Z-800F Instructions for Use.
P/N 800F-IFU-2602, Rev. O
Z-800F Infusion Pump
INSTRUCTIONS FOR USE
Zyno Medical LLC
177 Pine Street
Natick, MA 01760
1-508-650-2008
Fax: 1-508-650-2006
Web: www.zynomed.com
1

Advertisement

Table of Contents
loading

Summary of Contents for Zyno Medical Z-800F

  • Page 1 Z-800F Infusion Pump INSTRUCTIONS FOR USE Zyno Medical LLC 177 Pine Street Natick, MA 01760 1-508-650-2008 Fax: 1-508-650-2006 Web: www.zynomed.com Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 2 WARNING: Only use Zyno Medical provided IV sets with the Z-800F Infusion Pump. There are risks associated with using any IV Sets other than Zyno Medical IV Sets with this device. Zyno’s warranty for its device will be null and void and Zyno will assume no responsibility for any incidents that may occur if the device is not utilized strictly in accordance with its product labeling.
  • Page 3: Table Of Contents

    SAVE INFUSION PARAMETERS AS PROTOCOL ............ 46   ALARMS AND TROUBLESHOOTING ................47   DEFINITIONS ......................... 47   ALARMS ......................... 48   WARNINGS ........................50   ERRORS .......................... 52   MAINTENANCE ........................ 53   Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 4 INSPECTION REQUIREMENTS .................. 76   SERVICE INFORMATION .................... 78   WARRANTY ........................79   APPENDIX .......................... 80   ALARM TESTING PROCEDURE ................. 80   APPROVED ADMINISTRATION SETS ..............84   ACCESSORIES LIST ..................... 86   Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 5: Introduction

    The Z-800F Infusion Pump must only be operated utilizing Zyno Medical’s proprietary administration sets. The sets are designed for use with the Z-800F Infusion Pump as well as for gravity-flow stand-alone use. For specific IV Administration Set instructions, refer to the Directions For Use provided with the set.
  • Page 6: Features

    INTRODUCTION FEATURES Flow Rates The Z-800F Infusion Pump flow rate range is from 1 to 1200 ml/h. Free Flow Protection The Z-800F Infusion Pump has a built in free flow clamp to prevent inadvertent free flow when the set is loaded in the pump.
  • Page 7: Symbols

    IPX1 Protection against fluid ingress: Drip Proof. Attention: Refer to accompanying documentation. Rx Only Federal (U.S.A) law restricts this device to sale by or on the order of a physician. Fuse: 2X2TA. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 8: Getting Started

    WARNING: Zyno Medical assumes no responsibility for incidents that may occur if its product is not used in accordance with its product labeling. WARNING: Only use Zyno Medical provided IV Sets with the Z-800F Infusion Pump. WARNING: The Z-800F Infusion Pump operation is strictly limited to trained operators whose competency in safe Z-800F Infusion Pump operation and safe IV therapy practices has been tested and proven.
  • Page 9 WARNING: Do not operate this device in the presence of Flammable Anesthetics mixture with Air, Oxygen or Nitrous Oxide. WARNING: Do not expose the Z-800F Infusion Pump to X-Rays, Gamma Rays or other Radiation, or to strong Electric or Magnetic Fields.
  • Page 10 VTBI, Secondary Rate, Secondary VTBI) with the prescription before starting infusion. CAUTION: Grounding reliability can only be achieved when equipment is connected to an equivalent receptacle marked "hospital only" or Hospital grade. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 11: Device Description

    GETTING STARTED DEVICE DESCRIPTION Pump Front View – Door Closed Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 12 GETTING STARTED DEVICE DESCRIPTION (Continued) Pump Front View - Door Open Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 13 GETTING STARTED DEVICE DESCRIPTION (Continued) Pump Back View Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 14: Installation Procedure

    3. Check to ensure the pump door operates freely. 4. Check for any loose parts. NOTE: If any of the following conditions are observed, the Z-800F Infusion Pump must be removed from use and inspected by qualified personnel:  Look for any signs of physical damage from shipping.
  • Page 15: Preparing An Infusion

    Powering On the System 1. Connect the Z-800F Infusion Pump Power Cord to an AC power source. 2. Press and hold the ON/OFF key on the Z-800F Infusion Pump for 2 seconds. 3. System self-test begins:  The diagnostics test causes all LED display segments and Status indicator lights to illuminate briefly.
  • Page 16: Loading Primary Administration Set

    Open the administration set package, remove the set and close Step 1 the roller clamp. WARNING: Make sure to close Roller Clamp before removing IV Administration Set from Z-800F Infusion Pump. Loading Primary Administration Set 1. Make sure the Anti-Free Flow Pinch Clamp is attached on the primary set and it’s in open position.
  • Page 17 12. Open the IV set roller clamp and confirm that there is NO FLOW in the IV set drip chamber. Connect the administration set to the patient’s injection site. LOADING IV SET INTO Z-800F PUMP WARNING: Make sure to close roller clamp before removing administration set from Z-800F pump.
  • Page 18: Infusion Mode Information

    3. Press the 10 Up data entry key twice to increase the 10 digit of Pri RATE to 20. Press the single digit Up data entry key five times to increase the single digit of Pri RATE to 5. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 19 Titrate Flow Rate or VTBI During a Primary Infusion From the PAUSE screen, press the PROG key; the pump will display primary infusion programming screen. The user is then able to modify the primary RATE and primary VTBI parameters. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 20 NOTE: The VINF field will record cumulative infusion volume infused unless cleared by the user pressing the CLR key or turning the pump off/on and starting a new infusion. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 21 8. Hang the secondary fluid container at least 8 inches above the primary solution container. WARNING: Make sure to close roller clamp before removing administration set from Z-800F pump. WARNING: The bottom of the secondary solution should be at least 19” above the top of the pump.
  • Page 22  Sec VTBI: Secondary VTBI 3. Keys: Allow user to access pump configuration and primary infusion programming screen.  CONFIG: Access pump configuration  PRI: Access primary infusion programming  Up/Down Arrows: Select parameter items Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 23 2. Press the 100 Up data entry key once to increase the 100 digit of Sec VTBI to 100. 3. Press 10 Up data entry key five times to increase the 10 digit of Sec VTBI to 50. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 24  Upon completion of secondary infusion, a switchover audio alert sounds 4 short beeps. The Main Display will switch-over to PRI (primary) State. WARNING: Do not over-program VTBI. Program the actual amount of the fluid in the IV bag. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 25 From PAUSE state, press the RUN/STOP key to resume infusion. Changing Rate or VTBI During a Secondary Infusion From the PAUSE screen, press PROG key; the Z-800F Infusion Pump will enter secondary infusion programming screen. The user is able to modify the secondary RATE and secondary VTBI parameters.
  • Page 26: Time/Vtbi Programming

    INFUSION MODE INFORMATION TIME/VTBI PROGRAMMING Access TIME/VTBI Programming Option The Z-800F Infusion Pump provides TIME/VOLUME programming options for Continuous Mode Infusion. The user may choose the infusion programming parameter option in the pump Infusion Mode selection screen. 1. Press Home key to access the main page. If a confirmation page showing “Current infusion...
  • Page 27 TIME/VTBI infusion parameters may be programmed by using the UP/DOWN arrow keys and Data Entry keys. WARNING: Do not over-program VTBI. Program the actual amount of the fluid in the IV bag. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 28: Infusion Complete

    Press the RUN/STOP key to acknowledge the infusion complete message. The pump will pause and the KVO infusion will be stopped. WARNING: Make sure to close roller clamp before removing administration set from Z-800F Infusion Pump. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 29: Tpn Mode Infusion

    3. In pump Infusion Mode menu, use the Up/Down Arrow keys to highlight 3. TPN MODE and press the SELECT key. 4. The pump will display the Programming screen of the TPN Mode infusion. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 30 The TPN Ramp Time parameters may be programmed by using the UP/DOWN Arrow keys and Data Entry keys. WARNING: Do not over-program VTBI. Program the actual amount of the fluid in the IV bag. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 31: 10-Step Mode Infusion

    Arrow keys to highlight .10 STEP MODE R/V or 10 STEP MODE T/V and press the SELECT key. 4. The pump will display the Programming screen of the 10 STEP Mode infusion. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 32 RUN/STOP to begin the infusion locks in the list. To edit the list after an infusion is underway, press RUN/STOP to pause, and then press the PROG key to edit the list. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 33 RUN/STOP to begin the infusion locks in the list. To edit the list after an infusion is underway, press RUN/STOP to pause, and then press the PROG key to edit the list. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 34: Intermittent Mode Infusion

    The sum of the “run” and the “idle” time is the cycle time, as illustrated below. Period Period Period IDLE IDLE IDLE CYCLE TIME Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 35 Entry keys. Note: During idle period, the pump runs at KVO rate. Volume infused during idle period is counted. The pump alarms “Infusion Completed” once VINF (volume infused) equals programmed Total VTBI. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 36 6 hours (Cycle Time). The run time for each period is 10 minutes (Period Time); idle time in each cycle is 5 hours 50 minutes. The parameters can be programmed as follows: Total VTBI: 200mL Cycle Time: 06:00 Period VTBI: 50mL Period Time: 10 minutes Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 37: Blood Infusion

    3. In pump Infusion Mode menu, use the Up/Down Arrow keys to highlight 6. BLOOD INFUSION and press the SELECT key. 4. The pump will display the Programming screen of the BLOOD INFUSION mode. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 38  CONFIG: Access pump configuration.  Up/Down Arrows: Highlight the selected parameter item. The RATE and VTBI parameters may be programmed by using the UP/DOWN Arrow keys and Data Entry keys. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 39: Managing Infusion

    The maximum prime volume is 10mL. Upon reaching the maximum prime volume, the prime will stop. WARNING: Patient must be disconnected before utilizing the PRIME key. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 40: Keypad Lockout

    To unlock the keypad, press the CLR key and immediately after that the 10’s DOWN data entry key. An audio prompt tone will sound and a “PANEL UNLOCKED” message will be displayed for 3 seconds. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 41: Clear Volume Infused

     Press the NO key to abort the action. 4. Upon user confirmation, the current VINF parameter will be reset to 0 ml. NOTE: The VINF field will record cumulative infusion volume infused unless cleared by pressing the CLR key. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 42: Powering Off

    MANAGING INFUSION POWERING OFF Stop the Z-800FInfusion Pump by pressing the RUN/STOP key during an infusion. Press and hold down the ON/OFF key for 2 seconds. The pump will turn off. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 43: Changing Primary Solution Container

    MANAGING INFUSION CHANGING PRIMARY SOLUTION CONTAINER 1. Stop the Z-800F Infusion Pump by pressing the RUN/STOP key during an infusion. 2. Close the Roller Clamp. 3. Remove the empty solution container. 4. Spike the new container. 5. Go to the primary infusion programming screen.
  • Page 44: Changing And Reloading Iv Set During Infusion

    MANAGING INFUSION CHANGING AND RELOADING IV SET DURING INFUSION 1. Stop the Z-800F Infusion Pump by pressing the RUN/STOP key during an infusion. 2. Close the Roller Clamp. 3. Disconnect the IV Administration Set from the patient. Step 5 4. Open the Pump Door.
  • Page 45: Protocol Mode Infusion

    Start Infusion Verify the displayed infusion parameter entries (Primary Rate, Primary VTBI, Secondary Rate, and Secondary VTBI). If a clamp is engaged, remove the clamp. Press RUN/STOP key to start the infusion. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 46: Save Infusion Parameters As Protocol

    4. Press SAVE key to associate the current programmed infusion parameters with the protocol name as entered. NOTE: Loading factory default settings will delete all saved protocols. (Refer to Maintenance->Configurable settings->Load default on how to load factory default). Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 47: Alarms And Troubleshooting

    ALARMS AND TROUBLESHOOTING ALARMS AND TROUBLESHOOTING To enhance safety and ease of operation, the Z-800F Infusion Pump provides a full range of audio and visual alarms, warnings, and prompts. DEFINITIONS Error An audio and visual signal indicates a failure has been detected.
  • Page 48: Alarms

    3. Verify that the fluid then alarms). container is not empty. Press the RUN/STOP key again to resume infusion. Battery The Z-800F pump is Continuously flashes (blue 1. Plug the power cord into an Empty operating on battery or red) notification until AC power outlet.
  • Page 49 2. The infusion is paused. pump. 3. Load the IV set based The infusion is clamp correctly. Press the stopped. RUN/STOP key again to resume infusion. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 50: Warnings

    ALARMS AND TROUBLESHOOTING WARNINGS Warning Meaning Secondary Alarm/ Response Warning Light (optional) Z-800F Infusion Pump is Continuously flashes Plug the power cord Battery operating on battery power (blue or red) into an AC power and low battery condition is notification until user outlet.
  • Page 51 The pump red) notification until Press the RUN/STOP automatically switches user key. to KVO mode. acknowledgement. The Z-800F Pump will be The visual warning paused. message will appear and the audio alarm will sound constantly until user acknowledgement. Pump will...
  • Page 52: Errors

    Infusion acknowledgement. stop using the affected stops. pump. The pump must be taken out of service until serviced by qualified service personnel. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 53: Maintenance

    Test fluid is distilled water at room temperature (21+/-1C) at ambient air pressure;  Head height is set according to instructions in the IFU;  Test set is Zyno proprietary set at specifications;  Programmed volume during test is no less than 10 mL. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 54 Temperature: +10°C to +40 °C (+59 °F to + 113 °F) Humidity: 30 % to 75% R.H. at +40 °C, non-condensing Air pressure: 70kPa to 106kPa IV Administration Set: Use only Zyno Medical approved Administration IV Sets. Alarms: Air-in-Line Occlusion...
  • Page 55: Configurable Settings

    Press the QUIT key again to exit the configuration menu. WARNING: The factory default settings should be used unless qualified clinical personnel determine that other customized settings are appropriate and safe. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 56 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Adjusting the Air-In-Line Alarm Threshold The Z-800F’s Air-In-Line sensor detects an air bubble in the administration set tubing. Air-In-Line alarm threshold can be adjusted in the configuration menu. Alarm threshold can be adjusted to air bubble size from a volume of 2 micro-liters to 280 micro-liters.
  • Page 57 Press the QUIT key again to exit the CONFIGURATION menu. WARNING: The factory default settings should be used unless qualified clinical personnel determine that other customized settings are appropriate and safe. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 58 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Enable/Disable NEAR END WARNING The Z-800F Infusion Pump has a feature to detect that the current infusion is about to end in 3 minutes. The user may enable or disable the near end warning in the configuration menu. When enabled, the audio and visual warning message will appear every 5 seconds until infusion completion.
  • Page 59 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Enable/Disable Drip Sensor Option The Z-800F pump has an optional external drip sensor, which may be used to detect a fluid container empty event. The user may enable or disable the drip sensor in the configuration menu. To enable or disable this option, refer to “Customize the BIOMED OPT Menu”...
  • Page 60 When in manual continuous combined infusion, changing HIDE SEC to YES results in the infusion parameters being zeroed out. When in manual non- continuous infusion, changing HIDE SEC has no effect on selected infusion. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 61 Press the QUIT key again to exit the CONFIGURATION menu. WARNING: The factory default settings should be used unless qualified clinical personnel determine that other customized settings are appropriate and safe. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 62 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Adjusting the LCD Brightness The Z-800F Main LCD Display’s brightness level can be adjusted by the user. 1. From the configuration menu, Select 2 . FACTORY SET 2. In the FACTORY SET menu, highlight the BRIGHTNESS option by pressing the Up/Down Arrow.
  • Page 63 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Adjusting the Contrast The Z-800F Main LCD Display’s contrast level can be adjusted by the user. 1. From the configuration menu, Select 2. FACTORY SET 2. In the FACTORY SET menu, highlight the CONTRAST option by pressing the Up/Down Arrow.
  • Page 64 MAINTENANCE CONFIGURABLE SETTINGS (Continued) Adjusting the Alarm Volume The Z-800F Infusion Pump is able to generate multiple audio alert tones to indicate a pump alarm, warning, error or prompt. Audio alarm volume may be adjusted by the user to Low, Mid or High setting.
  • Page 65 After the user has finished the customization, he or she can choose “SELECT” to save the changes and return to the configuration menu. Pressing “QUIT” will return to the configuration menu without saving the changes. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 66 After the user has finished the customization, he or she can choose “SELECT” to save the changes and return to the infusion mode menu. Pressing “QUIT” will return to the infusion mode menu without saving the changes. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 67 “SELECT” to save the changes and return to the “ LOAD PROTOCOLS ” main page. Pressing “QUIT” will return to the “ LOAD PROTOCOLS ” main page without saving the changes. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 68 After the user has finished the customization, he or she can choose “SELECT” to save the changes and return to the “BIOMED OPT” main page. Pressing “QUIT” will return to the “BIOMED OPT” main page without saving the changes. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 69 After the user has finished the customization, he or she can choose “SELECT” to save the changes and return to the “FACTORY SET” main page. Pressing “QUIT” will return to the “FACTORY SET” main page without saving the changes. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 70 DEFAULT option and press “SELECT”. A confirmation page will be displayed: “Confirm Restoration Of Factory Default Parameters?” Pressing “YES” would reset the pump to factory default parameters. Press “NO” to cancel the action. Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 71: Storage And Transportation

    The AC indicator light pictured left will be on whenever the Z-800F Infusion Pump is plugged in. When transporting the pump, please apply sufficient protection to the pump to prevent physical damages. It is recommended to use a box size of 10x10x12 for an individual pump.
  • Page 72: Battery Care And Maintenance

    When the pump runs on battery power, the POWER indicator is off, and the BATTERY indicator is on. The Z-800F Pump is equipped with a standard configuration of 9.6 volt 4500mAh Nickel-Metal-Hydride battery. The battery charges whenever the pump is plugged into an AC outlet.
  • Page 73 This will be observed by user as low apparent run times and premature low battery warning and alarms. The lowered capacity is not permanent, but may require 2-3 full discharge/charge cycles to recover. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 74 The battery capacity should be checked at least once every 12 months. CAUTION: Battery replacement should be If the Z-800F pump is to be stored at temperatures in performed by qualified excess of 86 °F (30 °C) for one or more months, the battery service personnel while the should be removed and placed in an environment between instrument is not in use.
  • Page 75: Cleaning

    2. Use soft cloth dampened with warm water and a mild nonabrasive cleaning solution to clean all exposed surfaces. Do not spray any fluids directly on the instrument. For sanitizing or antibacterial treatment, use 10% bleach solution and water. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 76: Inspection Requirements

    Start-Up Sequence section of this in improper instrument instruction for use before each usage of the instrument. operation. Regular inspections are not covered under the contract or agreement offered by Zyno Medical and must be performed by the user. REGULAR INSPECTIONS PROCEDURE FREQUENCY...
  • Page 77 Medical for the performance of all required maintenance. WARNING: The manufacturer cannot assure the accuracy and/or safety of the pump if it is not regularly maintained as recommended in the Z-800F Instructions for Use Manual. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 78: Service Information

    MAINTENANCE SERVICE INFORMATION If a Z-800F Pump fails to respond as described in this Instruction for Use and the cause cannot be determined, do not use the affected instrument. Contact qualified service personnel. Within the United States, application and service information may be obtained by writing to Zyno Medical LLC.
  • Page 79: Warranty

    Zyno Medical LLC. (hereinafter referred to as “Zyno Medical”) warrants that: A. Each new Zyno Medical Z-800F infusion pump is free from defects in material and workmanship under normal use and service for a period of one (1) year from the date of delivery by Zyno Medical to the original purchaser.
  • Page 80: Appendix

    Main Display will show “AIR-IN-LINE”. The infusion will stop. 5. Press the RUN/STOP key again to acknowledge the alarm condition. The Z-800F Pump will be in PAUSE state. The “PAUSE” message will appear on the display screen.
  • Page 81 Display will show alarm screen with “Occlusion”. The infusion will stop. 5. Press the RUN/STOP key to acknowledge the alarm condition. The Z-800F Pump will be in PAUSE state. The “PAUSE” message will appear on the screen. Note: At slower infusion rates, the occlusion alarm 6.
  • Page 82 6. Install the IV set based clamp in pump. Press the RUN/STOP key again to resume infusion. Battery Test 1. Connect the Z-800F pump to an approved AC power outlet for at least 18 hours to allow the battery to fully charge.
  • Page 83  Time interval between the Battery Empty Alarm and the pump shut down is less than 2 minutes.  Pump configuration setting changed after battery depletion.  Any of the current infusion parameters is not preserved. Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 84: Approved Administration Sets

    WARNING: Use only administration sets labeled as Zyno Medical with the Z-800F Infusion Pump System. The use of any other set for use with Z-800F system may cause improper instrument operation, resulting in inaccurate fluid delivery or other potential hazards.
  • Page 85 Check Valve, Two Needleless Injection Sites, one at the proximal side, one at the distal side, Pinch clamp, Roller Clamp, 0.22 micron IVEX filter, Slide clamp, Male Luer Lock Adaptor, Luer Lock Tip protector Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 86: Accessories List

    Drip Chamber-Based Flow Sensor Accessory Secondary Red Secondary Red Alarm/Warning Optional 6612B Alarm/Warning Light Light Visual Indicator (magnetic) Accessory Secondary Blue Secondary Blue Alarm/Warning Optional 6638A Alarm/Warning Light Light Visual Indicator (magnetic) Accessory Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...
  • Page 87 This page is intentionally left blank Z-800F Instructions for Use. P/N 800F-IFU-2602, Rev. O...
  • Page 88 Zyno Medical LLC. 177 Pine Street Natick, MA 01760 Phone: 1-508-650-2008 Fax: 1-508-650-2006 Z-800F Instructions for Use P/N 800F-IFU-2602, Rev. O...