Maintenance - Quoticare BP101U Instruction Manual

Fully automatic upper arm style
Table of Contents

Advertisement

Troubleshooting and Maintenance
Troubleshooting and Maintenance

Maintenance

To protect your unit from damage, please observe the following:
Do not disassemble the unit.
Do not fold the cuff or tubing tightly.
Do not wash the cuff or immerse it in water.
Do not use volatile liquids to clean the main unit.
Do not use petrol, thinners or similar solvents to clean the cuff.
Do not subject the unit to strong shocks or vibrations (for example,
dropping the unit on the floor).
Do not subject the main unit and the cuff to extreme temperatures,
humidity, moisture or direct sunlight.
Do not carry out repairs of any kind yourself. If a defect occurs,
Please contact the local distributor.
If needed, please use ultraviolet radiation for disinfection.
gasoline
thinner
The unit should be cleaned with a soft, dry cloth.
Use a soft, moistened cloth and soap to clean the arm cuff.
Substitution of a cuff different from that supplied might result in
measurement error.
We will provide circuit diagrams, component part lists, descriptions,
calibration instructions, or other information to assist service
personnel to repair parts.
The service personnel needs to be the personnel who is
trained by the manufacture and has got the relevant
qualification certificate.
The accuracy of this blood pressure monitor has been carefully
tested and is designed for a long service life. We recommend that the
performance be checked every 2 years and after maintenance and
repair, by utilizing the manometer mode and verifying the accuracy
of the manometer at least at 50 mmHg (6,7 kPa) and 200 mmHg
(26,7 kPa). Please contact the ocal distributor for details.
24
BP101U
2×90 Sets of Measurement Values
3 AA Alkaline Batteries
Within 2 Minutes
2 Years
Cuff, Instruction Manual
Note: Subject to technical modification without prior notice.
180
Battery Indicator
105
(50 ~ 104 F )
10~40℃
15 ~ 90% RH ( Noncondensing )
80~105 kPa
-20~60℃ (-4 ~ 140 F )
10 ~ 95% RH ( Noncondensing )
25