4.0

Specifications

Sample Flow Rate
Calibration Gas Requirement
Operating Pressure Drop At 10 l/m
Maximum Gas Temperature At Inlet
Filter Oven Temperature
Dimensions
Weight
Input Power Requirement
Input Voltage Requirement
Blowback Tank Volume
Blowback Duration
Blowback Timer Period
Blowback Solenoid Valve Voltage
Maximum NH
Concentration
3
Enclosure
Sample Probe
Ammonia Converter
Sample Line Connection
Filter Heater Type
Oven Insulation Material
Filter Chamber Material
Filter Element Types
Ammonia Converter Heater Type
Converter Canister
270NH3.MAN.REVD.062019
OPERATING SPECIFICATIONS
0 to 10 l/m
Sample flow rate plus 10%
12" water column
700°F (370°C)
350°F (176°C)
18" H x 30" W x 10" D
50lbs (Plus Probe)
1600 watts
115/230VAC, 50/60 Hz
0.7scf when at 100psig
2 seconds suggested
Controlled externally
As specified
20 ppm by volume
MATERIAL SPECIFICATIONS
Fiberglass or 304SS (NEMA 4X)
Painted steel (NEMA 4X)
3" Heat shrink boot
Rod heaters in aluminum tube, controlled with integral
thermal switch
Jacketed fiberglass
316SS
2 µm ceramic or 316SS filter elements; ceramic element
with 0.1 µm coating inside
Ceramic fiber heater (controlled remotely, sensed with Type
K Thermocouple)
316 Stainless Steel