are also entitled to have the goods transportation. Medistore will return the repaired or replaced if the goods fail to be...
  • Page 16 Medistore has a policy of continuous product improvement and reserves the right to amend specifications presented in this brochure. This product carries the CE mark in accordance with EC Directive on Medical Devices (93142lE EC) . Control Number 95CG. Classified by Underwriter’s Laboratory Inc.