Summary of Contents for SynCardia Companion 2 Driver System
Page 1
CAUTION: In the United States, the use of the SynCardia 70cc TAH-t for destination therapy is investigational. CAUTION: In the United States, the use of the Companion 2 Driver System with the 70cc TAH-t for the destination therapy indication or with the 50cc TAH-t is investigational.
Page 2
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Chapter 15. Switching from Primary Companion 2 Driver to Backup Companion 2 Driver ................97 Chapter 16. Switching from CSS Console to Companion 2 Driver System ..99 Chapter 17. Switching from Companion 2 Driver System to CSS Console ..103 Chapter 18.
Page 4
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Page 5
Table of Figures Figure 1-1 – 50cc or 70 cc TAH-t with Companion 2 Driver System ......9 Figure 1-2 – 70cc TAH-t and 50cc TAH-t ..............10 Figure 1-3 – Companion 2 Driver System Major Components ........ 10 Figure 1-4 – Companion 2 Driver ................11 Figure 1-5 –...
Page 6
Figure 20-3 – Driver Side Panel with Filter Cover Removed ........115 Figure 20-4 – SETUP Menu ..................116 Figure 20-5 – Verification Screen ................116 Figure 20-6 – SETUP Menu with Filter Maintenance Alarm Cleared ..... 117 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 6 of 132...
Page 7
Table 8-1 – O.R. Mode Parameter Adjustment Range and Increments ....45 Table 9-1 – I.C.U. Mode Parameter Adjustment Range and Increments ....60 Table 11-1 –Symbols Used in Companion 2 Driver System Labeling ..... 67 Table 12-1 – List of Locations and Sizes of Fuses ........... 69 Table 14-1 –...
Page 8
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 8 of 132...
For more information on the TAH-t, see the SynCardia temporary Total Artificial Heart (TAH-t) Instructions for Use with the Companion 2 Driver System, for the 50cc or 70cc TAH-t, as appropriate.
The Companion 2 Driver System (Figure 1-3) is a multi-component electro- mechanical unit designed to provide pneumatic power to the implanted 50cc or 70cc TAH-t. The Companion 2 Driver System operates and monitors the 50cc or 70cc TAH-t throughout the implantation, surgical recovery phase and ambulatory phase of patient support.
Cart with the External Batteries facing outward. o To remove the Driver from the Hospital Cart, simply lift the Driver straight up out of the Hospital Cart. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 11 of 132...
Release Button found at the back of the Caddy. Once the Driver Release Button is depressed, the Driver can be removed from the Caddy by lifting straight up. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 12 of 132...
Driver is not immediately available. The Companion Drivelines connect to the CPC connectors on the Hand Pump. Clinical operators should pump at 110 strokes per minute to provide adequate Cardiac Output. See the SynCardia Hand Pump Operator Manual for additional details. Hand...
Page 14
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Companion 2 Driver System Intended Use The Companion 2 Driver System, used with the 50cc TAH-t or the 70cc TAH-t, is intended for use inside the hospital and on the hospital grounds.
Page 16
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 16 of 132...
• Patients who do not have sufficient space in the chest area vacated by the natural ventricles. • Patients who have a body surface area (BSA) of > 1.85 m • Patients who cannot be adequately anticoagulated on the 50cc TAH-t. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 17 of 132...
Page 18
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Prior to use, refer to the SynCardia temporary Total Artificial Heart (TAH-t) Instructions for Use with the Companion 2 Driver System for the 50cc or 70cc TAH-t. The Drivelines are sterile components and are intended for single use only.
Page 20
This is performed through driver touch screen calibration procedure (see Section 8.1.10). 4.14 Always keep the Companion 2 Driver System away from any ignition source including open flame and smoldering cigarettes. 4.15 If a docked Driver user interface displays the following in the Power Status...
Surgical, nursing, and perfusion staff responsible for the SynCardia TAH-t program at each hospital must complete the SynCardia Certification program. The SynCardia TAH-t and Drivelines are provided sterile; caution must be taken in opening the package. Do NOT resterilize. Do NOT use if package is damaged.
Page 22
5.16 To facilitate Hospital Cart movement over an obstacle that is 20 mm in height (approximately 0.8”) it is recommended that the user (Clinical Engineer/Nurse) and an assistant, if needed, pull upwards on the Hospital Cart Support Handles. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 22 of 132...
The power management electronics assembly is designed to continue to route power to the motor assemblies in the event of a loss of one power channel. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 23 of 132...
Page 24
Driver is placed in the Hospital Cart or Driver Caddy. There is a port on the right side of the Driver for connecting to External air. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 24 of 132...
Chapter 14, Alarms and Notifications, contains additional information. The Graphical Display Window shows pressure and flow waveforms on a beat-to-beat basis and the cardiac output trend SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 25 of 132...
Page 26
– I.C.U. Environment, Operating Modes – Ambulatory Mode, respectively. The Air Icon indicates if the Driver is connected to external air or is powered by the compressors. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 26 of 132...
Solid green indicates the Emergency Battery is not in use. A flashing green emergency LED indicates the Emergency Battery is being charged. Flashing red indicates a very high alarm state. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 27 of 132...
Page 28
Driver and DO NOT attempt to dock the Driver again into the same Hospital Cart or Caddy. A malfunction of the Hospital Cart or Caddy may exist and it must be returned to SynCardia for evaluation. The Driver should be docked into another Hospital Cart or Caddy.
(counter will take approximately one minute to update the estimated battery support time based on Driver settings. Battery Release Latch Battery Fuel Gauge Battery Test Button Figure 6-3 – Driver External Battery SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 29 of 132...
Page 30
As with all rechargeable batteries, the full capacity of the External Batteries and Emergency Battery will decrease as a result of battery age and charge/discharge cycling. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 30 of 132...
Driver functionality, however, the internal compressor may activate intermittently. When the compressor activates, the external air icon on the display will not be illuminated. Figure 7-1 – External Air Connection SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 31 of 132...
Page 32
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Driver software has loaded, the system will show the Ambulatory Mode Screen (Figure 8-1). CAUTION Do not rapidly rotate the key between the ON setting and OFF setting. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 33 of 132...
Page 34
Driver and DO NOT attempt to dock the Driver again into the same Hospital Cart or Caddy. A malfunction of the Hospital Cart or Caddy may exist and it must be returned to SynCardia for evaluation. The Driver should be docked into another Hospital Cart or Caddy.
CONFIRM to change to O.R. Mode. This can be verified by the Mode Change temporary confirmation window displayed for two seconds and the O.R. Mode descriptor in the upper right corner of the display (Figure 8-2). SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 35 of 132...
A reminder audible signal will be generated every two hours to indicate that the system is in O.R. Mode. All visual alarm indicators will continue to function SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 36 of 132...
Driver. The calibration procedure can be repeated as necessary to improve the touch screen performance. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 37 of 132...
Begin the automated portion of the System Check by navigating to the System Check icon in the menu. Select MENU, SYSTEM then SYSTEM CHECK (Figure 8-4). Figure 8-4 – System Check Menu Screen SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 38 of 132...
Page 39
Systems, LLC. Obtain the backup Driver and prepare it for use as indicated in this section. (15) Send a copy of the completed form to SynCardia Systems, LLC by fax at the number indicated on the form. 8.1.12. Set Date and Time...
Screen displayed when selecting MENU, HELP in order to provide users with access to any contact numbers entered in this screen. Select CONFIRM when entry is completed. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 40 of 132...
• Patient ID • TAH-t Serial Number • Hospital The following measured or calculated values are recorded in the log file(s): • Time • Rate SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 41 of 132...
Page 42
Press COPY to start. o Copying will take a few seconds. o Press DONE, then remove USB device. o To exit at any time, select CANCEL. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 42 of 132...
Select “+” to increase the setting and “-” to decrease the setting. Pressing the “+” or “-” button once will change the value by increments of one. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 44 of 132...
All adjustments in O.R. Mode occur without delay when the “+” increase or “-” decrease button is selected. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 45 of 132...
When the Rate is below 40, “- -” is shown for the Rate and single pulse actuation becomes an option. Select the One Time icon, “1X” (beneath the “+” button) to perform a single pulse actuation (Figure 8-10). SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 46 of 132...
TAH-t rate is zero. The stroke volumes will be low until the patient is completely off bypass. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 47 of 132...
MENU, SETUP, FLIP SCREEN (Figure 8-11). This will change the orientation of the Driver display only and will not affect the Hospital Cart display. Figure 8-11 – Flip Screen Menu SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 48 of 132...
Full Eject Flag Figure 8-12 – Zoomed View of Pressure Waveform Showing Full Eject Flag SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 49 of 132...
Page 50
65mL or greater (Figure 8-13). An acceptable partial fill for the 70cc TAH-t is fill volume between 50 and 60 cc. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 50 of 132...
• Right Fill Volume High Fill volume for the 50cc TAH-t should be monitored and Driver settings adjusted if the patient is experiencing or approaching a full fill condition (> 40cc) SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 51 of 132...
TAH-t is operating at a partial fill condition. It may not be necessary to adjust all parameters to resolve the issue. • Increase rate • Decrease vacuum • Increase systolic duration SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 52 of 132...
No password is required to change from O.R. Mode to I.C.U. Mode because there are more restrictions in I.C.U. Mode than in O.R. Mode. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 53 of 132...
Power on the backup Driver. (Note: Alarms are disabled for the first 2 minutes). Once the Driver software has loaded, the system will show the Ambulatory Mode Screen (Figure 8-16). Figure 8-16 – Ambulatory Mode SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 54 of 132...
Page 55
Change to I.C.U. Mode by selecting MENU, I.C.U. MODE and enter the password, 9722. Press the LOCK Icon in the lower right corner to unlock the touch screen. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 55 of 132...
Page 56
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Adjustment Locked/Unlocked The default condition for the I.C.U. Mode is “locked,” which is indicated by the LOCK Icon (Figure 9-1). Figure 9-1 – I.C.U. Mode Screen – Locked SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 57 of 132...
The adjustment window in I.C.U. Mode shows both the current setting and the new setting, as shown in Figure 9-3. Figure 9-3 – I.C.U. Mode Parameter Adjustment SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 58 of 132...
Page 59
UNLOCKED icon to return to the locked position. 9.2.4. Holding the “+” or “-” button for two seconds will cause the parameter to immediately change by increased increments, as shown in Table 9-1. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 59 of 132...
* Actual vacuum achieved is dependent on the patient systemic pressures and resistance and the Driver parameter settings. In general, increasing vacuum increases fill volumes and decreases in vacuum decrease fill volume. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 60 of 132...
Once the Driver software has loaded, the system will show the Ambulatory Mode Screen (Figure 10-1). Alarm Alarm Mute Status Bar Button Power Status Window Average Cardiac Output Window Maximize Display Button Figure 10-1 – Ambulatory Mode SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 61 of 132...
• Percent Systole • Left Cardiac Output • Right Cardiac Output • Left Hourly Cardiac Output Average • Right Hourly Cardiac Output Average • Left Fill Volume SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 62 of 132...
Page 63
• Press COPY to start. • Copying will take a few seconds. • Press DONE, then remove USB device. • To exit at any time, select CANCEL. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 63 of 132...
“+” or “-” button to change to the desired value. Select ENTER to accept the change (Figure 10-4). Figure 10-4 – Set Date/Time SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 64 of 132...
Page 65
Change the operating mode to I.C.U. Mode by completing the following steps: (1) Select MENU, MODE, and I.C.U. MODE. (2) Enter the I.C.U. Mode password (9722) and press CONFIRM to change to I.C.U. Mode. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 65 of 132...
Page 66
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Chapter 11. List of Symbols Table 11-1 – Symbols Used in Companion 2 Driver System Labeling Symbol Description Symbol Description "OFF" for a part of "ON" (power) equipment Temporary Audible Alarm Audible Alarm Disabled silenced Protected against objects Refer to instruction manual/ IP30 >...
Page 68
19: Maximum diameter of cell (in mm) 65: Maximum overall height of the diameter (in mm) 5: Number of battery cells connected in parallel within a cell pack SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 68 of 132...
Investigate and correct the cause of any alarm. 12.3. Use of Backup Driver A SynCardia approved Driver must be available as a backup to be used in the event of a failure of the primary Driver. 12.4. Maintenance of Electrical Safety The Companion 2 Driver System is designed to meet AAMI/NFPA electrical safety requirements.
Page 70
12.5. Fuses Fuses are user replaceable. 12.6. Reduction of Explosion Hazard Do NOT operate the Companion 2 Driver System in the presence of flammable anesthetics or other flammable gases. 12.7. Maintenance of Patient Safety Verify that all pneumatic and electrical connections are in place, secure, and connected to the Driver in use before attempting to operate the System.
Page 71
12.11. Companion 2 Driver System Cleaning Exercise extreme care in cleaning the Companion 2 Driver System. DO NOT soak any item during cleaning or disinfection. DO NOT allow water or solvent to come in direct contact with electrical connectors. See Chapter 20, Equipment Maintenance and Care, for more information.
Page 72
• The Companion 2 Driver is docked in either a Hospital Cart or Driver Caddy. • A backup Driver is available that has successfully completed System Check. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 72 of 132...
• Make sure that the External Power Indicator on the Driver turns green. If the light does not turn green, the Driver is using battery power. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 73 of 132...
Hospital Cart Power Inlet Figure 13-2 – Connecting External Power with the Hospital Cart Driver Caddy Power Inlet Figure 13-3 – Connecting External Power with the Companion Driver Caddy SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 74 of 132...
Driver and DO NOT attempt to dock the Driver again into the same Hospital Cart or Caddy. A malfunction of the Hospital Cart or Caddy may exist and it must be returned to SynCardia for evaluation. The Driver should be docked into another Hospital Cart or Caddy.
Battery Indicator begins flashing yellow and an alarm sounds every 30 seconds. The Battery symbol in the Power Status Window on the display shows a red battery. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 76 of 132...
Companion 2 Driver display (counter) will take approximately one minute to update the estimated battery support time based on Driver settings. Power Status Window Counter Figure 13-6 – Battery Power Status Window SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 77 of 132...
Symbol with an “X” (Figure 13-8). Battery 2 Removed Battery 1 Full Charge Wall Power Connected Figure 13-8 – Battery Power Status Window with One Battery Removed SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 78 of 132...
Figure 13-9, flashes green when the batteries are being charged. Alarm Indicator External Power Indicator External Battery Indicators Emergency Battery Indicator Figure 13-9 – Driver Power Status LED Array SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 79 of 132...
The system must be reconnected immediately to an alternate power source, such as the wall power or External Batteries or the patient must be switched to a SynCardia approved Backup Driver. SynCardia Systems, LLC Companion 2 Driver System Operator Manual...
Table 14-1 for alarms and their corrective actions. For alarms observed while the Driver is not supporting a patient, contact SynCardia for additional information. If more than one alarm occurs, the highest priority alarm message is shown first in the Alarm Bar.
Page 82
11 hours. Cycle power on the Driver. If Battery Error alarm persists, return the Driver to Emergency Battery 70-80 dBA SynCardia. Indicator: Flashing Red Notify SynCardia at 1-866-771-9437 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 82 of 132...
Page 83
Assess patient conditions that may Imbalance contribute to a high Cardiac Output. (high) 70-80 dBA If alarm persists, replace Driver with a SynCardia approved backup Driver. Notify SynCardia at 1-866-771-9437 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 83 of 132...
Page 84
Left Ventricle seconds Over Alarm Indicator LED : Red If alarm persists, replace Driver with a Pressure SynCardia approved backup Driver. 70-80 dBA Notify SynCardia at 1-866-771-9437 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 84 of 132...
Page 85
Alarm Bar: Flashing 14.4). Right Yellow Pressure If alarm persists, replace Driver with a 70-80 dBA Incorrect Alarm Indicator: Yellow SynCardia approved backup Driver. Notify SynCardia at 1-866-771-9437 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 85 of 132...
Page 86
Battery Yellow gauge. Battery Indicator: Yellow 70-80 dBA If alarm persists, replace Driver with a SynCardia approved backup Driver. Notify SynCardia at 1-866-771-9437 SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 86 of 132...
Page 87
2 beeps every 3½ Replace Driver with a SynCardia Alarm Bar: Yellow Driver minutes approved backup Driver. Requires Alarm Indicator LED: Notify SynCardia at Service Yellow 1-866-771-9437 70-80 dBA SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 87 of 132...
Figure 14-1 – Alarm Status Screen The Alarm Status Screen may be accessed by pressing MENU, ALARMS or by pressing on the alarm bar anywhere other than the Alarm Mute Button. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 88 of 132...
The Alarm History can be viewed by touching the Alarm Status Bar and then selecting the Alarm History Tab or by pressing MENU, ALARMS, and ALARM HISTORY. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 89 of 132...
Scroll Bar Figure 14-3 – Alarm Settings Screen Alarm settings can be adjusted by scrolling up or down, selecting the alarm name and pressing the EDIT button. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 90 of 132...
Alarm limits and settings are stored in the Driver and persist during short or extended power interruptions. This allows the user to configure alarm settings on the backup Driver as directed in Section 8.2.8, Setup of Backup Driver. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 91 of 132...
Battery Low Left Ventricle 0 to 20 300 mmHg Overpressure mmHg mmHg Right Ventricle 0 to 20 90 mmHg 170 mmHg Overpressure mmHg Very Low External minutes Battery SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 92 of 132...
Page 93
Left Fill Volume High alarm and Right Fill Volume High alarm only applies to Driver use with the 70cc TAH-t (this alarm is not present with the 50cc TAH-t). SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 93 of 132...
Solid “X” over the Alarm Disabled in O.R. reminder signal procedure, otherwise change to Mute Button Mode Reminder every 2 hours I.C.U. Mode or Ambulatory Mode to activate audible alarms SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 94 of 132...
Page 95
External air Connection No audible signal No action required Connected Icon enabled (black) External air External air Connection No audible signal No action required Disconnected Icon disabled (grey) SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 95 of 132...
Page 96
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Driveline connector. Attach the Driveline connector to the connection on the backup Driver in the same direction (as shown in Figure 15-1). Figure 15-1 – Properly Connected Drivelines SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 97 of 132...
Page 98
Driver. Turn off the primary Driver. The key will lock into place when the primary Driver is set to the OFF position. 15.2.6. Notify SynCardia Systems that the Driver has been switched. Report any alarms or problems at this time.
Setup of Backup Driver, including the System Check. If the driver does not pass the Setup, do not connect the patient to the Companion 2 Driver and contact SynCardia immediately. 16.2.2. Connect the Companion Drivelines to the Driver, using the single Driveline connector.
Drivelines of the Companion 2 Driver within reach. 16.3.3. Perform steps 16.3.4 and 16.3.5 simultaneously in order to disconnect the Drivelines of the CSS Console and connect the Drivelines of the Companion 2 Driver. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 100 of 132...
The Wire Ties around the Cannulae and CPC connectors must be secured using a Wire Tie Gun. Secure each Cannula with two Wire Ties Figure 16-4 – Securing Cannulae to CPC Connectors with Wire Ties SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 101 of 132...
Page 102
16.3.7. It is recommended that the CSS Console Drivelines be retained or disposed of in accordance with hospital procedures. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 102 of 132...
NOTE: The procedure described below to switch drivers is the same for both the 50cc TAH-t and 70cc TAH-t. 17.1. Materials Prior to switching from the Companion 2 Driver System to the CSS Console, assemble the following materials: • One set of CSS Console Drivelines •...
Page 104
17.3.4. Reinsert a locking tie under the metal release button on the CPC connections to prevent inadvertent disconnect. Note: The TAH-t is now connected to the CSS Console via CPC connectors. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 104 of 132...
NOTE: The Freedom Driver is set up not to activate a Fault Alarm for low Cardiac Output for 15 minutes after it has been initially turned on, to allow time to connect to a TAH-t. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 105 of 132...
Page 106
When performing a switch from the Companion 2 Driver to the Freedom Driver, make certain the Drivelines are properly connected to the TAH-t Cannulae, and do not kink the Drivelines. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 106 of 132...
CPC Connector that secures TAH-t Cannula to the Companion 2 Driveline (Figure 18-2). Gently pull to remove the Wire Tie and discard. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 107 of 132...
• Immediately insert the Cannula into the Freedom Driveline, • Insert until a click is heard. • Lightly tug on the connection to make sure that it is secure. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 108 of 132...
Press Pull Pull Figure 18-3 – Disconnecting Companion 2 Drivelines from the Cannulae Push in Push in Figure 18-4 – Connecting the Freedom Drivelines to the Cannulae SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 109 of 132...
The patient’s name and the date of the switch must be written on the bag and retained by the Hospital to switch the patient back to the Companion 2 Driver, if necessary. The Hospital should notify SynCardia Systems that the patient has been switched. SynCardia Systems, LLC...
Backup Driver, including the System Check. 19.2.2 Connect the Companion Drivelines to the Driver using the single Driveline connector. 19.2.3 Position the Driver that is ON next to the patient. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 111 of 132...
Page 112
19.3.4 Cut off the excess ties and cover with Velcro. 19.3.5 The cannulae from the TAH-t are now connected to the Drivelines on the Companion 2 Driver. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 112 of 132...
Check (Appendix A) upon receipt, prior to use, and every 6 months if the Driver is not in use. In addition, SynCardia will track the Driver runtime and notify the center when a Driver is due for scheduled service. The Driver Caddy may require air to be added to the tires periodically.
20.4.2 To access the filter, press the Filter Cover Latch (Figure 20-2) and pull the Filter Cover away from the Driver body. Filter Cover Latch Filter Cover Figure 20-2 – Driver Side Panel and Filter Cover SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 114 of 132...
20.4.5 After cleaning or replacing the filter, follow these steps to clear the alarm: (1) Select MENU (2) Select SETUP (3) The FILTER MAINT. REQUIRED button should be active (see Figure 20-4) SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 115 of 132...
(4) Select the FILTER MAINT. REQUIRED button (5) This will bring up a verification screen asking “Has the filter maintenance been completed” (see Figure 20-5). Select YES. Figure 20-5 – Verification Screen SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 116 of 132...
(6) Once the alarm has been cleared, the FILTER MAINT. REQUIRED button will be grayed out (see Figure 20-6). Figure 20-6 – SETUP Menu with Filter Maintenance Alarm Cleared SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 117 of 132...
Page 118
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 118 of 132...
21.1.8 Once the internal Emergency Battery and External Batteries are charged, the system is ready for clinical use. CAUTION Do not rapidly rotate the key between the ON setting and OFF setting. SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 119 of 132...
Page 120
Driver and DO NOT attempt to dock the Driver again into the same Hospital Cart or Caddy. A malfunction of the Hospital Cart or Caddy may exist and it must be returned to SynCardia for evaluation. The Driver should be docked into another Hospital Cart or Caddy.
Chapter 22. Companion 2 Driver System Specifications 22.1 Safety Standards The Companion 2 Driver System, as part of the SynCardia TAH-t Companion 2 Driver System, has been thoroughly tested pursuant the requirements of European safety standard IEC 60601-1: 2006. This standard requires making the following declarations and stating the type and degree of protection for listed hazards.
Guidance Concerning Electromagnetic Emissions The Companion 2 Driver System components are intended for use in the electromagnetic environment specified in Table 22-2. The customer or the user of the device should assure that it is used in such an environment.
22.1.2 Declaration and Guidance Concerning Electromagnetic Immunity for all Companion 2 Driver System The Companion 2 Driver System components are intended for use in the electromagnetic environment specified in Table 22-3. The customer or the user of the components should assure that they are used in such an environment.
Page 124
The customer or the user of either device should assure that they are used in such an electromagnetic environment. Table 22-4 – Declaration and Guidance Concerning Electromagnetic Immunity for Life- Sustaining Companion 2 Driver System, including the TAH-t, Companion 2 Driver and Batteries 60601...
Page 125
80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended decrease likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into the patient areas. For this reason, an SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 125 of 132...
Page 126
The Companion 2 Driver System has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2:2001 Medical electrical equipment − Part 1-2: General requirements for safety −...
Page 127
• Battery Storage Temperature: 0° to 40°C (32° to 104°F) • Transportation/Storage Pressure: 700 to 1060 mBar (525 to 795 mmHg) • Transportation/Storage Humidity: 10 – 100% RH (non- condensing) SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 127 of 132...
Unidirectional USB port protected with a USB Port Driver side panel rubber cap that allows information to be copied from the Driver SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 128 of 132...
Page 129
• Externally-Powered Equipment when connected to an external AC (wall) power source The Driver is intended for CONTINUOUS OPERATION 22.10 Software Versions in the Companion 2 Driver System • Embedded Supervisor: v1.25 • Graphical User Interface: v1.0.2160 SynCardia Systems, LLC...
Page 130
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Appendix A – Companion 2 Driver System Check Form SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 131 of 132...
Page 132
SynCardia Systems, LLC Companion 2 Driver System Operator Manual Page 132 of 132...
Page 133
Verify the system passed all Tests 9. Upon completion of System Check, press “Close” 10. Proceed to Battery Tests on Page 2. Comments: ____________________________________________________________________________ ____________________________________________________________________________ __________ SynCardia Systems, LLC Revision Date: 2017-02 Companion 2 Driver System Check Form: C2-900005 Rev 014 Page 1 of 2...
Page 134
Fail Verify the driver displays and LED lights indicate that the wall power is connected 10. Turn the system off. If the Driver failed any of the above steps, contact SynCardia Systems, LLC Help Line at (866) 771-9437. 11. Sign, date and fax form to 520-903-1782 12.