ILC Dover S-5000 Series User Instruction Manual

Sentinel xl papr cbrn system

Advertisement

SENTINEL XL™ PAPR CBRN SYSTEM
USER INSTRUCTION MANUAL
FOR MODEL NO. S-5000
∆ WARNING
IMPROPER USE OF THIS RESPIRATOR MAY RESULT IN PERSONAL INJURY OR DEATH.
IMPROPER USE INCLUDES, BUT IS NOT LIMITED TO, USE IN ATMOSPHERES
CONTAINING UNKNOWN CONTAMINATES AND/OR UNKNOWN CONCENTRATIONS OF
CONTAMINANTS, USE WITHOUT ADEQUATE TRAINING, DISREGARD TO THE
WARNINGS AND INSTRUCTIONS CONTAINED HEREIN AND FAILURE TO INSPECT AND
MAINTAIN THIS RESPIRATOR.
THE RESPIRATOR IS NOT INTENDED FOR USE IN ATMOSPHERES WHERE THE
IDENTITY AND/OR CONCENTRATION OF THE CONTAMINANT IS UNKNOWN. THIS MAY
RESULT IN IMMEDIATE DANGER TO PERSONNEL.
THIS RESPIRATOR DOES NOT SUPPLY OXYGEN AND MUST NOT BE USED IN OXYGEN-
DEFICIENT ENVIRONMENTS CONTAINING LESS THAN 19.5% OXYGEN.
S-2018
REV. D

Advertisement

Table of Contents
loading

Summary of Contents for ILC Dover S-5000 Series

  • Page 1 SENTINEL XL™ PAPR CBRN SYSTEM USER INSTRUCTION MANUAL FOR MODEL NO. S-5000 ∆ WARNING IMPROPER USE OF THIS RESPIRATOR MAY RESULT IN PERSONAL INJURY OR DEATH. IMPROPER USE INCLUDES, BUT IS NOT LIMITED TO, USE IN ATMOSPHERES CONTAINING UNKNOWN CONTAMINATES AND/OR UNKNOWN CONCENTRATIONS OF CONTAMINANTS, USE WITHOUT ADEQUATE TRAINING, DISREGARD TO THE WARNINGS AND INSTRUCTIONS CONTAINED HEREIN AND FAILURE TO INSPECT AND MAINTAIN THIS RESPIRATOR.
  • Page 2: Table Of Contents

    TABLE OF CONTENTS ITEM PAGE SAFETY PRECAUTIONS ......................... INTRODUCTION ............................DESCRIPTION............................GENERAL SAFETY INFORMATION ..................... REPLACEMENT PARTS .......................... PREPARATION FOR USE (USING ADJUSTABLE WAIST BELT OR DECON AND QUICK RELEASE BELT AND CLIP) ..............PAPR BACKPACK COMPONENT INTEGRATION ................PREPARATION FOR USE (USING BACKPACK ASSEMBLY) ............HEAD HARNESS SIZING INSTRUCTIONS ..................
  • Page 3: Safety Precautions

    SAFETY PRECAUTIONS The Danger, Warnings, Cautions, and Notes contained in this instruction manual have the following significance. ∆ DANGER Danger indicates an imminently hazardous situation, which if not avoided, will result in serious personal injury, serious illness or death. ∆ WARNING Warning indicates a potentially hazardous situation, which if not avoided, could result in serious personal injury, serious illness or death.
  • Page 4: Introduction

    INTRODUCTION This manual provides instructions for the proper use and care of the Sentinel XL™ PAPR CBRN. Before attempting to use the Sentinel XL™ PAPR CBRN in an emergency you must read, understand, and be able to follow all instructions as specified in this Sentinel XL™ PAPR CBRN Instruction Manual. A NIOSH approval label is included as part of this manual.
  • Page 5: Description

    DESCRIPTION The ILC Dover SENTINEL XL™ PAPR CBRN is a blower-assisted air-purifying respirator known as a Powered Air Purifying Respirator or PAPR. The SENTINEL XL™ PAPR CBRN is comprised of the...
  • Page 6 Duracell batteries need to be purchased to operate the Sentinel XL blower. Batteries have a long term shelf life and are easily accessible. Cartridge P/N S-2016 ILC Dover SENTINEL XL™ PAPR CBRN utilizes three NIOSH approved CBRN Cartridges that have been tested and approved to filter certain CBRN agents. Adjustable Waist Belt P/N S-2007 Supports the blower assembly on the user’s waist and is adjustable.
  • Page 7 ILC DOVER SENTINEL XL™ PAPR CBRN SYSTEM (WITH ADJUSTABLE WAIST BELT) OR DECON AND QUICK RELEASE BELT AND CLIP Alkaline Battery Adaptor Decon and Quick Release Belt and Clip Not to Scale - 6 - S-2018 REV. D...
  • Page 8 ILC DOVER SENTINEL XL™ PAPR CBRN SYSTEM (WITH BACKPACK AND ADJUSTABLE WAIST BELT) Alkaline Battery Decon and Adaptor Quick Release Belt and Clip - 7 - S-2018 REV. D...
  • Page 9: General Safety Information

    GENERAL SAFETY INFORMATION Intended Use: The ILC Dover SENTINEL XL™ PAPR CBRN blower can be mounted to the Adjustable waist belt, Decon and Quick Release Belt and Clip, or optional backpack, and is connected to the hood assembly via an integrated flexible breathing tube.
  • Page 10 INSTRUCTION AND OTHER APPLICABLE USER INSTRUCTIONS. ∆ WARNING BEFORE USING THE ILC DOVER SENTINEL XL™ PAPR CBRN SYSTEMS, EACH PERSON MUST READ AND UNDERSTAND THE INFORMATION IN THIS USER INSTRUCTION. USE OF THE RESPIRATOR SYSTEMS BY UNTRAINED OR UNQUALIFIED PERSONS, OR USE THAT IS NOT IN ACCORDANCE WITH THESE INSTRUCTIONS, MAY ADVERSELY AFFECT RESPIRATOR PERFORMANCE AND MAY RESULT IN INJURY OR DEATH.
  • Page 11 NIOSH APPROVAL For a listing of the components in the ILC Dover respirator systems (ILC Dover SENTINEL XL™ PAPR CBRN) that are NIOSH approved, refer to the NIOSH-approved label insert. NIOSH CAUTIONS AND LIMITATIONS Not for use in atmospheres containing less than 19.5 percent oxygen.
  • Page 12: Replacement Parts

    REPLACEMENT PARTS Part No. Component Cartridge S-2016 CBRN Toxic Chemical Cartridge (with Instructions) Battery Packs and Charger S-2003 Rechargeable NiMH Battery Pack Assembly S-2004 Non Rechargeable LiMnO2 Battery Pack Assembly S-4013 (Optional) Alkaline Battery Adaptor (Batteries NOT included) S-2009 Fast Battery Charger (with Instructions) Blower S-2002 Standard Blower Assembly...
  • Page 13: Preparation For Use

    PREPARATION FOR USE (USING ADJUSTABLE WAIST BELT OR DECON BELT) 1. Open the package/box and take out the components: • Blower S-2002 Battery S-2003 Rechargeable NiMH Battery Pack Assembly or S-2004 Non Rechargeable LiMnO2 • Battery pack Assembly Alkaline Battery Adaptor S-4013 (Optional) Batteries NOT included •...
  • Page 14 RESPIRATOR SYSTEM. BEFORE USE, THE USER MUST BE TRAINED BY THE EMPLOYER IN THE PROPER USE AND MAINTENANCE OF THE ILC DOVER SENTINEL XL™ PAPR CBRN SYSTEM. SUCH TRAINING MUST BE IN ACCORDANCE WITH THIS USER INSTRUCTION. EACH PERSON USING THESE RESPIRATOR SYSTEMS MUST FIRST READ AND UNDERSTAND THIS ENTIRE USER INSTRUCTION.
  • Page 15 7. Cartridge Installation. Verify that the Cartridges are the appropriate type for the application by examining the label attached to the cartridge. • Before removing the Cartridges, verify that the foil package is vacuum sealed and inspect the foil bag for any damage. Do not use the Cartridges if the foil bag has been damaged.
  • Page 16 ∆ WARNING BEFORE ENTERING A HAZARDOUS ATMOSPHERE WEARING THIS RESPIRATOR SYSTEM, THE RESPIRATOR SYSTEM MUST BE INSPECTED. COMPLETE AN AIRFLOW CHECK, AND DON THE SYSTEM ACCORDING TO THE INSTRUCTIONS IN THIS MANUAL. FAILURE TO DO SO MAY NOT ADVERSELY AFFECT RESPIRATOR PERFORMANCE AND MAY RESULT IN INJURY OR DEATH.
  • Page 17: Papr Backpack Component Integration

    PAPR BACKPACK COMPONENT INTEGRATION The Sentinel XL PAPR CBRN system is shipped with components pre-installed in the backpack. Use the following instructions for general storage or addition of replacement components in the backpack. 1. Open the package/boxes and take out the components: •...
  • Page 18 5. Insert the foil wrapped Cartridges into the three net receptacles on the backpack. 6. Undo the strap buckles and unfold the blower compartment cover. Close the blower compartment using the zipper (see picture below). The backpack is now in stand-by storage configuration.
  • Page 19: Preparation For Use (Using Backpack Assembly)

    PREPARATION FOR USE (USING BACKPACK ASSEMBLY) 1. Open the blower compartment using the zipper (see picture to the right). Blower compartment zipper 2. If using a rechargeable battery pack, remove the battery charger from the backpack. To charge the PAPR: •...
  • Page 20 ANY RESPIRATOR SYSTEM. BEFORE USE, THE USER MUST BE TRAINED BY THE EMPLOYER IN THE PROPER USE AND MAINTENANCE OF THE ILC DOVER SENTINEL XL™ PAPR CBRN SYSTEM. SUCH TRAINING MUST BE IN ACCORDANCE WITH THIS USER INSTRUCTION. EACH PERSON USING THESE RESPIRATOR SYSTEMS MUST FIRST READ AND UNDERSTAND THIS ENTIRE USER INSTRUCTION.
  • Page 21 ∆ WARNING DO NOT OVER TIGHTEN CARTRIDGE. OVER TIGHTENING MAY RESULT IN DISTORTION OR DISPLACEMENT OF THE SEAL AND MAY ALLOW CONTAMINATED AIR TO ENTER THE RESPIRATOR HEADPIECE. THIS MAY RESULT IN INJURY OR DEATH. 6. Adjust the backpack’s waist belt length to your waist size when dressed in your protective garments. 7.
  • Page 22 8. Take the hood out of the designated hood Compartment. Remove hood from fabric bag. 9. Connect the Hood’s Breathing Tube to the Blower. • Confirm that the breathing tube thread has the black rubber gasket seal present. • Confirm that the thread in the blower unit is in good condition and clear of foreign matter. •...
  • Page 23: Head Harness Sizing Instructions

    Head Harness Adjustment and Donning the Butyl Hood Note: If possible, the ‘buddy system’ should be utilized to aide in effective and fast donning of the system. A second person can help the wearer by insuring that the head harness is adjusted and donned properly, and to check proper positioning of the hose and hood skirt.
  • Page 24: Head Harness Replacement Instructions

    HEAD HARNESS REPLACEMENT INSTRUCTIONS STEP 1 STEP 2 D-ring Hood Slot Slide the head harness webbing through the Slide the black webbing through the D-ring slot. hood slot; make sure the webbing is not twisted and is able to lie flat. Medium Hood Slot Small...
  • Page 25 STEP 5 Female snap STEP 6 Male snap Male snap Female snap Center the head harness in the hood slot after Fasten the remaining male loose snap making the sizing adjustment. Fasten the male (located next to the D-ring) of the head loose snap (opposite side of the D-ring) of the harness into the female end located on the head harness into the female end located on the...
  • Page 26 • The pictures below illustrate how to incorrectly and correctly cinch the drawstring cord. Incorrect Correct Incorrect • Tuck the inner skirt under your protective clothing and allow the outer skirt to hang outside. • With the respirator in operation and donned according to the instructions, enter the contaminated area, breathing normally.
  • Page 27: Decontamination Procedures

    IF YOU HAVE ANY DOUBTS ABOUT THE APPLICABILITY OF THE EQUIPMENT TO YOUR JOB SITUATION, CONSULT AN INDUSTRIAL HYGIENIST OR CALL THE TECHNICAL SERVICE DEPARTMENT OF ILC DOVER at 1-800-631-9567. DO NOT REACH YOUR HAND INTO THE HOOD IN AREAS WHERE THE AIR IS CONTAMINATED.
  • Page 28: General Cleaning

    GENERAL CLEANING Solvents should not be used to clean the PAPR blower unit, battery pack, waist belt, hose and hood. Liquid solvents may chemically weaken the plastic. Wipe the PAPR blower unit, battery pack, waist belt, hose and hood with a mild cleaning solution. The decontamination solutions may be applied via cloth, paper towels, or spraying methods.
  • Page 29 Cartridge. ∆ WARNING IF RUBBER GASKETS APPEAR TO BE POSITIONED IMPROPERLY WITH TWISTS, RAISED AREAS, OR SURFACE DEFECTS, DO NOT USE BLOWER. CONTACT THE ILC DOVER CUSTOMER SERVICE DEPARTMENT IMMEDIATELY.
  • Page 30: Changing Battery Pack

    CHANGING BATTERY PACK (P/N S-2003 and P/N S-2004) 1. Make sure the blower is turned off. Latch 2. Open the battery pack compartment by pressing Connector on the latch connecting it to the blower. Battery Pack Battery Pack Seal 3. Disconnect the connector parts and separate the battery pack from the blower.
  • Page 31: Installing The Alkaline Battery Adaptor

    INSTALLING THE ALKALINE BATTERY ADAPTOR (P/N S-4013) 1. Make sure the blower is turned off. 2. Open the battery pack compartment by pressing on the latch connecting it to the blower. 3. Disconnect the battery wire connector from the blower and separate the battery from the blower.
  • Page 32 6. Press the adapter plate assembly firmly to ensure the white Velcro hook adheres to the inside wall of the blower. Then gently pull the blue fabric strap to pull the plate assembly out of the unit. The Velcro hook (color white) should remain adhered to the inside of the blower.
  • Page 33 9. After batteries are placed into the Alkaline Battery Adaptor install the part into the back of the blower. The pull strap should remain outside as shown below, and is used to remove the alkaline battery pack for exchange of battery cells or replacement with alternative battery options. Pull Strap 10.
  • Page 34: Removing The Alkaline Battery Adaptor

    12. Install the back plate included in the Alkaline Battery Kit by aligning the cover hinge to the hinge pin of the blower. Ensure that you do not pinch the battery leads or fabric loop when closing the battery pack. 13.
  • Page 35 ∆ WARNING FOR ALKALINE BATTERY ADAPTOR OPTION, P/N S-4013, USE ONLY DURACELL BATTERIES. DO NOT USE A COMBINATION OF DURACELL OR OTHER BRANDS. IT IS IMPORTANT THAT ALL FIVE BATTERIES ARE DURACELL D-CELL BATTERIES. WHENEVER BATTERIES ARE CHANGED, CHANGE ALL FIVE D-CELL BATTERIES AT THE SAME TIME.
  • Page 36: Storage

    Remove plug and install third CBRN filter. PAPR blower malfunction. Switch to a different blower unit. Blower gasket is raised, Gasket is not properly seated Contact ILC Dover Customer Service twisted, or has any into the blower and may produce Department immediately visible surface defects...
  • Page 37: Customer Service Information

    NO. The Breakthrough is determined by which quantity, NO or NO, reaches breakthrough first. CUSTOMER SERVICE INFORMATION If you need additional information on the Sentinel XL™ PAPR CBRN or any ILC Dover product, consult your distributor or contact: Customer Service ILC Dover...
  • Page 38: Warning

    WARRANTY AND LIMITATION OF LIABILITY 1. ILC DOVER warrants all of its products against defects in workmanship, materials, and construction under normal use and service for the following periods of time and subject to the following conditions and limitations: a. Promptly after receipt of each shipment of product, the BUYER shall inspect the products. All claims for shortages and visible, in-transit damage must be made within ten (10) days after receipt of the shipment by the BUYER.
  • Page 39: Warning

    In no event shall the liability of ILC Dover exceed the purchase price of the product, and, in no event, including the case of latent defects, negligence or strict liability, shall ILC Dover be liable for incidental or consequential damages.
  • Page 40 - 39 - S-2018 REV. D...