Page 1
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 2
Signal Conditioning SC-2311 User Manual Connectivity for 5B and SSR Modules SC-2311 User Manual February 2001 Edition Part Number 322906A-01 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 4
Warranty The SC-2311 shielded carrier, SSR Series modules, and 5B Series modules are warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period.
Page 5
Conventions The following conventions appear in this manual: This icon denotes a note, which alerts you to important information. This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. bold Bold text denotes items that you must select or click on in the software, such as menu items and dialog box options.
Contents Chapter 1 Introduction About the SC-2311, 5B Series Modules, and SSR Series Modules ......1-2 What You Need to Get Started ..................1-3 Unpacking ........................1-4 Chapter 2 Configuring and Setting Up the SC-2311 Configuring the SC-2311....................2-1 Setting SW5.....................2-4 Setting SW4.....................2-4 Setting SW3.....................2-4 Setting SW2.....................2-5...
Page 7
SSR-OAC-5A, 240 VAC ................4-5 SSR-ODC-5, 3–60 DVC ................. 4-5 Appendix A Specifications Appendix B 5B Series Module Data Sheets Appendix C Power Requirements for the SC-2311 with 5B and SSR Series Modules Appendix D Technical Support Resources Glossary Index SC-2311 User Manual ni.com...
Page 8
Contents Figures Figure 2-1. SC-2311 Top Cover Removal ...............2-2 Figure 2-2. SC-2311 Parts Locator Diagram ............2-3 Figure 2-3. NRSE, and RSE or DIFF Mode Circuits..........2-4 Figure 2-4. Installing an I/O Panelette ..............2-8 Figure 2-5. Installing an I/O Panelette Label ............2-9 Figure 2-6.
SC-2311 shielded carrier, 5B Series modules, SSR Series modules, and accessory terminal blocks. It also describes how to configure, connect, and customize the SC-2311 for use with National Instruments E Series data acquisition (DAQ) devices. The SC-2311 shielded carrier interfaces with the following signal conditioning modules: •...
The SSR Series modules are high-performance, fully encapsulated devices that you can insert into any of the eight available slots on the SC-2311. The SC-2311 also has light emitting diodes (LEDs) that indicate the status of each SSR Series module.
Chapter 1 Introduction What You Need to Get Started To set up and use the SC-2311 you need the following items: ❑ SC-2311 shielded carrier ❑ Up to eight of the following 5B Series modules for analog signal conditioning –...
One or more I/O panelettes and label sheet Unpacking Your SC-2311 shielded carrier is shipped in a cardboard box. The 5B and SSR Series modules are shipped in antistatic packaging to prevent electrostatic damage to the modules. Electrostatic discharge can damage several components on these products.
• • • You must configure these slide switches prior to operating the SC-2311. To access the slide switches, remove the eight top cover screws and the top cover as shown in Figure 2-1. See Figure 2-2 for switch locations.
Chapter 2 Configuring and Setting Up the SC-2311 1 Top Cover Screws 2 Top Cover 3 SC-2311 Figure 2-1. SC-2311 Top Cover Removal SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Chapter 2 Configuring and Setting Up the SC-2311 Setting SW5 Setting SW5 to the AIGND/AISENSE SHORT position allows the E Series DAQ device to operate in nonreferenced single-ended (NRSE) mode. Use this switch setting when you are using 5B Series modules to take measurements.
When using the Board +5V (5B) setting, it is necessary to know the current requirements of the 5B Series modules and the current available from the DAQ device you are using. See Appendix C, Power Requirements for the SC-2311 with 5B and SSR Series Modules, for more details.
Page 18
Connections, including power signals to ground and vice versa, that exceed any of the maximum signal ratings on the SC-2311 device can create a shock or fire hazard, or can damage any or all of the boards connected to the SC-2311, the host computer, and the SC-2311 device.
Select configurable I/O panelettes appropriate for the I/O connectivity needs of your application. The I/O panelettes have standard prewired I/O connectors. You can install I/O panelettes on the front of the SC-2311 enclosure. You can install up to 18 single-width I/O panelettes in the SC-2311 enclosure.
Figure 2-4. Installing an I/O Panelette I/O Panelette Labels Each SC-2311 ships with a sheet of labels for you to apply to your I/O panelettes as shown in Figure 2-5. The label sheet has both preprinted and blank labels. You can customize the blank labels to suit your application.
Figure 2-6. Rack-Mount Option Refer to Figure 2-7 and use the following steps to install the stack-mount kit on your SC-2311 configurable connector enclosure: Attach the stack-mount brackets to both ends of the lower enclosure with the 4-40 × 1/4 in. flathead screws from the stack-mount kit.
Chapter 2 Configuring and Setting Up the SC-2311 × 1 4-40 1/4 in. Flathead Screws 2 Stack-Mount Brackets Figure 2-7. Stack-Mount Option You can stack additional enclosures by using additional stack-mount kits. You can remove the stack-mount kit handles, if necessary, by removing the two screws that attach the handle.
0–20 mA sources, RTDs, thermocouples, strain gauges, potentiometers, and frequency inputs. Install panelettes in the front of the SC-2311 to connect to your I/O signals. Connect the panelettes to the input terminal blocks of corresponding 5B channels. After the signals are conditioned, they are routed through a shielded 68-position cable to an E Series DAQ device for measurement.
20 Ω resistor that is included with the 5B32 module. These modules scale the input current to a 0–5 V output signal that is routed from the SC-2311 to the E Series DAQ device for measurement.
SC-2311 accessory, to compensate for the parasitic thermocouples formed at the input screw terminals of the SC-2311. The 5B37 scales the input signals from the thermocouples into a 0–5 V output signal that is routed from the SC-2311to the E Series DAQ device for measurement.
5B38 accepts input signals from full- and half-bridge 300 Ω to 10 kΩ transducers. The 5B38 also provides a +10 V excitation and produces an output signal of –5 to +5 V that is routed from the SC-2311 to the E Series DAQ device for measurement.
SC-2311 accessory, to compensate for the parasitic thermocouples formed at the input screw terminals of the SC-2311. The 5B47 linearly scales the thermocouples temperature into a 0–5 V signal that is routed form the SC-2311 to the E Series DAQ device for measurement.
– SSR-OAC-5A, 240 VAC – SSR-ODC-5, 3 to 60 VDC See Appendix C, Power Requirements for the SC-2311 with 5B and SSR Series Modules, to find the power requirements for your SSR Series modules. Input Modules Input SSR Series modules sense voltages from a field source and return a logic value to the E Series DAQ device.
Ground Figure 4-1. SSR-IAC AC Input Diagram You can insert and use the SSR-IAC-5 in any of the eight SSR sockets inside the SC-2311. SSR-IAC-5A, 240 VAC The SSR-IAC-5A input module senses 180–280 VAC/DC from a field source and converts it into a TTL logic signal. When a signal in this voltage range is detected at its input, the SSR generates a logic low signal on the corresponding DIO line of the E Series DAQ device.
Ground Figure 4-2. SSR-IDC DC Input Diagram You can insert and use the SSR-IDC-5 in any of the eight SSR sockets inside the SC-2311. Output Modules SSR Series output modules use DIO lines on an E Series DAQ device to control external devices.
You can insert and use the SSR-OAC-5 in any of the eight SSR sockets inside the SC-2311. SSR-OAC-5A, 240 VAC The SSR-OAC-5A output module uses a DIO line from the E Series DAQ device to switch an external 24–280 VAC load at up to 3.5 A.
– Figure 4-4. SSR-ODC DC Output Diagram You can insert and use the SSR-ODC-5 in any of the eight SSR sockets inside the SC-2311. SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 36
Appendix A Specifications for SC-2311 Environment Operating temperature ......0 to 55 °C Storage temperature ........–20 to 70 °C Relative humidity ........10 to 90% noncondensing Safety Designed in accordance with IEC61010-01, UL 3111-1, and CAN/CSA C22.2 No.1010.1 for electrical measuring and testing equipment.
Page 38
8.33 mS 5B Series For 5B Series modules specifications, see Appendix B, 5B Series Module Data Sheets. Maximum working voltage ....300 V Category II SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 40
The output common must be kept within ±3 V of power common. SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 44
The output common must be kept within ±3 Vdc of power common. SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 48
The output common must be kept within ±3 V of power common. SC-2311 User Manual B-10 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 52
The output common must be kept within ±3 V of power common. Specifications subject to change without notice. Copyright 1995-2000 Analog Devices, Inc. All rights reserved. SC-2311 User Manual B-14 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 56
±0.2˚C/% of V Power Supply Sensitivity, RTI Mechanical Dimensions 2.275" x 2.375" x 0.595" (57.8 mm x 59.1 mm x 15.1 mm) SC-2311 User Manual B-18 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 58
5B36 is mix-and-match and hot swappable with all other 5B Series modules, so can be inserted or removed from any socket in the same backplane without disrupting system power. SC-2311 User Manual B-20 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 62
5B36 is installed on a 5B Series backplane. Specifications subject to change without notice. Copyright 1995-2000 Analog Devices, Inc. All rights reserved. SC-2311 User Manual B-24 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 64
J, K, T, E, R, S, B, N, C Output Range 0 to +5 V Figure 2. 5B37 Input Field Connections (-5 V to +5 V - custom) SC-2311 User Manual B-26 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 66
The output common must be kept within ±3 V of power common. Specifications subject to change without notice. Copyright 1995-2000 Analog Devices, Inc. All rights reserved. SC-2311 User Manual B-28 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 72
The output common must be kept within 3 Vdc of power common. SC-2311 User Manual B-34 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 78
-3 dB @ 100 Hz (90 dB per decade > 100 Hz) Input, Excitation, and Sense Resistor Protection Continuous 240 V rms, maximum SC-2311 User Manual B-40 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 80
The output common must be held within ±3 V of power common. SC-2311 User Manual B-42 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 82
< 5 V peak-peak Large Signal, V > 5 V peak-peak 40 k || 330 pF Power Off 40 k || 330 pF SC-2311 User Manual B-44 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 84
The Output Common must be kept within ±3 V of Power Common. Specifications subject to change without notice. Copyright 1995-2000 Analog Devices, Inc. All rights reserved. SC-2311 User Manual B-46 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 86
J, K, T, E, R, S, B Output Ranges 0 to +5 V Figure 2. 5B47 Input Field Connections (+1 V to +5 V - custom) SC-2311 User Manual B-48 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 88
The output common must be kept within ±3 V of power common. Specifications subject to change without notice. Copyright 1995-2000 Analog Devices, Inc. All rights reserved. SC-2311 User Manual B-50 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
SC-2311 with 5B and SSR Series Modules Use Tables C-1 and C-2 to determine how much power your SC-2311 consumes and whether or not the external power option is necessary. Most E Series DAQ devices can supply +5 V at 1 A with the exception of DAQCard E Series devices.
Appendix C Power Requirements for the SC-2311 with 5B and SSR Series Modules Table C-1. 5B Series Current Requirements (Continued) 5B Series Module (Each) Current Requirement at 5 VDC 5B45 110 mA 5B46 110 mA 5B47 30 mA Table C-2. SSR Series Current Requirements...
Page 91
The NI Developer Zone at is the essential resource for ni.com/zone building measurement and automation systems. At the NI Developer Zone, you can easily access the latest example programs, system configurators, tutorials, technical news, as well as a community of developers ready to share their own techniques.
Page 92
National Instruments corporate. Phone numbers for our worldwide offices are listed at the front of this manual. SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 98
Also called nonreferenced signal sources. Some common example of floating signal sources are batteries, transformers, or thermocouples. FREQ_OUT frequency output signal SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 100
Also called handshaked digital I/O. SC-2311 User Manual ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 101
NIDAQ32.LIB is a library that contains NI-DAQ functions. The NI-DAQ function set is broken down into object modules so that only the object modules that are relevant to your application are linked in, while those object modules that are not relevant are not linked.
Page 102
PWB used by the customer for signal connection passband the range of frequencies which a device can properly propagate or measure SC-2311 User Manual G-10 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 104
Also called a grounded measurement system. resistance temperature detector—a metallic probe that measures temperature based upon its coefficient of resistivity SC-2311 User Manual G-12 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 108
SC-2311 User Manual G-16 ni.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 109
B-34 to B-35 5B Series modules overview, 3-2 overview, 1-2, 3-1 to 3-2 specifications, B-36 to B-37 power requirements for SC-2311 shielded 5B42 process current input module with 20 VDC carrier with 5B Series modules, loop C-1 to C-2...
Page 110
Index conventions used in manual, iv customer education, D-1 parts locator diagram for SC-2311 shielded carrier, 2-3 physical specifications, SC-2311 shielded carrier, A-1 digital I/O specifications, SC-2311 shielded power requirements for SC-2311 shielded carrier, A-1 carrier with 5B and SSR series modules,...
Page 111
A-3 to A-4 B-27 to B-28 stack-mount kit, 2-10 to 2-11 5B38 strain-gauge input module, switch settings for SC-2311 shielded carrier B-31 to B-33 SW2, 2-5 5B42 process current input module with SW3, 2-4 to 2-5 20 VDC loop, B-40 to B-41...
Page 112
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Need help?
Do you have a question about the SC-2311 and is the answer not in the manual?
Questions and answers