UVpad E
Manual
Version: 1.1.2

Advertisement

Table of Contents
loading

Summary of Contents for Opsytec UVpad E

  • Page 1 Opsytec Dr. Gröbel GmbH Am Hardtwald 6-8 76275 Ettlingen Tel.: 07243 / 94783 - 50 Fax: 07243 / 94 83 - 65 info@uv-groebel.de UVpad E Manual Version: 1.1.2...
  • Page 2: Table Of Contents

    The UVpad E displays „MAX“ ................31 12.2 The UVpad E displays too little irradiance values ..........31 12.3 The UVpad E displays too high or too low UVC Values ........31 12.4 The UVpad E displays “ - “ ..................32 12.5 The measured value is not reproducible ..............
  • Page 3: Change History

    Meaning: Failure to observe the mentioned instructions can re- sult in a damage of the device. Meaning: Instructions are to be observed for the regular opera- tion. Optional functions, not available for every UVpad E...
  • Page 4: Function Overview

    4 Function Overview 4 Function Overview The UVpad E combines the benefits of the spectral measuring technology with a handheld and easy to operate radiometer. Use the spectroradiometer UVpad E for an accurate measurement of the UV irradiance and for comparisons of different UV lamps and UV LEDs. 512 photodi- odes measure the spectrum in the wavelength range of 200 –...
  • Page 5: Technical Data

    5 Technical Data 5 Technical Data 5.1 General technical data Size (L x W x H) 160 x 100 x 20 mm³ Weight ~500 g Power battery powered, 3 x CR2032 Max. operational temperature 60°C Display graphical, 128 x 64 px Optically active surface Diameter: 10 mm Optically active surface position...
  • Page 6: Usb

    5 Technical Data 5.4 USB / hardware-software-requirements USB connector MiniUSB PC requirements min Intel CORE i3, 2 GB Ram, >40 Gb HDD operating system Windows 7 with .NET Framework > 4.0 5.5 Effect related measurements Photoinitiator / Sensitivity curve optional...
  • Page 7: Intended Use

    Operation only allowed in dry environments. The assembly position is arbitrary. When using the UVpad E, light and UV radiation can be reflected and distributed at the housing. Appropriate protection measures for the protection against radiation need to be applied, if neces- sary.
  • Page 8: Important Information

    The maximum housing temperature of the sensor can reach > 60 °C and there is the risk of skin burns on contact. Please allow for the UVpad E to cool down and use protective gloves to avoid skin burns.
  • Page 9: Operation

    9 Operation 9 Operation The UVpad E is operated via four keys. These are arranged on the front panel below the dis- play. The keys and the functions are illustrated as follows: Function  Back / Cancel / Off ...
  • Page 10: Standby Menu

    After 2 minutes of not using the UVpad E, the device switches off automatically. 9.2 Standby Menu Via the standby menu, all functions of the UVpad E are available for you. These are: • Start measurement • Show last measurement •...
  • Page 11: Start Measurement

    9 Operation 9.2.1 Start Measurement From the standby menu, a measurement is started with the key . First, the spectrum is displayed. Use ‘”Auto” mode to get the best integration time or, in fixed mode use  and  keys to adjust integration time.
  • Page 12  you can return to the standby screen. Note: If the irradiance is too low for the measurement range, then the UVpad E dis- plays no measured values. The irradiance and the dosage values are then dis- played with “-“.
  • Page 13: Start Manual Dark Measurement

    9 Operation 9.2.2 Start manual dark measurement For manual darkness measurement, select the menu item "dark+radiometric" under Settings, Measurement mode. If your UVpad E is equipped with several effective functions, manual dark measurement may be available for further effective functions. Note: Manual darkness measurement is only active in the sensitivity mode "fixed".
  • Page 14 9 Operation 4. The irradiance is displayed. 5. To save the measurement, press the  button for at least two seconds. After release, the irradiation values are dis- played. Note: The irradiance is not displayed if the darkness measurement is not done or the integration time has been changed.
  • Page 15: Show Last Measurement

    If using fixed integration time, during a measurement you can set up the sensitivity of the UVpad E with the keys  and . Select  for a lower sensitivity and  for a higher sensitiv- ity. The sensitivity is not saved.
  • Page 16 9 Operation Select a higher sensitivity (), if the spectrum is noisy or does not fill the display completely. Select a lower sensitivity (), if single lines of the spectrum are cut on top. The measurement range is then exceeded and is shown in the display MAX. Examples: Sensitivity too low Sensitivity correct...
  • Page 17: Recording Time

    9.2.6 Settings In the menu „Settings“, you can set up the sensitivity, the effective function, the menu lan- guage and the date / time. Furthermore, you can reset the UVpad E to the factory settings. 9.2.7 Information Select the menu item “Information” from the standby menu, to display the serial number, the firmware version, as well as the date of calibration.
  • Page 18: Settings

    Reset to factory settings 9.3.1 Sensitivity The UVpad E adjusts the sensitivity automatically, in order to ensure the optimal signal-noise ratio. The sensitivity is hereby set via the sampling rate. With high irradiances, the sampling rate is increased in order to not exceed the measurement range. With low irradiances the sam- pling rate is decreased.
  • Page 19: Averaging

    This reduces sensitivity by factor of 5. 9.3.2 Averaging The UVpad E can calculate the irradiance from the moving average. Hereby are signal-to-noise ratio, and thus measurement accuracy can be enhanced. For integration times below 500 ms we recommend averaging over 16-32 measurements, while for integration times over 500 ms smaller settings may be useful.
  • Page 20 “(ef)” for “effective”. Radiometric and effect-related measure- ments are illustrated subsequently. Example: radiometric Example: with effective function The function “Effective Function” is only available, if your UVpad E has been programmed with a customized effective function at factory.
  • Page 21: Language

     and confirm with . 9.3.5 Date / Time The UVpad E has a real-time clock to check the validity of the calibration. You can set up the real-time clock in the menu „Date / Time“. First set the day with the keys  and . Se- lect ...
  • Page 22: Data Export / Pc-Software

    Connect the UVpad E with the PC and switch it on. The screen displays the message “USB connected”. Once the UVpad E is connected with the PC, it does not react to any keystroke. Disconnect the connection with the PC and the UVpad E continues working autonomously.
  • Page 23: Viewer Software

    Importing can take a few seconds. During this time, a progress bar displays the import: If the UVpad E is not connected or not ready yet, then an error message pops up. Restart the import after a few seconds via the button „Connect“.
  • Page 24 You can select and compare different measurements. You can also add a comment in the table before exporting data. Note: The UVpad E Viewer displays measurement settings and results as sensitivity, saturation and errors. Use mouse over function in table view this.
  • Page 25: Export Of Measurements

    10 Data Export / PC-Software 10.4 Export of measurements Measurements can be exported individually, in a group or completely. For this purpose, mark up to 10 measurements. Button „Deselect all“ deselect marked measurements. Select “Export selected measurements” or “Export all measurements” for the data export. The measurements are exported as a text file, separated by a comma, and can thus be imported to a table calculation program, such as Microsoft EXCEL.
  • Page 26: Displaying And Comparing Measurements

    10 Data Export / PC-Software 10.5 Displaying and comparing Measurements Select the desired display via the tab: Below is an example with two irradiance measurements. Up to 10 measurements are selectable at the same time and are labeled in the display area with date / time.
  • Page 27: Deleting Measurements

    10.7 Deleting Measurements You can delete the entire memory of the UVpad E via the button “Delete all measurements”. Confirm the confirmation prompt with OK. Individual measurements cannot be deleted. Tip: Via the button „Info “, the serial number and the assignment of memory...
  • Page 28: Remote Control / Live Measurement

    11 Remote control / Live measurement 11 Remote control / Live measurement In addition to the export functions, the UVpad E software also has a remote control called "Live measurement". The button with the same name takes you to a separate user interface for remote control of the UVpad E.
  • Page 29 In the "Auto" mode a dark current measurement is done directly in the device. In "Auto" mode, dark current correction is carried out as usual in the UVpad E device. “Calibration + Auto” mode corresponds to the measurement result of the UVpad E.
  • Page 30 11 Remote control / Live measurement When exporting all measurements, in addition to the averaged measurement result, all individ- ual measurements as well as the measurement results for irradiance are exported to a *. csv file. The export of a "single measurement" only provides the averaged measurement result. All measurement settings are also saved in the export file.
  • Page 31: Notes For Optimal Measurement Results

    The UVpad E achieves optimal measurement results, if the measurement range is well adjusted to the irradiance and thus to the UV system. This is generally taken out automatically. The following notes are there to support you with the setting of the UVpad E, so that optimal measurement results can be achieved.
  • Page 32: The Uvpad E Displays

    12.4 The UVpad E displays “ - “ Just like each measurement device, the UVpad E has a fixed measurement range. If this range is fallen below, then the UV radiation cannot be captured correctly. In that case, the UVpad E displays “ – “.
  • Page 33: Cleaning, Maintenance & Troubleshooting

    13 Cleaning, Maintenance & Troubleshooting 13.1 Calibration The calibration of the UVpad Es is checked during the initialization. If the UVpad E is not calibrated properly anymore, then this is indicated by a note after the initialization process. Measurements are also possible at any time with an expired calibration.
  • Page 34: Battery Change

    The measured data will be stored. Date and time should be adjusted. The UVpad E is operated with lithium batteries. For a battery change, switch the UVpad E off and turn it with the rear panel facing towards you. Loosen the three marked screws with a screwdriver (Torx T6) and remove the cover.
  • Page 35: Troubleshooting

    Change batteries. switched on Interruption Restart the device. For this purpose, remove all batteries and replace them after 5 minutes. The UVpad E is in delivery status. The measured value is not Automatic, low or high sen- Use fixed sensitivity reproducible...
  • Page 36: Declaration Of Conformity

    14 Declaration of Conformity 14 Declaration of Conformity Hereby we, Opsytec Dr. Gröbel GmbH Am Hardtwald 6-8 76275 Ettlingen declare that the following product spectral Radiometer Type designation: UVpad E have been developed, constructed and produced under our sole responsibility and that the...
  • Page 37: Notes

    15 Notes 15 Notes...