Device Serial Number Label

Each ActiGraph device is equipped with a label stuck to it with pertinent device information inside the
serial number. For example the serial number MRA3A29120016 means:
MRA3 = device model
A = device model version
29 = week of the year it was manufactured (29
12 = year it was manufactured (2012)
0016 = number of device manufactured that week

Cleaning the Device

ActiGraph devices are approved to be cleaned with a non-abrasive cloth and isopropyl alcohol. The USB
cap is to be inserted into the device and tightened prior to cleaning so no liquid can enter into the
device and damage any of the internal electronics. For optimal results, apply isopropyl alcohol to the
cloth, clean the device for 15-20 seconds, and let air dry. ActiGraph devices are not to be sterilized
because damage may occur. Though ActiGraph devices are not required to be cleaned on any
predetermined schedule, it is advised that the devices be cleaned after each use by a user.

Device Storage and Life Expectancy

There is no specific requirement for storing ActiGraph devices but a dry location with temperature
limitations between 0°C and 50°C is suggested. For optimal battery life expectancy, routinely charge the
device according to Table 4.
The only limiting factor of an ActiGraph device when calculating life expectancy is the battery. The
battery has a life expectancy of five (5) years but can be replaced by ActiGraph personnel thus extending
the life of the device. The performance of the components inside the device does not degrade over
time so the life expectancy of a device is immeasurable.

Device Maintenance

ActiGraph's devices do not require routine maintenance or calibration. Calibration is done during the
manufacturing of the device and will last for the life of the device. If maintenance is required on a
device, or a piece of a device has gone missing, call ActiGraph Customer Support or visit the support
section on the ActiGraph website
Any critical components that require maintenance are to be performed by trained ActiGraph personnel
only. All necessary training manuals, schematics, parts listings, and other necessary documentation, will
be made available to the trained ActiGraph service technicians.

Accessories

Accessories sold at ActiGraph are constantly changing due to the ever changing features available with
the wGT3X-BT device. For a full list of current accessories available for the wGT3X-BT device, please visit
ActiGraph's website
www.actigraphcorp.com/product-category/accessories
49 E. Chase St.
www.actigraphcorp.com
Pensacola, FL 32502
th
week)
and initiate an RMA for your device.
. Use the accessories
tel 850.332-7900
www.theactigraph.com
P a g e
| 20
fax 850-332-7904

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for ActiGraph wGT3X-BT

Table of Contents