Contents 1. The installation and components of equipment ..............1 2. Installation and adjustment .....................5 3. Maintenance and sterilization ..................6 4. Precaution ........................10 5. After service .........................12 6. Symbol instruction .......................12 7. Environmental protection .....................13 8. Manufacturer’s right .....................13 9. European authorized representative ................13 10.
Ultrasonic scaler S6 LED is composed of electrocircuit, water way and ultrasonic transducer. 1.2.3 Scope of application Ultrasonic scaler S6 LED is used for the dental calculus elimination and root canal treatment. 1.3 The main technical specifications Technical description is included in the instructions for use.
Page 4
1.3.15 Degree of protection against harmful ingress of water: Ordinary equipment, the foot switch is drip-proof equipment (IPX1) 1.3.16 Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air, Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.
Page 5
1.4.2 Back of the main unit sketch map Picture 2 1.4.3 Instruction of using wrench Picture 3...
Object Remark Type No./model No./ Manufacturer/Trademark Detachable handpiece HW-8L Scaler tips Power suplly See picture 5 ZMN-03 Guilin Woodpecker F2 food switch See picture 6 Woodpecker Endochuck TD-E1 Torque wrench TD-1L Endo wrench 2. Installation and adjustment 2.1 Operation 2.1.1 Open the packing box, make sure that all the parts and accessories are complete according to the packing list, take the main unit out of the box, and put it on the the stable plane facing to the operator.In addtion,install the device to an operator-friendly position.
2.1.13 After finishing operation, keep the machine working for 30 seconds with the water supply to clean the handpiece and the tip. 2.1.14 Unscrew the scaling tip and sterilize it. 2.1.15 After operating, turn off the power switch to cut off the power, then pull off the plug of power adapter. Note: Don't screw the scaling tips when stepping on the foot switch, and the machine is working.
3.3 Disinfection and sterilization of torque wrench and Endo wrench 3.3.1 Torque wrench and Endo wrench can be sterilized with any commonly used non-corrosive neutral disinfectant, or they can be sterilized by autoclaving (Please refer to 3.2.1). 3.3.2 Do not disinfect or sterilize the torque wrench in the following ways: a) Put it in solution and cook it;...
Page 10
Fault Possible cause Solutions The scaling tip vibrates but there is no spray when The water control knob is not on. Turn on the water contrl knob [note 1]. stepping on the foot switch. The tip hasn't been screwed on to the handpiece tightly. Screw the tip on the handpiece tightly (See Picture 4).
Page 11
Fault Possible cause Solutions There is no water coming out from the handpiece There is air in the water pipe. Turn the water control to the Max, reinsert the bottle. (automatic water supply mode). If the problem still can't be solved, please contact with local dealer or manufacturer. Manufacturer will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to service personnel in parts repair.
I.The maintenance, repair and modification are made by the manufacturer or the authorized dealer. II. The changed components are original of "DTE" and operated correctly according to instruction manual. III. Maintenance is forbidden while the device is used for treatment.
Page 13
4.2 Contraindication 4.2.1 The patient who has hemophilia is not allowed to use this equipment. 4.2.2 The patient or doctor who with heart pacemaker is forbidden to use this equipment. 4.2.3 The heart disease patient, pregnant woman and children should be cautious to use the equipment. 4.3 Storage and maintenance 4.3.1 The equipment should be handled carefully and lightly.
5. After service Within one year from the date of sale, the device enjoys one year free repair by providing warranty card. The repair of the equipment should be carried out by professional technician. We are not responsible for any irretrievable damage caused by non- professional person.
Guidance and manufacturer’s declaration - electromagnetic emissions The models S6 LED are intended for use in the electromagnetic environment specified below. The customer or the user of the models S6 LED should assure that it is used in such an environment.
Page 16
If the variations on power 40 % UT for 0.5 cycle user of the models S6 LED require continued supply input lines (60% dip in UT) for 5 cycles 40 % UT operation during power mains interruptions, it is IEC 61000-4-11.
Page 17
Guidance & Declaration - Electromagnetic immunity The models S6 LED are intended for use in the electromagnetic environment specified below. The customer or the user of the models S6 LED should assure that it is used in such an environment.
Recommended separation distances between portable and mobile RF communications equipment and the models S6 LED The models S6 LED are intended for use in electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the models S6 LED can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the models S6 LED are recommended below, according to the maximum output power of the communications equipment.
Need help?
Do you have a question about the S6 LED and is the answer not in the manual?
Questions and answers