SETUP MENU contd.
SlEEP MODE
When in sleep mode, the meter
automatically shuts down after
20 minutes if no buttons have
been pressed during that time.
is displayed if sleep mode
is enabled.
1. While in Setup Menu, press
SLP is displayed.
2. Press enTer until ON or OFF is displayed.
3. Press
or
until desired setting is
displayed.
4. Press enTer to store the setting in memory.
Note: If sleep mode is not enabled, the meter
does not automatically shut OFF, and
will not show on the display.
SPECIFICATIONS
General
Battery
Operating Temperature
Storage Temperature
Temperature range
K-type Thermocouples
IR Temperature

display resolution

K
SeTUP
or
until
9V
32°F to 122°F
(0°C to 50°C)
Noncondensing
14°F to 122°F
-10°C to 50°C
-328°F to 2501°F
-200°C to 1372°C
-22°F to 1022°F
-30°C to 550°C
0.1(°C•°F•K)<1000
1(°C•°F•K)>1000
MAINTENANCE

rePLaCInG THe BaTTerY

The thermometer must be OFF when changing
the battery.
1. Press
to power OFF the thermometer.
2. On the back side of the thermometer, locate
the
symbol.
3. While holding the thermometer, place your
thumb on the
push downward and pull backward; the battery
cover will begin to open. Use your other hand
to grasp the cover and pull downward. Lift the
cover up and off.
4. Lift out the 9-volt battery and disconnect it from
the harness.
5. Connect the new 9-volt battery to the harness,
and place it into the battery cavity.
6. Install the battery cover by placing it over the
battery and pushing upward until it snaps into
place.
7. Dispose of the old battery according to
local, state, and federal regulations.
Temperature accuracy
T1, T2, T3, T4
Above -148°F (-100°C)
T1, T2, T3, T4
Below -148°F (-100°C)
T1-T2, T3-T4
Ir
IR above 14°F (-10°C)
IR below 14°F (-10°C)
T1-Ir & T2-Ir
IR above 14°F (-10°C)
IR below 14°F (-10°C)
T3-Ir & T4-Ir
IR above 14°F (-10°C)
IR below 14°F (-10°C)
. Using gentle pressure,
± [0.15% rdg + 1.8°F (1°C)]
± [0.5% rdg + 3.6°F (2°C)]
± [0.5% rdg + 1.8°F (1°C)]
± [2.0% rdg + 3.6°F (2°C)]
± 9°F (5°C)
± [2.0% rdg + 5.4°F (3°C)]
± 11°F (6°C)
± [2.0% rdg + 5.4°F (3°C)]
± 11°F (6°C)