Specifications - 3M Versaflo TR-600 Series User Instructions

Chargers and battery packs for the powered air purifying respirator assemblies
Hide thumbs Also See for Versaflo TR-600 Series:
Table of Contents

Advertisement

Available languages

Available languages

Inspection
An inspection of the entire charger assembly and completion of all necessary replacement or repairs must be made before each use.
1. Inspect the power base and cradle for cracks or other damage.
2. Inspect the power cord for frayed wires or other damage.
3. Inspect the battery pack case for cracks or other damage.
Remove from service and dispose of properly if any damage is noted.
4. Ensure electrical contacts are clean and dry.
5. Ensure the battery is able to securely connect to the motor/blower unit.
Storage
Range: -22°F to 122°F (-30°C to 50°C)
Optimal: 59°F (15°C)
Disposal
Dispose of Lithium-ion battery packs according to local environmental regulations. Do not crush, disassemble, dispose of in standard waste bins, in
a fire, or send for incineration. Failure to properly dispose of battery packs may lead to environmental contamination, fire or explosion.
The charger contains electronic components. Do not throw in trash. Dispose of product as directed by local regulations.

SPECIFICATIONS

Battery Pack
Type
Weight
• 3M™ Standard Battery TR-630
• 3M™ High Capacity Battery TR-632
Electrical
• 3M™ Standard Battery TR-630
• 3M™ High Capacity Battery TR-632
Service Capacity for New Battery Pack after
Completing Charge Cycle
• 3M™ Standard Battery TR-630
• 3M™ High Capacity Battery TR-632
Battery Pack Charging Temperature
Conditions
Time to recharge
• 3M™ Standard Battery TR-630
• 3M™ High Capacity Battery TR-632
Battery Life
Battery Pack Storage Temperature
Battery Pack Shelf Life
Maximum Continuous Time on Charger
W WARNING
Lithium-ion
NOTE: Consult a transportation specialist for any requirements or limitations prior to transporting
lithium-ion battery packs.
TR-630 and TR-632 are shipped compliant with UN 38.3
Approx. 1 lb (450 grams)
Approx. 1.3 lbs (620 grams)
Nominal 11.1 Vdc, 5.2 Ah., 58 Wh
Nominal 11.1 Vdc, 7.8 Ah., 87 Wh
Approx. 7-13 hours*
Approx. 10-19 hours*
* Estimated system run time based on testing with a new battery and filter at 68°F (20°C). Longest
run times with HE filter. Actual system run time may be extended or shortened depending on flow
level selected, system configuration and environment.
Optimal: 68°F to 77°F (20°C to 25°C)
Range: 32°F to 104°F (0°C to 40°C). Do not charge outside of this range
Typically less than 3.5 hours
Typically less than 4.5 hours
Approx. 250 full charge/discharge cycles equivalents until 80% of the original capacity*
* Over the course of the first year of service when used under recommended conditions
Range: -22°F to 122°F (-30°C to 50°C)
Optimal: 59°F (15°C)
1 year suggested shelf life from the date of manufacture when stored in original packaging. The LOT
code is on the battery housing. LOT number format (YYWWX); YY = year of manufacture; WW = week
of year. (See Fig. 5)
Battery pack may be stored indefinitely on the charger
4
(English)

Hide quick links:

Advertisement

Table of Contents
loading