Operating Procedure; Performance Specification - O-Two Equinox II User Manual

Nitrous oxide/ oxygen 50%/50% administration system
Table of Contents

Advertisement

1.4 PERFORMANCE SPECIFICATION

DEMAND VALVE
(Minimum) Peak Flow rate .......................................................... 140 L/min
Exhalation Resistance .....................................0 to 6 cmH
Inhalation Resistance................................0 to -6 cmH
Input Pressure ..........................................................................................50 PSI
Initiation Pressure .............................................................. -2 to -2.5 cmH
Operating Temp ....................................-18
Storage Temp ...............................-40
Input Connection ........................................................................... 5/8"-18 TPI
Patient Connector .............................................................................15/22 mm
Weight ....................................................................................... 0.21 Kg (7.4 oz)
Dimensions ..................................................Dia. 51mm x 81mm (2.0" x
3.2")
PRESSURE REGULATOR
Inlet Connection..........................CGA 965 N
Outlet Connector...........................................................................5/8" Thread
Operating Pressure...............................................................500 -3000 PSI
Output Pressure................................................................................50 -60 PSI
Weight.........................................................................................0.45 Kg (1.0 lb)

2. OPERATING PROCEDURE

2.1 DEMAND BREATHING
The O-Two Equinox® II Demand Valve is designed to operate on
compressed Nitrous Oxide/ Oxygen gas. It is intended to provide
50/50% Nitrous Oxide/Oxygen mixture to the patient at a safe
delivery pressure and adequate flow rate upon demand.
An inspiratory effort by the patient will activate the demand valve
and Nitrous Oxide/ Oxygen will flow to the patient at a rate and
volume in line with their inspiratory effort.
Equinox® II Administration System manual
2
C to 50
C (0
0
0
C to 60
C (-40
0
0
O/O
Pin Indexed Yoke
2
2
O @ 60 L/min
2
O @ 60 L/min
O
2
F to 122
F)
0
0
F to 140
F)
0
0
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents