Operation and Service Guide

General Information

General Information
This operation and service guide contains information about initial inspec-
tion, operation, performance tests, troubleshooting and repair of the E4412A
and E4413A power sensors.
Warranty
The power sensors are warranted and certified as indicated on the inside
cover of this manual.
Instruments Covered by Manual
These instruments have a two-part serial number: the prefix (two letters and
the first four numbers), and the suffix (the last four numbers). The two letters
identify the country in which the unit was manufactured. The four numbers
of the prefix are a code identifying the date of the last major design change
incorporated in your Agilent Technologies product. The four-digit suffix is a
sequential number and, coupled with the prefix, provides a unique identifica-
tion for each unit produced. The contents of this manual apply directly to all
serial numbers unless otherwise indicated
Description
The E4412A and E4413A power sensors are diode power sensors. They are
intended for measurement of CW microwave power levels in a wide
+
dynamic range from
70 dBm to
20 dBm (100 pW to 100 mW). The
E4412A measures at frequencies from 10 MHz to 18.0 GHz. The E4413A
measures at frequencies from 50 MHz to 26.5 GHz. These are high-speed
power sensors, and do not incorporate narrow-bandwidth averaging used in
average-power sensors. Signals with digital, pulse, or other forms of ampli-
tude modulation may introduce measurement errors. Multi-tone signals
(containing multiple frequency components), or signals with significant har-
monic content (>
45 dBc) may introduce measurement errors at high power
levels. (Specifications for the power sensors are in Table 1.)