5 Memory / 6 Miscellaneous
5.3
Deleting memory
5.4
Switching off
the unit

6 Miscellaneous

6.1
Guide to
Display Symbols
6.2
Troubleshooting
24
The data can be deleted from the memory selected by pressing and holding the
MEMORY button
for about 7 seconds. After 7 seconds, the display
shows "
" ( CLEAR ) which indicates that all data have been deleted.
The unit switches off automatically after about 1 minute, if no button is
pressed during this period. It can also be switched off by pressing the START
button
. The time and the personal memory selected are constantly shown
in the display.
Err
Error during measurement
Lo
Pulse is lower than 40 beats/minute
Determining pulse
Unit has detected an arrhythmia
Batteries low. Please replace batteries.
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse could be
determined.
Check the connections between the cuff and the main unit.
Err 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
If the same error occurs repeatedly with further readings, please contact
MEDISANA.