Table 6.23 Draining your ultrasonic bath
Step
3
4
5
100-214-294 REV. 05
Hand tighten the drain valve onto the drain pipe over the white teflon sealing
tape. Finish tightening the valve in place using an adjustable or a 21mm
wrench. Tighten the valve no more than one full turn when using the wrench
until the handle is on top.
CAUTION
Over tightening of the valve can cause damage to the ultrasonic tank. Always
use teflon sealing tape or a sealing paste designed for use with stainless steel
if retightening or refitting of the drain valve is required.
Hand tighten the hose adaptor into the end of the drain valve. Slide the drain
tube over the barbed hose adaptor end.
Close the drain valve by turning the handle perpendicular to the valve body
and the unit is ready to fill with solution. To open the valve and drain the
tank, turn the handle so that it is in line with the valve body.
Action
55