Operation - Scion Instruments 456-GC Reference Manual

Transformer gas analyzer
Hide thumbs Also See for 456-GC:

Advertisement

Scion Instruments

OPERATION

Operational method is preloaded in the instrument.
Listed below is an example method; settings determining the application
may be different.
Sample delivery:
V1 (GSV)
Initial
0.1
3.0
9.9
Flow/Pressure:
Front EFC-24 (Ar):
Pressure (psi)
Initial
Total Flow:
Middle EFC-24 (Ar):
Pressure (psi)
Initial
Total Flow:
Column Oven:
Temp (ºC)
Initial
Valve oven:
Temp
Methanizer oven:
Temp:
Hydrogen flow:
Detector (front TCD):
Temp:
Range:
Time cons:
Fila temp:
Scion Instruments
V2 (CS)
-
-
+
-
+
+
-
-
Hold (min)
27.5
10
60 ml/min
Hold (min)
38.5
10
60 ml/min
Rate (ºC/min)
70
100(ºC)
400 °C
20 ml/min
200 ºC
0.05
slow
254 °C
Transformer Gas Analyzer Reference Manual
Hold (min)
--
10
Total (min)
10
Page: 7