Page 3
Before using your device, please make sure you have read all the precautions, warnings and safety information, followed by the guidance for proper setup. If in doubt of the suitability of the Parasym device, consult your doctor before use. ...
Parasym device is in close proximity while being used. Consult with your HCP before use because the device may cause changes in cardiac rhythm. Safety and efficacy of the Parasym device have not been evaluated in the following patients, and therefore it is not indicated for:...
Page 5
Instructions for Use may not be enough to fully explain the safe and effective use of the device. Ask your HCP or contact Parasym support if you have any questions about how to use the device or require any further clarification of these Instructions.
Page 6
POTENTIAL RISKS AND COMPLICATIONS The Parasym device is considered low risk when used in accordance with the instructions for use. A low incidence of skin irritation has been reported from using the Parasym device. No serious adverse events have been reported. If you experience any side effects from device use stop treatment immediately. If in doubt, contact your healthcare provider.
Page 9
Electrode Placement Connect the electrode to the tragus of the left ear, running the cable behind the neck over the right shoulder. ...
Page 11
Quick Start Guide 1. Insert the two AA alkaline batteries Take care in doing this, be careful not to strip the material covering the batteries as this can lead to a short circuit. 2. Plug the lead included into channel 1 3.
Page 12
Parasym Settings (a) Pulse Rate = 2 5 Hz (b) Pulse Width = 2 50 μs (c) Intensity = S ee section below (d) Recommended Time = 6 0m/day Mode: C onst ...
Page 13
(c) Finding your Intensity level After setup, repeat for each treatment session 1. Ensure electrode has a firm connection to the left tragus (See p.7 image). 2. Using the up arrow on the left of the keypad, for channel 1, slowly increase the intensity by 1 level every 2 seconds to the point where a ...
Page 14
Common Intensity Questions My stimulation intensity changes frequently throughout my treatment session, is this ok? It is common for sensory thresholds to change throughout sessions as well as between sessions. This is because the variables that affect ...
Page 15
Troubleshooting If the unit does not appear to be operating correctly, refer below to determine the cause. Should none of these measures correct the problem, please contact your supplier. The unit does not function at all Check the batteries are inserted correctly. ...
Page 16
Visit our Help Center Our Help Centre has answers to most common questions. You can access this at: ww.parasym.co/hc ...
Page 17
Reach out to us via our support messaging widget on our website www.parasym.co and we’ll get right back to you. You can also email us directly at h ello@parasym.co and this will also go through to our support team. ...
If there is a fault or if electrode is found to no longer work, please contact Parasym Support. ...
Page 19
shown in the image on page 8). Note: the safety and efficacy of the Parasym device has not been ...
Page 20
List of Accessory: ● 2 X 1.5 volt AA Alkaline ● 1 X earclip electrode (Model: GE-1) Parasym Specification Output: Maximum 36 mA (average value/Pulse Width100μS) across 500 Ohm load, RMS Value 6.08mA Pulse Width: From 50μS to 250μS adjustable...
15º. Vertically falling drops shall have no harmful effects when the enclosure is tilted at any angle up to 15’’ on either side of the vertical. European Representative Parasym Ltd 56 Shoreditch High Street London E16JJ England legal@parasym.co...