1. Introduction
1.1. Generalization
The Meter is a battery-operated process maintenace field device which combined both digital-parameter function with
process-signal source together.
Functioning as follows:
Measurement function of TC voltage, DC voltage , Ohms, DC, Continuity and Diode; Data reserving function;
Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.
Output function of DC (Constant, Manual-Step and SIMULATE modes)
Loop-circuit testing function; Powering the loop circuit with 24 Voltage and testing the current; Built-in 250Ω HART
loop resistance
Safety: complied with IEC 61010-1 safety requirements for measurement, control and electronic instruments for lab
use.
1.2. Opening the Meter Case
Check the product to see whether it is damaged in the shipment or not. Check the materials to see whether they are the
same as shown in the packing list. Keep the packing materials for late delivery.
Standard and chosen accessories supplied are listed as follows. Chosen ones are bought at purchaser's options.
Standard accessories:
1x testing wire(with alligator clips)
1x user's manual
2x 1.5V (LR6)Alkaline batteries
2x 63mA/250V Fast Fuses
1.3. Safety Information
The design, manufacture and test of the Meter reach the IEC 61010-1 Safety Requirements. This Manual contains all
warnings and safety regulations that must be followed to ensure safe operation and retain the meter in safe condition.
Read the following instructions before operation.
Mark
in the Meter means the operator needs to make reference to related parts in the Manual to ensure safe operation.
Warning - indentifies conditions and actions that pose hazard(s) to the user; Caution - indentifies conditions and actions
AX-C705

Advertisement

loading
Need help?

Need help?

Do you have a question about the AX-C705 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for AXIOMET AX-C705

  • Page 1 AX-C705 1. Introduction 1.1. Generalization The Meter is a battery-operated process maintenace field device which combined both digital-parameter function with process-signal source together. Functioning as follows: • Measurement function of TC voltage, DC voltage , Ohms, DC, Continuity and Diode; Data reserving function;...
  • Page 2 that may damage your meter. Note - indentifies symbols from which you can understand operation and features of your meter. For related international symbols used by the Meter and the Manual, please read table below for reference. Warning • Do not use the meter if it is damaged. Before using the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
  • Page 3 Fuse Safety rules CAT III - Overvoltage of class three, secondary pollution (according to IEC61010) refers to the level of protection for the pulse compression. Typical installation location includes: distribution of three-phase circuit (including a single commercial lighting circuit) position-fixed equipment; interior lighting equipments in large buildings, lines; industrial field devices.
  • Page 4 1.6. Using display hold mode Under output mode, the Meter freezes data displayed in the reading part when pressing <BACK> button to enter into display hold mode (screen displays symbol). Repress <BACK> button to exit from display hold mode. 2. Operation Instructions 2.1.
  • Page 5 • The screen displays ”OL” if the open circuit or resistance of the device under measurement exceeds the maximum range of the Meter. • Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA. •...
  • Page 6 current value //Displayed In percentage form -22.000mA //162.50% 0.000mA // 25.00% 4.000mA // 0.00% 20.000mA // 100.00% 22.000mA // 112.50% 2.8. Measuring loop-circuit current The function may measure current value powered with 24VDC constant voltage. 24V loop-circuit measurement function may test transmitter loop circuit. The meter may be connected to transmitter, but not to transmitter or signal adaptor. Warning Typical value of loop power supply is 24VDC.
  • Page 7 Do not apply any voltage to output terminals except in SIMULATE mode. Any wrong application may damage internal circuit. 2.11. Constant current output (SOURCE mode) • Turn the rotary switch to “”; the screen displays “OUTPUT” and “LOOP POWER”; output is set as 0mA. •...
  • Page 8: General Maintenance

    3. General Maintenance 3.1. Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents. • Take out the batteries if the meter won’t be used for a long time. • Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA. •...
  • Page 9: Specification

    To avoid personnel injury or damage to the meter, use only the specified fuse. The specification is 6 3mA 250V fast-melt. 3mA 250V-fuse protection is in the mA current output terminal. To check whether fuses are blown: • Turn the rotary switch to mA •...
  • Page 10 Automatic power-off - The default value is 5 mins if no operation which is adjustable. Preheating time - 10min Closed – case calibration No internal adjustments needed Battery access door - Battery or fuse replaceable without voiding calibration Measurement - 180(L)×90(W)×47(D)mm Weight - about 500g Calibration period - one year 4.3.
  • Page 11 4.4. Outputting detailed accuracy indicator Output function - Range - Output designated range - Frequency - Accuracy - Remark DCI DC current - 20mA - 0.000~22.000mA - 0.001mA - 0.2%+4 • Max. load: 1kΩ at 20mA SIMULATE Simulate transmitter - -20mA - 0.000~-22.000mA - 0.001mA - 0.2%+4 •...