Procedures and Processes ................3-1 Daily Start Up Procedures ................3-1 Check Status and Clear Sample Racks ............3-2 When to Do ....................3-2 Result ......................3-2 UniCel DxC 600/800 In-Lab Training Manual A24183 Table of Contents May 2005 Page 1 of 4...
Page 4
Quick Tips ....................3-18 Additional References ................3-18 Procedure ....................3-18 Programming and Processing Patient Samples Manually ......3-19 When to Do ....................3-19 Result ....................... 3-19 Table of Contents UniCel DxC 600/800 In-Lab Training Manual A24183 Page 2 of 4 May 2005...
Page 5
Instrument Commands ................4-8 Answer Key ....................4-9 Reagents ..................... 4-9 Calibration ....................4-10 Samples ....................4-10 UDR ......................4-10 Instrument Commands ................4-10 UniCel DxC 600/800 In-Lab Training Manual A24183 Table of Contents May 2005 Page 3 of 4...
Page 6
Table of Contents UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4 of 4 May 2005...
The STOP button is located near the Off Load Tray and it is used to mechanically stop all instrument movements. UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005 Page 1-3...
Home position. The interior of the probe is rinsed and the waste is collected while at the Home position. System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-4 May 2005...
Detailed information regarding the Obstruction Detection Assembly can be found in the UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 2, System Description. UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005 Page 1-5...
DxC 600 systems have one-cup module for the rapid Module(s) analysis of glucose. DxC 800 systems have 6 individual cup modules for the rapid analysis of glucose, BUN/UREA, creatinine, phosphorus, albumin and total protein. System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-6 May 2005...
The EIC mixes the sample with Electrolyte Buffer reagent prior to the delivery (of sample) to the flow cell. 1. Waste Outlet 4. Reference Inlet 2. Flow Cell Outlet 5. DI H 0 Inlet 3. Buffer Inlet Figure 1.6 System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-8 May 2005...
• Stir bar and mixer motor assembly - Effectively mixes the contents of the reaction cup • A detection system - (Chemistry dependent) serves to detect light transmitted or electrical changes UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005 Page 1-9...
(Optional) Note: Also known as Near Infrared Particle Immuno Assay (NIPIA). Cuvette Wash Station Washes cuvettes at the completion of analysis in preparation for the next chemistry. System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-10 May 2005...
Both reagent probes on the UniCel DxC work in tandem to provide maximum throughput and efficient reagent delivery. UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005...
Photometer assembly. Additional information about the LPIA assembly may be found in the UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 2, System Description, Cuvette Reaction System. System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-12 May 2005...
Water. The last probe removes DI water from the cuvette, leaving it dry and ready for the next chemistry. 1. Probe #1 4. Probe #4 2. Probe #2 5. Wash station 3. Probe #3 Figure 1.10 Cuvette Wash Station UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005 Page 1-13...
DxC System. For detailed information on the operation of these components, please refer to the individual console computer, monitor or printer’s Operations Manual. UniCel DxC 600/800 In-Lab Training Manual A24183 System Components May 2005...
Page 22
System Console Cuvette Wash Station System Components UniCel DxC 600/800 In-Lab Training Manual A24183 Page 1-16 May 2005...
5. Sample Status Icons 2. Host Indicator 6. Rack Status 3. System Status Indicator 7. Function Bar 4. Function Select Icons Figure 2.1 UniCel DxC Main Operator Screen UniCel DxC 600/800 In-Lab Training Manual A24183 Software Overview May 2005 Page 2-1...
These icons provide access to each of the major functional areas of the system. Table 2.1 Function Select Icons Icon Functional Area Main Screen Sample Programming Results Recall Reagents/Calibration Set Up Utilities Status Instrument Commands (1 of 2) Software Overview UniCel DxC 600/800 In-Lab Training Manual A24183 Page 2-2 May 2005...
Carousel. As each rack is loaded, the rack number is displayed at the top of the Rack Status box. All samples on the rack are listed below the rack number. UniCel DxC 600/800 In-Lab Training Manual A24183 Software Overview May 2005...
Page 26
Main Operator Screen Rack Status Software Overview UniCel DxC 600/800 In-Lab Training Manual A24183 Page 2-4 May 2005...
• Program (if necessary) and process controls. • Review control results to verify system operation. • Program (if necessary) and process patient samples. • Review patient results. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-1...
For more information about this procedure, refer to the UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 3, Preparing Samples for Analysis, How to Use Reserved Racks, Clearing Samples. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-2 May 2005...
Highlight the Racks(s) Field and type the rack number(s) to be cleared. Select <OK> to clear the racks. Select <OK> to confirm the rack(s) to be cleared. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-3...
• Adequate # of tests remaining for CC reagents. • Reagents do not have a reagent status of "Reagent Expired" or "Days Exceeded." Unload/Load reagents as indicated. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-4 May 2005...
Page 31
• Load the new reagent bottle in the MC Bulk Reagent Compartment. Repeat Steps 5–7/8 for unloading and loading of additional MC reagents. Select Done [F10]. Before processing patient samples, calibrate and run QC. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-5...
• UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 5, Reagent Load/Calibration. • SYNCHRON Clinical Systems Chemistry Information Manual, refer to chemistry specific Chemistry Information Sheet. • Reagent Preparation Quick Reference. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-6 May 2005...
Clear [F1]. • Do not close the reagent carousel door. Continue with the next step to load a reagent. (1 of 2) UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-7...
Page 34
If multiple positions were selected for unloading/loading, repeat Step 5 and Step 6. Before processing patient samples, calibrate and run QC if needed. (2 of 2) Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-8 May 2005...
Select the chemistries to be calibrated. Use the <Page Up> and <Page Down> buttons to select more chemistries. Select Cal [F4]. Select List [F5] to review the Calibrator Load List. (1 of 2) UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-9...
Page 36
Place the rack(s) in the autoloader with the rack’s bar code label to the right. Press the RUN button on the analyzer. (2 of 2) Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-10 May 2005...
For more information about this procedure, refer to the UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 3, Preparing Samples for Analysis. All trademarks are the property of their respective owners. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005...
• Remove the cap, if present. • Determine if volume is adequate. • Check for fibrin. • Ensure that container is placed in a reserved or non-reserved rack. (1 of 3) Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-12 May 2005...
Page 39
Unreadable Bar Code Patient Samples" damaged and cannot be in this manual. replaced Not programmed • Go to "Programming and Processing Patient Samples Manually" in this manual. (2 of 3) UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-13...
Page 40
A014600L.EPS Figure 3.1 Bar Code Label Placement (3 of 3) All trademarks are the property of their respective owners. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-14 May 2005...
Additional References For more information about this procedure, refer to the UniCel DxC Synchron Clinical Systems Instructions For Use manual, CHAPTER 6, Quality Control. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-15...
Pour the control material into a 0.5 mL or 2.0 mL sample cup. Place the sample cup in the assigned rack/cup position. Place the rack in the autoloader tray. Press the RUN button on the analyzer. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-16 May 2005...
Load rack in the Autoloader tray. Press the RUN button on the analyzer. UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-17...
• Enter the Rack number. • Enter the SAMPLE IDs. • Select <OK>. Load rack in the Autoloader tray. Press the RUN button on the analyzer. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-18 May 2005...
If provided, enter Demographics: • Select Demog [F2]. • Enter in the information provided. • To return to the Program Sample screen, select Next [F10]. (1 of 2) UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005 Page 3-19...
Page 46
If the sample does not have a readable bar code label: • Refer to "Processing Unreadable Bar Code Patient Samples" in this manual. To program additional samples, repeat Steps 2–8. (2 of 2) Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-20 May 2005...
Press the OFF power switch at the instrument. Press the OFF power switch at the console (i.e., computer). UniCel DxC 600/800 In-Lab Training Manual A24183 Procedures and Processes May 2005...
To safely restore power to the system follow these steps: Step Action Press the ON power switch at the console (i.e., computer). Press the ON power switch at the instrument. Procedures and Processes UniCel DxC 600/800 In-Lab Training Manual A24183 Page 3-22 May 2005...
Cal Time Left and Days Left. Request and process calibration for all ISE Chemistries, GLUCm and MG. Using the Cal Load List, determine Calibrator IDs and Calibrator Lot #/Level. UniCel DxC 600/800 In-Lab Training Manual A24183 Training May 2005 Page 4-1...
Page 50
Program and process 2 levels of Quality Control using bar code labels. Manually program and process 5 samples without bar code labels (included information: routine/stat, patient ID #, sample type, tests and demographics). (1 of 2) Training UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4-2 May 2005...
Page 51
Prepare the instrument to "clean flow cell and cups." Perform "CC Probe Cleaning" procedure. Determine Chloride’s Calibration Span and frequency for replacement of chloride’s electrode tip. (1 of 2) UniCel DxC 600/800 In-Lab Training Manual A24183 Training May 2005 Page 4-3...
Page 52
What is Beckman Coulter’s Call Center phone number? _____________________________ Describe the laboratory’s procedure for restarting the Host Communications. Status Tech’s Trainer’s Initials / Initials/Date Date View Current Temperature status. View Current Hydropneumatics status. Training UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4-4 May 2005...
9. Replace any reagent that is <10 tests remaining or has < 10% volume remaining. 10. Which MC reagent is replaced monthly? 11. What is the purpose of "No Foam" reagent? UniCel DxC 600/800 In-Lab Training Manual A24183 Training May 2005 Page 4-5...
Bar Code Label ID? 21. Process 2 levels of control for all general chemistries on the instrument. Samples 22. Describe a scenario wherein the PRIORITY LOAD button is used. Training UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4-6 May 2005...
28. Recall results for Sample ID # 2222. Print another report for this sample. 29. Request a re-run for GLUCm on Sample ID # 1111. For this exercise, inform the DxC analyzer that you had performed 1:2 dilutions on the sample. UniCel DxC 600/800 In-Lab Training Manual A24183 Training May 2005...
30. Give two differences between a "Beckman Coulter" chemistry and a "UDR" chemistry. Instrument 31. List the steps that you would perform (on the DxC Console) to Home the Commands instrument. Training UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4-8 May 2005...
15. IgG, IgA, IgM, and TRF require entering calibration parameters with every new lot of reagent. The parameter card is included with each reagent cartridge. The parameters may be entered manually or with the bar code reader. False True UniCel DxC 600/800 In-Lab Training Manual A24183 Training May 2005 Page 4-9...
31. List the steps that you would perform (on the DxC Console) to Home the Commands instrument. • At the DxC Console, select the Instrument Commands icon • Select Option 1: Home Training UniCel DxC 600/800 In-Lab Training Manual A24183 Page 4-10 May 2005...
Page 59