AMI Trides
Instrument Setup
4.4.
36
Correction of Trides sensor
If you measure concentrations > 0.1 ppm chlorine or 0.01 ppm
ozone correct the instrument after 24 h running-in. A zero point cal-
ibration is not necessary.
The standard method to correct the AMI Trides sensor is the DPD
photometric method. Use a high quality photometer (e.g. Swan
Chematest), to determine the reference value. Perform 3 manual
measurements and calculate the average value.
The manual sample must be taken from the grab sample outlet of
the flow cell.
If you measure concentrations < 0.1 ppm chlorine or 0.01 ppm
ozone, let the instrument run continuously for at least 5 days in nor-
mal operation before performing a zero point calibration!
A-96.250.111 / 211019
Need help?
Do you have a question about the AMI Trides and is the answer not in the manual?