Troubleshooting Error Codes; Product Specifications - HoMedics THERA-P TE-120 Manual

Deluxe one-second ear thermometer
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING/ERROR CODES
Condition
Possible Cause
Consistent low
Temperature
temperature
reading is below
readings ("Lo")
measurement range
(below 50˚F/10°C).
The probe is not
positioned properly
in ear.
The waterproof probe
is dirty.
Consistent high
Temperature
temperature
reading is above
readings ("Hi")
measurement range
(above 122˚F/50°C).
Consistent error
Operating temperature
messages ("Err")
is not within the
permissible range for
measurement.
Low battery warning
Battery power
is low.
Low battery warning
Battery power is too
with "Lo" symbol
low for the
thermometer to
function.
Corrective Action
Product Design-Measurement
range is 50˚– 122˚F (10° – 50°C).
Contact your physician if you
have or suspect any medical
problems.
Make sure probe is properly
positioned against the opening
of the ear canal
Clean the probe with a
cotton swab moistened with
alcohol.
Product Design-
Measurement range is
50˚– 122˚F (10° – 50°C). Contact
your physician if you have or
suspect any medical problems.
Product Design-
Operating temperature is
60.8˚– 95˚F (16°– 35°C).
Replace battery (see "Inserting
and Replacing Batteries").
Replace battery
immediately (see "Inserting and
Replacing Batteries").
18

PRODUCT SPECIFICATIONS

Model Number
Power Source
Battery Life
Sensor
Temperature Indication
Accuracy
Measurement Range
Weight
Operating Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Contents
NOTE: These specifications are subject to change without notice. This ear
thermometer meets requirements established in ASTM Standard E 1965-98.
Model Number TE-120
DC 3 Volt (lithium battery CR2032)
1,700 Measurements in day mode or
700 Measurements in night mode.
Thermopile
In 4 digits, every 0.1˚F or ˚C
+/- 0.4˚F/0.2°C between 96.8˚– 102.2˚F
(36.0° – 39.0°C)
+/- 0.5˚F/0.3°C less than 96.8˚F (36.0°C) or
greater than 102.2˚F (39.0°C)
50˚– 122˚F (10˚– 50˚C)
0.12 lbs (with battery)
60.8˚– 95˚F (16˚– 35˚C)
Relative humidity up to 95%
(non-condensing)
-4˚– 122˚F (-20˚– 50˚C)
Relative humidity up to 95%
(non-condensing)
Ear Thermometer & Battery, Storage Case,