How can I change the shock protocol on my LIFEPAK CR-T training system to
display more scenarios than "no shock advised"?

Explanation:

The LIFEPAK CR-T training system can be programmed to follow one of four preset training scenarios or one of two custom set scenarios
you can configure.
The four preset training scenarios are:

Steps to Resolve:

To change the shock protocol on the trainer:
1. Turn on the CR-T Training System using the Master Power switch. It is located on the right side when facing the front of the trainer. The
switch is next to this symbol:
2. Press the GREEN SELECT button on the remote control. The LED on the back of the trainer will flash green.
3. Select the shock protocol you wish to use by pressing the corresponding button on the remote: F1, F2, F3, F4, C1, or C2. The LED on
the back of the trainer will display the shock protocol selected. Green LED indicates Shock Advised and red LED indicates No Shock
Advised.
4. Press the GREEN SELECT button on the remote again to confirm and save the shock protocol. The signal sensor flashes, indicating the
selection is saved, and the Trainer enters a simulated OFF status.
5. The trainer will now follow the shock protocol you selected.
TRAINING SYSTEM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LIFEPAK CR-T and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Physio Control LIFEPAK CR-T

  • Page 1 “no shock advised”? Explanation: The LIFEPAK CR-T training system can be programmed to follow one of four preset training scenarios or one of two custom set scenarios you can configure. The four preset training scenarios are:...
  • Page 2 References: Please refer to “Configuring Shock Protocols” in the set-up procedures of the Operating Instructions for your LIFEPAK CR-T training system. That section includes instructions to program a custom shock protocol setting for Custom #1 (C1) or Custom #2 (C2).

Table of Contents