Set UVpad for high belt speeds ................29 11.3 The UVpad displays too little irradiance values ............. 30 11.4 The UVpad displays too high or too low UVC Values ........... 30 11.5 The UVpad displays “ - “ ..................30 11.6 The measured value is not reproducible ..............
Meaning: Failure to observe the mentioned instructions can re- sult in a damage of the device. Meaning: Instructions are to be observed for the regular opera- tion. Optional functions, not available for every UVpad...
Here, the UVpad offers enormous application advantages compared to the common spectral radiometers. Not only is the UVpad the worlds flattest autonomic spectral radiometer, it also offers a high- quality spectral measurement technology in a robust and easy to operate measurement device.
4 Function Overview 4.1 Quick start guide This quick start guide is intended to explain the function of the UV pad by means of an operat- ing procedure. Turn on the UV pad with any key on the front Step 1 panel.
Page 7
Connect the device to your PC using a Mini Step 8 USB cable. With the software UVpad Viewer you can now transfer your measurements to the PC. For more information, see chapter 10...
5 Technical Data 5 Technical Data 5.1 General technical data Size (L x W x H) 160 x 100 x 14,4 mm³ Weight 375 g Power battery powered, 3 x CR2032 Max. operational temperature 70°C Display graphical, 128 x 64 px Optically active surface Diameter: 6 mm Optically active surface position...
5 Technical Data 5.3 spectral ranges The classification of the UV regions in the spectral radiometer is carried out according to CIE as follows: UVC 200 – 280 nm; UVB 280 – 315 nm; UVA 315 – 400 nm; VIS 200 – 440 nm. according UV-C 200 - 280 nm...
It can be used in UV curing systems, if the maximum operating temper- ature of 70 °C is not exceeded. For this reason, the UVpad is equipped with a thermal shield on the rear panel. The rear panel of the UVpad may be exposed to intensive UV radiation. The front panel must not be exposed to UV radiation.
The maximum housing temperature can reach > 60 °C and there is the risk of skin burns on contact. Please allow for the UVpad to cool down and use pro- tective gloves to avoid skin burns.
9 Operation 9 Operation The UVpad is operated via four keys. These are arranged on the front panel below the display. The keys and the functions are illustrated as follows: Function Back / Cancel / Off Down ...
After 2 minutes of not using the UVpad, the device switches off automatically. 9.2 Standby Menu Via the standby menu, all functions of the UVpad are available for you. These are: Start measurement Show last measurement Scope measurement ...
9 Operation 9.2.1 Start Measurement From the standby menu, a measurement is started with the key -key. The measurement device displays that a measurement is in progress. Stop the measurement by pressing -key. Turn the measurement device and position it for the measurement. 1.
Page 15
For optimal measurement results, the measurement range has to be fully modulated. For this purpose, the UVpad has a 12-bit analogue/digital converter, which scans the spectrum continu- ously. Excellent measurement results are achieved, when the measurement range is used to at least 50%.
-key or -key you can return to the standby screen. Note: If the irradiance is too low for the measurement range, then the UVpad dis- plays no measured values. The irradiance and the dosage values are then dis- played with “-“.
9 Operation Delete all measurements: The internal memory can be deleted directly on the UVpad or via a PC. For this purpose, please select the prior menu item “Clear memory”. Confirm with Yes (-key). Return with -key or -key. 9.2.3 Scope Measurement A scope measurement is a real-time view of the spectral irradiance.
Protect your skin and eyes during the work at and with UV radiation sources. 9.2.4 Settings In the menu „Settings“, you can set up the sensitivity, the effective function, the menu lan- guage and the date / time. Furthermore, you can reset the UVpad to the factory settings.
Reset to factory settings 9.3.1 Sensitivity The UVpad adjusts the sensitivity automatically, in order to ensure the optimal signal-noise ratio. The sensitivity is hereby set via the sampling rate. With high irradiances, the sampling rate is increased in order to not exceed the measurement range. With low irradiances the sam- pling rate is decreased.
Page 20
This setting is suitable for most applications. In the setting „low“, the sampling rate is al- ways between 10 ms and 100ms. The UVpad can measure UV-systems with a very high irradiance or significant irradiance modifica- tions (e.g. with shutter operation or with very high tape speed).
Example: radiometric Example: with effective function The function “Effective Function” is only available, if your UVpad has been programmed with a customized effective function at factory. 9.3.3 Measurement delay Select a measurement delay in order to start a new measurement delayed.
-key and -key and confirm with -key. 9.3.5 Date / Time The UVpad has a real-time clock to check the validity of the calibration. You can set up the real-time clock in the menu „Date / Time“. First set the day with the keys -key and -key.
Windows 7. 10.2 Connecting with the PC Connect the UVpad with the PC and switch it on. The screen displays the message “USB con- nected”. Once the UVpad is connected with the PC, it does not react to any keystroke. Disconnect the...
Importing can take a few seconds. During this time, a progress bar displays the import: If the UVpad is not connected or not ready yet, then an error message pops up. Restart the im- port after a few seconds via the button „Connect“.
Page 25
You can select and compare different measurements. You can also add a comment in the table before exporting data. Note: The UVpad Viewer displays measurement settings and results as sensi- tivity, saturation and errors. Use mouse over function in table view this.
10 Data Export / PC Software 10.4 Export of measurements Measurements can be exported individually, in a group or completely. For this purpose, mark up to 10 measurements. Button „Deselect all“ deselect marked measurements. Select “Export selected measurements” or “Export all measurements” for the data export. The measurements are exported as a text file, separated by a comma, and can thus be imported to a table calculation program, such as Microsoft EXCEL.
10 Data Export / PC Software 10.5 Displaying and comparing Measurements Select the desired display via the tab: Below is an example with two irradiance measurements. Up to 10 measurements are selectable at the same time and are labeled in the display area with date / time.
10.7 Deleting Measurements You can delete the entire memory of the UVpad via the button “Delete all measurements”. Confirm the confirmation prompt with OK. Individual measurements cannot be deleted. Tip: Via the button „Info “, the serial number and the assignment of memory...
The measurement time for the UVpad is set via the integration time. Typical integration times are 10 ms to 200 ms. If the UVpad is moved underneath an UV system at slow speed, then the maximum irradiance (directly underneath the UV lamp) is being determined, since several measurements are available for the irradiance.
Furthermore, the UV lamps generally emit visible light and IR radiation apart from the UV radiation. Both can generate scattered light in the UVpad, which is then detected. It is also pos- sible that the UVpad heats up during the measurement and thus the reading noise increases.
11 Notes for optimal measurement results 11.6 The measured value is not reproducible Measurements are only stored and displayed in UVpad when the saturation is between 10% and 100%. This can cause to saturation direct under the light source, however correct meas- urement may be recorded before or after that.
12 Cleaning, Maintenance & Troubleshooting 12.1 Calibration The calibration of the UVpads is checked during the initialization. If the UVpad is not calibrat- ed properly anymore, then this is indicated by a note after the initialization process. Measure- ments are also possible at any time with an expired calibration.
The measured data will be stored. Date and time should be adjusted. The UVpad is operated with lithium batteries. For a battery change, switch the UVpad off and turn it with the rear panel facing towards you. Loosen the three marked screws with a screw- driver (Torx T6) and remove the cover.
Page 34
Change batteries. switched on Interruption Restart the device. For this purpose, remove all batteries and replace them after 5 minutes. The UVpad is in delivery status. The measured value is not Automatic, low or high sen- Use fixed sensitivity reproducible...
Page 35
13 Declaration of Conformity 13 Declaration of Conformity Hereby we, Opsytec Dr. Gröbel GmbH Am Hardtwald 6-8 76275 Ettlingen declare that the following product spectral Radiometer Type designation: UVpad have been developed, constructed and produced under our sole responsibility and that the...