DUALER IQ Pro™
MANUAL
(800) 985-8324 or (385) 695-5000
Fax: (385) 695-5001
www.jtechmedical.com
© 2015 JTECH Medical MN056_F
MN056_F.indd 1
5/18/2015 8:48:02 AM

Advertisement

Table of Contents
loading

Summary of Contents for JTECH Medical DUALER IQ Pro Series

  • Page 1 DUALER IQ Pro™ MANUAL (800) 985-8324 or (385) 695-5000 Fax: (385) 695-5001 www.jtechmedical.com © 2015 JTECH Medical MN056_F MN056_F.indd 1 5/18/2015 8:48:02 AM...
  • Page 2 This Page Intentionally Left Blank © 2015 JTECH Medical MN056_F MN056_F.indd 2 5/18/2015 8:48:02 AM...
  • Page 3: Table Of Contents

    Simulated Goniometer . . . . . . . . . . . . . . . . . . . 13 Viewing and Changing Testing Defaults . . . . . . 14 © 2015 JTECH Medical MN056_F MN056_F.indd 1...
  • Page 4: Publication Information

    Fax (385) 695-5001 www.jtechmedical.com © 2015 JTECH Medical. All rights reserved. This manual is provided as a service to our customers, any other use is strictly prohibited. This manual may not be reproduced by any means, physically or electronically, without the express written consent of JTECH Medical.
  • Page 5: Graphic Symbol Definitions

    An EN 980 symbol Identifying the Manufacturer An EN 980 symbol Identifying the Authorized Representative for European Communities An IEC 60878 symbol Signifying Alternating Current An IEC 60878 symbol Signifying Direct Current © 2015 JTECH Medical MN056_F MN056_F.indd 3 5/18/2015 8:48:03 AM...
  • Page 6: Conventions Used Throughout This Manual

    • Enter Button - The singular button on the top of the Dualer IQ Pro Primary Inclinometer, which can be used to enter the test zero and repitition values. • Device - This refers to a singular JTECH Medical testing device, i.e. only the Dualer IQ Pro Primary Inclinometer or only the Dualer IQ Pro Secondary Inclinometer.
  • Page 7: Primary Inclinometer Led States

    There Are No Saved Tests This message is displayed when attempting to enter the review mode and no saved tests exist in memory © 2015 JTECH Medical MN056_F MN056_F.indd 5 5/18/2015 8:48:04 AM...
  • Page 8: Warnings And Notifications

    Opening of housings by anyone other than an authorized JTECH service representative will void your warranty. Warning: Changes, or modifications, not expressly approved by JTECH Medical could void the user’s authority to operate the equipment. Warning: Use only a factory-supplied charger. Use of another charger may result in electrical shock or equipment damage.
  • Page 9: Emc Notice

    Safety and EMC requirements for medical electrical systems (based on the IEC 60601-1 and IEC 60601-1-2 standards) JTECH Medical and its manufacturing facility are certified to the following technical standards. ISO 9001:2008 ISO 13485:2003 with Canadian Medical Devices Conformity Assessment System (CMDCAS)
  • Page 10: Dualer Iq Pro (Cm101/Cm102)

    Connection Cable (7WR020) – The Connection Cable is used to connect the Primary and Secondary inclinometers for dual inclinometry use. © 2015 JTECH Medical MN056_F MN056_F.indd 8 5/18/2015 8:48:15 AM...
  • Page 11: System Setup

    2 . Highlight “System,” and press the Select Button. 3 . Press Up/Down until “Info” is displayed, and press the Select Button. 4 . The LCD screen will show the firmware revision, battery voltage, and calibration information. © 2015 JTECH Medical MN056_F MN056_F.indd 9 5/18/2015 8:48:15 AM...
  • Page 12: Using The Dualer Iq Pro

    Settings - Select this option to set test defaults, for more information see “Settings” above or “Global System Settings” on page 14. For menu structure see “Menu Tree” on the next page © 2015 JTECH Medical MN056_F MN056_F.indd 10 5/18/2015 8:48:15 AM...
  • Page 13: Menu Tree

    Select – Change / Cycle through Testing Defaults Tests – 1-20 Reps – 1-6 Mode – Auto. Rep, Static, Dynamic Sound – On / O Info Calibrate Restore Clear Tests Exit © 2015 JTECH Medical MN056_F MN056_F.indd 11 5/18/2015 8:48:15 AM...
  • Page 14: Performing An Exam

    4 . Press select to cycle through the planes of motion prior to beginning a test.  Note: The Dualer IQ Pro does not require a full calibration of the Inclinometers prior to starting each exam. © 2015 JTECH Medical MN056_F MN056_F.indd 12...
  • Page 15: Static Testing Mode

    The Dual Inclinometers are very accurate for measuring extremity ranges of motion because it effectively eliminates goniometer alignment errors and substantially increases inter- and intra-rater repeatability. For instructions on performing specific tests, please refer to the optional Multimedia Help System © 2015 JTECH Medical MN056_F MN056_F.indd 13 5/18/2015 8:48:15 AM...
  • Page 16: Viewing And Changing Testing Defaults

    • Calibrate - This puts the Dualer IQ Pro into user calibration mode. • Restore - This overwrites user calibration data with factory calibration data. • Clear Tests - This deletes all saved test data. © 2015 JTECH Medical MN056_F MN056_F.indd 14 5/18/2015 8:48:16 AM...
  • Page 17: Calibration Settings

     Note: Restoring the factory calibration only overwrites the user calibration data with the factory calibration data stored on the instrument.  Note: For information on recommended annual factory recalibration, contact JTECH Customer Service at 1-800-985-8324, ext. 4 or service@jtechmedical.com © 2015 JTECH Medical MN056_F MN056_F.indd 15 5/18/2015 8:48:16 AM...
  • Page 18: Performing A Full Calibration

    16 . Press the Enter Button to save this position. 17 . Connect the Secondary inclinometer and repeat the process  Note: For information on recommended annual factory recalibration, contact JTECH Customer Service at 1-800-985-8324, ext. 4 or service@jtechmedical.com © 2015 JTECH Medical MN056_F MN056_F.indd 16 5/18/2015 8:48:16 AM...
  • Page 19: Features And Benefits

    Module automatically calculates test validity and impairments to the whole-person level based on your test data . For more information on upgrades, contact your JTECH sales representative or call JTECH at 1-800-985-8324 or 1-385-695-5000. © 2015 JTECH Medical MN056_F MN056_F.indd 17 5/18/2015 8:48:16 AM...
  • Page 20: General Information

    (e.g. weeks) can reduce the overall life of the battery. Leaving instrument connected over the weekend should not affect the life of the batteries. © 2015 JTECH Medical MN056_F MN056_F.indd 18 5/18/2015 8:48:16 AM...
  • Page 21: Troubleshooting Your Dualer Iq Pro

    What happens if my instrument does not recharge? Because of the sensitive electronics inside, the Dualer IQ Pro is not designed for battery replacement in the field. If an instrument fails to hold a charge, contact JTECH Medical Customer Service for repair or replacement options.
  • Page 22: Storage And Cleaning Of Instrument

    Dualer IQ Pro devices have no associated methods of preventative inspection. Maintenance Dualer IQ Pro is not user serviceable and does not include a service manual. For any service including maintenance, repair, and calibration, contact JTECH Medical Customer Support at 1-800-985-8324 or 1-385-695-5000 © 2015 JTECH Medical MN056_F MN056_F.indd 20...
  • Page 23: Calibration Of Instrument

    Dualer IQ Pro may require repair. If this unlikely event arises, please follow these instructions in order to ensure accurate and prompt servicing of your instrument. 1 . Contact JTECH Medical Customer Support at 1-800-985-8324 ext. 4 to obtain a Return Material Authorization (RMA) number.
  • Page 24: Customer Support

    (repair, replacement, shipping, etc.), after the initial warranty time period. You can buy extended warranties for your Dualer IQ Pro hardware. Please contact JTECH Medical Customer Service for more information on JTECH Guardian Hardware Maintenance plans. Before Calling for Support 1.
  • Page 25: Product Return Policy

    When returning merchandise, please include a copy of your original invoice or packing slip to ensure prompt issuing of credit. Return the Dualer IQ Pro (including all accessories) postage paid and insured to JTECH Medical.  Notice: JTECH Medical is not responsible for loss or damage during shipping.
  • Page 26: Hardware Limited Warranty

    PROVIDED BY STATE LAW. NO PERSON HAS ANY AUTHORITY TO BIND JTECH TO ANY REPRESENTATION OR WARRANTY EXCEPT AS SPECIFICALLY SET FORTH HEREIN.  For Government Use: Please refer to the hardware warranty on the installation media you received. © 2015 JTECH Medical MN056_F MN056_F.indd 24 5/18/2015 8:48:17 AM...
  • Page 27 This Page Intentionally Left Blank © 2015 JTECH Medical MN056_F MN056_F.indd 25 5/18/2015 8:48:17 AM...
  • Page 28: Technical Specifications

    Degree of Safety of Application in the This device is not suitable for use in the presence of flammable presence of flammable mixtures anesthetic mixture with air, or with oxygen or nitrous oxide IPX0 Symbols © 2015 JTECH Medical MN056_F MN056_F.indd 26 5/18/2015 8:48:17 AM...
  • Page 29 Degree of Safety of Application in the This device is not suitable for use in the presence of flammable presence of flammable mixtures anesthetic mixture with air, or with oxygen or nitrous oxide IPX0 Symbols © 2015 JTECH Medical MN056_F MN056_F.indd 27 5/18/2015 8:48:18 AM...
  • Page 30 Degree of Safety of Application in the This device is not suitable for use in the presence of flammable presence of flammable mixtures anesthetic mixture with air, or with oxygen or nitrous oxide Symbols IPX0 © 2015 JTECH Medical MN056_F MN056_F.indd 28 5/18/2015 8:48:19 AM...
  • Page 31: Appendix

    Alignment Rail (8MH068) – The Alignment Rail magnetically attaches to the Inclinometers for easy visual confirmation of alignment. VELCRO® Strap (47) – The 28 inch VELCRO® strap can be used for dynamic range of motion inclinometry testing. © 2015 JTECH Medical MN056_F MN056_F.indd 29 5/18/2015 8:48:24 AM...
  • Page 32: Accessories And Detachable Parts (Continued)

    Wrist Lanyard (7AC006) – The Wrist Lanyard is used to help prevent dropping devices when carrying or while performing some tests. Connection Cable (7WR020) – The Connection Cable is used to connect the Primary and Secondary inclinometers for dual inclinometry use. © 2015 JTECH Medical MN056_F MN056_F.indd 30 5/18/2015 8:48:27 AM...
  • Page 33 This Page Intentionally Left Blank © 2015 JTECH Medical MN056_F MN056_F.indd 31 5/18/2015 8:48:27 AM...
  • Page 34: Emc Guidance

    The Dualer IQ Pro instrument should not be used while stacked on, or adjacent to, other electrical or medical electrical equipment. If stacked or adjacent use is necessary, all electrical equipment should be observed to verify normal operation. © 2015 JTECH Medical MN056_F MN056_F.indd 32 5/18/2015 8:48:27 AM...
  • Page 35 3 A/m 3 A/m location in a typical magnetic field commercial or hospital environment. IEC 61000-4-8 NOTE U is the a.c. mains voltage prior to application of the test level. © 2015 JTECH Medical MN056_F MN056_F.indd 33 5/18/2015 8:48:28 AM...
  • Page 36 If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Dualer IQ Pro. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m. © 2015 JTECH Medical MN056_F MN056_F.indd 34 5/18/2015 8:48:28 AM...
  • Page 37 NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. © 2015 JTECH Medical MN056_F MN056_F.indd 35 5/18/2015 8:48:28 AM...
  • Page 38: Index

    Initial Battery Charge . . . . . . . . . . . . . . . . . . . . . . . . 9 Intended Use and Indications for Use . . . . . . . . . . 7 © 2015 JTECH Medical MN056_F MN056_F.indd 36...
  • Page 39 This Page Intentionally Left Blank © 2015 JTECH Medical MN056_F MN056_F.indd 37 5/18/2015 8:48:28 AM...