Description - dacell DN-AM210 User Manual

Dynamic strain amplifier
Table of Contents

Advertisement

3-3. Description

Power lamp
1
This lamp shows you the power supply to Amplifier.
Zero Hand-operated control volume
2
It is a volume to control the offset of residual quantity after finishing the auto
zero control.
Control range is within  1V.
Zero auto control switch
3
Use this switch to carry out auto zero control.
When this auto switch is pressed, "Auto" will appear on the LED and then
zero will be set.
L.P.F switch
4
It is a switch is to control the frequency of the active filter.
The frequency of Cut-off can be controlled as in 10Hz, 100Hz, 1kHz, 10kHz,
and WB(Wide-band).
GAIN control volume
5
It is a volume to control GAIN and it can be continuously controlled as in 을
from X1 up to X11 times.
⑥ GAIN selecting switch
It is the switch to control GAIN and it can be controlled as in X1, X10, X100,
X400, X1000 times.
⑦ CAL Strain value switch
It is a shunt switch to calibrate the bridge in two different types - 200,and
200strain.
⑧ Power switch
This switch is to supply power to Amplifier.