Emerson Branson M1800 Instruction Manual page 52

Ultrasonic bath
Table of Contents

Advertisement

Table 6.15 CPXH Series explanation of controls.
Control
44
Sonics
Press to activate ultrasonics. If running in Timed Mode, a timer will begin to
count down and ultrasonics will stop at 0 minutes. In Constant Sonics Mode
(Constant Sonics icon and "- - -" on the display), timer has no function.
Press Sonics key again to deactivate ultrasonics.
If running in Timed Mode, press Up and Down keys to adjust the ultrasonic
cycle time (adjustable from 1 to 99 minutes).
Degas
Press to degas the solution or to run a degas application. The degas timer
will begin to count down from its current setting and degassing will stop at
0 minutes.
Press Degas key again to stop degassing the solution.
During a degas cycle, press Up and Down keys to adjust degas time
(adjustable from 1 to 99 minutes).
NOTICE
Refer to
6.5.11 Solution
Auto
Press to begin an auto cycle. In Auto Mode, the following actions are
carried out automatically by the controller:
• Heater is turned on to bring bath to set temperature
• When set temperature is reached, ultrasonics are activated. The unit will abort
the auto cycle and flash the AUTO icon if set temperature is not reached within a
120-minute period
• When ultrasonics timer reaches 0 minutes, the auto cycle is finished
If at any point during an auto cycle the degas key is pressed, a degas cycle
will begin. If ultrasonics has already started, the ultrasonics timer will
restart after the degas period.
Function
for information on degassing.
100-214-294 REV. 05