Technical
Information
TI 177F/00/en
Radiometric Level Measurement
gammapilot FTG 671
Non-invasive, radiometric limit detection
Intrinsically safe signal circuit [EEx ib] IIC /
ATEX II (2) GD
For scintillation or Geiger-Müller detector
Application
The Gammapilot FTG 671 transmitter is
designed for
• non-invasive, level limit detection
in containers - mixers, reactors,
hoppers, silos, tanks - with inflammable,
abrasive, poisonous and aggressive
bulk material and liquids, e.g. in acid
tanks, boilers, cement silos, ballast
hoppers, cyclones, cupolas, rotary
furnaces etc.. Thanks to the fact that the
measuring system is fitted externally, it
also finds application in the food
processing industry.
Gammapilot transmitters may also be
used for applications in explosion
hazardous areas, and possess
intrinsically-safe sensor circuits
conforming to EEx ib IIC, or EEx d IIC,
EEx de IIC depending on the detector.
Endress
The Power of Know How
The Gammapilot
FTG 671 transmitter is a
19" plug-in card, shown
here in Monorack II
housing
Features and Benefits
• Suitable for scintillation detector
DG 57 or Geiger-Müller counters
DG 17 or DG 27
• Uses lowest possible source activity
with DG 57/100 mm or longer
• Digital signal transmission
- Interference-free function with
two-core installation cable up to
1000 m
• Intrinsically safe sensor circuits
- Ex-protection withoutZener barriers
approved as overspill protection
(VBF, WHG Germany) and for use
in explosion hazardous areas
• Simple operation with parameter
matrix
- Uniform handling as for all
Commutec transmitters
• Practical background functions
- Automatic compensation of source
decay
- Automatic adjustment of switch
point
• Commutec transmitter
- Simple connection to automatic
systems via Rackbus
- Up- and download of data
+
Hauser