PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Contents GENERAL WARNINGS ....................................4 1.1. SYMBOLS ........................................4 1.2. INTENDED USE ......................................5 1.2.1. CLASSIFICATION AND REFERENCE STANDARDS ........................5 1.2.2. ENVIRONMENTAL CONDITIONS..............................5 1.2.2.1. INSTALLATION CONDITIONS ................................5 1.2.3. WARRANTY .......................................5 1.2.4. DISPOSING THE EQUIPMENT WHEN NO LONGER USED ......................6 1.3.
Page 3
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL MAINTENANCE ......................................59 9.1. MAINTAINING THE INSTRUMENTS ................................ 59 9.2. DRAINING CONDENSATE ..................................59 9.3. CLEANING THE SURGICAL SUCTION FILTER ............................60 9.4. SURGICAL SUCTION ....................................60 9.5. CATTANI SURGICAL SEPARATOR ................................. 61 9.6.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL GENERAL WARNINGS • These instructions explain how to correctly use the following dental units: PUMA ELI, PUMA ELI CP, PUMA ELI CART, PUMA ELI ORTHO, PUMA ELI CART ISO, PUMA ELI A, PUMA ELI A CP.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 1.2. INTENDED USE • The dental units described in this manual are Medical Devices intended for dental treatment. • The dentist's board may hold up to 5 instruments. • The assistant’s board can hold 2 suction tubes and 2 instruments.
The floor (continuous) should meet the load-bearing capacity set forth by DIN 1055, sheet 3. PUMA ELI, PUMA ELI CP, PUMA ELI CART, PUMA ELI ORTHO, PUMA ELI CART ISO dental units: the weight of the dental unit is of approx. 350 kg (including a 190 kg patient).
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL The dental unit, once installed, is exposed to possible contaminants coming from the water mains. So, to effectively overcome this problem, it is advisable to install the dental unit only when its use will be daily and to perform the disinfecting procedures starting from the day in which it is installed by following the instructions set forth in the relevant sections.
Page 8
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Cleaning and disinfecting instructions. Clean and disinfect with disposable non-abrasive paper (avoid using recycled paper) or sterile gauze. Do not use sponges or, in any case, any material that can be reused.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL DESCRIPTION OF THE EQUIPMENT 2.1. IDENTIFICATION PLATES Dental unit model: PUMA ELI, PUMA ELI CP, PUMA ELI CART, PUMA ELI ORTHO. The plate is found on the arm between the dental chair and water unit.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 2.2. DENTAL UNITS Models: Model PUMA ELI RS version dentist's board (instruments will return to their original position through the pulling action of the spring-operated arms) attached to a double supporting arm, one of which is articulated and self-balancing.
Page 11
Connection box. i Foot control. o Height-adjustable cart. The dental units of the PUMA ELI A series come in the following models: Model PUMA ELI A RS version dentist's board (instruments will return to their original position through the pulling action of the spring-operated arms) attached to a double supporting arm, one of which is articulated and self-balancing.
Negatoscope for panoramic X-rays (optional). z DAMA A dental chair. 2.3. DENTAL CHAIR (1) DAMA dental chair (models PUMA ELI - PUMA ELI CP - PUMA ELI CART - PUMA ELI ORTHO) Description of the different parts: a Headrest.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 2.4. SPECIAL WARNINGS FOR PUMA ELI CART and PUMA ELI CART ISO models During installation, the cart board can be adjusted in height to 3 predefined working positions: MIN (80 cm) - MED (90 cm) - MAX (100 cm).
Page 14
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL • Complete the rotation to the left of the water unit by moving it to the standard use position for left-handed operators. NOTE: make sure that element ( k ) is correctly locking short arm in the new position.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL TURNING ON THE OPERATING UNIT Dental unit model: PUMA ELI PUMA ELI CP PUMA ELI CART Press the main switch ( f1 ) on the dental chair casing: Equipment ON: - electric system powered...
- with instrument extracted and working: all the chair movements are inhibited. • Dental units model PUMA ELI A and PUMA ELI A CP: water unit supporting arms are equipped with a safety device ( o ) that immediately stops the dental chair from moving down if an obstacle is...
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 4.2. EMERGENCY DEVICES WARNING: Use the devices below when movement of the equipment needs to be stopped: • Dental chair control buttons ( a ) or ( c ). By pressing any of the dental chair control buttons all movements will be stopped.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 4.4. MOVABLE ARMRESTS (OPTIONAL) Both armrests can be turned clockwise downwards so that the patient can more conveniently get on and off the chair. Safety warnings. WARNING: Maximum load supported by armrest: 68 kg.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL DENTIST'S BOARD OPERATION Layout of instruments. The positions of the instruments on the board are determined by the customer at the time of order. Activation of instruments. • The syringe is always on (see paragraph 5.3.).
Page 20
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Tray holder. The tray holder module ( f ) is made of stainless steel and can easily be removed from its support. WARNING: Maximum permitted load on the tray holder ( f ): 2 kg evenly distributed.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.1. DENTIST’S CONTROL CONSOLE Description of the buttons: INCREASE button: increases the settable values. DECREASE button: decreases the settable values. Fibre optics lighting ON/OFF button. Micromotor rotation direction inversion control button with relevant operating LED.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL SETTING THE DENTAL CHAIR "AUTOMATIC RETURN” AND "RINSE POSITION” 5.1.1. • Bring the dental chair into the desired position with the manual movement buttons. • Hold button “SAVE” for at least 2 seconds to activate save mode.
These dental chair functions are activated by keeping the pedal at the end of the travel for at least 2 seconds. NOTE: these functions are NOT active in PUMA ELI ORTHO models. Dental chair movement Joystick operation ( 3 ). It controls the following movements: Dental chair seat up.
NOTE: this function is NOT active in PUMA ELI ORTHO models. Right-hand button operation ( 5 ). • Holding down the button for at least 2 seconds with instrument extracted:...
Page 25
” NOTE: these functions are NOT active in PUMA ELI ORTHO models. • Holding down the button for at least 2 seconds with instrument extracted: Chip-air control: sends a jet of air to the Turbine or the Micromotor.
Activation of the “Automatic dental chair return” programme. NOTE: these functions are NOT active in PUMA ELI ORTHO models. • Holding down the button for at least 2 seconds with instrument extracted: Water Clean System Control: sends a jet of running water to instruments such as the Turbine, the Micromotor and the Scaler for rinsing the spray ducts.
Page 27
NOTE: the lever functions only if the instruments are in rest position.This function is NOT active in the PUMA ELI ORTHO models. For safety reasons, the selected function starts only after the lever has been briefly actuated and then released.
2 seconds. NOTE: these functions are NOT active in PUMA ELI ORTHO models. Operation with instrument removed • Operating lever ( 2 ) to the right: instrument is started and its speed/position is adjusted.
Page 29
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Activation/deactivation of spray water to instruments. Spray is delivered by keeping control ( 4 ) pressed with the instrument active ( lever ( 2 ) operated ) NOTE: the foot control can be set so that when the instrument is selected and lever ( 2 ) is in rest position, pressing the button ( 4 ) will enable operation to always be with spray (ON/OFF).
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.2.5. WIRELESS FOOT CONTROL The "side travel" and "push-pedal" foot controls can also be supplied in WIRELESS version. The wireless foot control contains a ZIGBEE transmitter module (module certified for Europe, Canada and the USA).
Page 31
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Limited battery warranty. The battery in the foot control is covered by a 6-month warranty from the date of installation. Recharging the battery. When the batteries in the WIRELESS foot control need to be recharged, Proceed as follows: •...
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.3. SYRINGE Description of the instrument. a Nozzle. b Handpiece. c Syringe release button. d Air button. e Water button. f Hot/cold selector (only for 6-function syringes). g Hot/cold indicator LED (only for 6-function syringes).
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.4. TURBINE Connecting the handpiece and changing the drill. Refer to the specific instructions provided with the handpiece. Use. WARNING: Read the instructions for use of the various turbines. • Operating times: work 5 min., rest 5 min.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.5. MICROMOTOR Installing the handpieces and changing the drill. Refer to the specific instructions provided with the micromotor and various handpieces. Use. WARNING: Also read the instructions for use of the various motors.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL • Never put the contra angle on the micromotor while it is running. • The drill release button must not be pressed down during operation! Friction between button and micromotor impeller overheats the head and may cause burns.
Page 36
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Warnings for use. WARNING: • The instrument is supplied NOT STERILE and must be sterilised before use. Carefully read the operating instructions supplied with the instrument before attempting to sterilise. • Before attempting to connect the handpiece, make sure the contacts are perfectly dry.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.7. T LED CURING LIGHT Technical specifications. Supply voltage: 24-36 VDC Max. power absorbed: 6 VA Light source: 1 5W LED Wavelength: 430-490 nm Acoustic signals: at cycle start, every 5 seconds, and at cycle end Type of operation: intermittent (work 3 consecutive cycles - rest 60 sec.)
Page 38
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Operation. WARNING: The instrument is supplied non-sterile. Before use, disinfect the lamp grip. The optical fibre and the eye protection can be sterilised in a steam autoclave at 135°C. • Put the fiber optics ( c ) in its housing until it clicks.
Page 39
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Warnings for use. WARNING: The LED is a Class 2 light source in accordance with IEC 62471 standard. DO NOT STARE AT THE BEAM. The light emitted may cause eye injury in case of direct radiation without eye protection.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 5.8. C-U2 DENTAL CAMERA The C-U2 dental camera system, complete with an extremely lightweight ergonomic handpiece, is specially designed for simple and well-conceived usability in examining the oral cavity. Auto-exposure and fixed focus features provide easy operation. This system is designed to allow the dentist to more efficiently show and explain to patients all oral conditions and reasons for planned treatment.
Page 41
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL • FREEZE function. Handpiece C-U2 is equipped with a touch-sensitive button ( g ). Tap this button or actuate the foot control to freeze the image of interest. Tap this button again (or actuate the foot control) to unlock the frozen image.
Page 42
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Handpiece status. An optical guide, illuminated by a multicolour LED indicator, found in the area near the control button ( g ), shows handpiece status as per the table given below:...
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL ASSISTANT’S BOARD OPERATION Dental units model PUMA ELI - PUMA ELI CP - PUMA ELI CART - PUMA ELI ORTHO: (1) Assistant's board with twin arm. Main features: • Two articulated arms ( c ) secure the board ( a ) to the water unit ( b ) allowing it to be placed in the most convenient work position.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Cleaning the sliding rollers. Push down and take off the sliding rollers (f). Clean the sliding rollers with a suitable product: refer to paragraph 1.4. 6.1. ASSISTANT’S BOARD CONSOLE Description of the buttons: Water-to-bowl button.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 6.3. T LED CURING LIGHT For detailed information regarding operation of this instrument, see paragraph 5.7. 6.4. C-U2 DENTAL CAMERA For detailed information regarding operation of this instrument, see paragraph 5.8.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Maintenance. Periodically lubricate the O-rings of the cannula holder terminals (see Paragraph 9.4.) using S1-Protective Lubricant for O-Rings CEFLA s.c.. Note about biocompatibility. Only use suction tubes supplied with the dental unit and original replacement tubes.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 6.7. TURBINE For detailed information regarding operation of this instrument, see paragraph 5.4. The cock ( f ), found under the board, is used to adjust the amount of water in the spray.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL WATER UNIT OPERATION 7.1. FILL CUP AND BOWL The bowl can be turned 305° on the water unit by hand. The bowl, the water-to-cup spout and the water-to-bowl spout can be removed to make the cleaning operations easier.
Page 49
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL Water-to-bowl controller modification upon recall of dental chair "Rinse position". The controller that delivers water to the bowl when the dental chair "Rinse Position" button is pressed can be turned on/off.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 7.2. S.S.S. SYSTEM. -ST Applied on models PUMA ELI , PUMA ELI CP e PUMA ELI CART, PUMA ELI A , PUMA ELI A CP, only. Description of the system. The system S.S.S. -ST (Separate Supply System) features a tank ( a ) for distilled water.
Description of the system. System features a tank ( a ) for distilled water. Tank has a total capacity of 1.1 litres (models PUMA ELI ORTHO) and 1.8 litres (models PUMA ELI SINGLE CART). Tank supplies: • the sprays of all the instruments present on dentist's and assistant's boards, •...
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 7.2.2. MANUAL DISINFECTION CYCLE The S.S.S. system allows performing a manual disinfection cycle of the water ducts of all the instruments on the dentist's board, on the assistant's board and on the water-to-cup duct (optional).
7.3. DISINFECTION SYSTEM Applied on models PUMA ELI , PUMA ELI CP e PUMA ELI CART, PUMA ELI A , PUMA ELI A CP, only. The AUTOSTERIL -ST system allows performing a disinfection cycle of...
Page 54
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL C) Performing the disinfection cycle: • Shortly press the "AS" button to activate the disinfection cycle. NOTE: an acoustic signal (1 BEEP) will confirm that cycle has started. • The system now replaces the water in the ducts.
Flushing cycle setting. • If you want to run a flushing cycle with distilled water, enable the S.S.S. system (see paragraph 7.2. or 7.2.1. based on the available PUMA ELI model). NOTE: it is advisable to execute the flushing cycle with tank (a) completely full.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 7.5. OPENING/CLOSING THE WATER UNIT SIDE COVER Opening the cover. • open the cover on the side of the water unit ( a ) after pushing up and releasing the lock lever ( b ).
The instructions for use and maintenance of the monitor are included with the equipment. 8.3. AUXILIARY TRAY HOLDER Applied on models PUMA ELI CART and PUMA ELI ORTHO, only. Tray holder can contain two standard size trays. Turn knob ( b ) to adjust vertical movement based on load: •...
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 8.4. NEGATOSCOPE FOR PANORAMIC X-RAYS An x-ray film viewer for panoramic x-rays can be mounted on all INTERNATIONAL version dentist's boards. The screen dimensions are as follows: H=210mm, L=300mm. Turn on negatoscope, simply turn special switch ( a ): Negatoscope on.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL MAINTENANCE Preventive maintenance CEFLA s.c., the manufacturer of the dental units, in accordance with applicable standards IEC 60601-1 3rd Ed. - 2007, IEC 62353 and directive MDD 93/42, and subsequent changes, for medical devices underlines that the preventive maintenance checks for the dental unit specified in the Technical care manual and Maintenance and warranty handbook are to be carried out by authorised personnel at least once every 12 months.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 9.3. CLEANING THE SURGICAL SUCTION FILTER This operation should be done daily at the end of work. WARNING: Always wear goggles and gloves to prevent contact with infected material when cleaning the suction filters.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 9.5. CATTANI SURGICAL SEPARATOR At the beginning of each work day. Insert inside filter ( d ) a tablet ( v ) of VF CONTROL PLUS (CEFLA s.c.) WARNING: Always wear gloves to prevent contact with infected material when carrying out this operation.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 9.6. CATTANI GRAVITY SURGICAL SEPARATOR Draining the separator’s container. • Move the dental chair completely up so that as much liquid as possible can be drained from the container. WARNING: Always wear goggles and gloves to prevent contact with infected material when carrying out this operation.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 9.9. DÜRR AMALGAM SEPARATOR The maintenance instructions for the DÜRR amalgam separator are enclosed with the equipment if the equipment comes with this type of separator. The separator control device is located in the water unit.
T 4 A 21 V~ : MULTIMEDIA power line. Dental chair card area. * Technical specifications refer to models: PUMA ELI, PUMA ELI CP, PUMA ELI CART, PUMA ELI ORTHO, PUMA ELI CART ISO, PUMA ELI A , PUMA ELI A CP.
PUMA ELI / PUMA ELI A – OPERATOR'S MANUAL 11. DENTAL UNIT OPERATING UNIT MAINTENANCE PLAN REFERENCE WHEN PART ACTION PARAGRAPH Drain condensate. Eliminate condensate from air ducts. See paragraph 9.2. Disinfection of spray water ducts with manual cycle (if the Spray ducts ( if the dental See paragraph 7.2.2.