Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of these Clinical IR thermometer
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Radiated RF IEC
61000-4-3
Conducted RF
IEC 61000-4-6
IEC 60601
Compliance
Test level
level
�6 kV
contact
Complies
�8 kV air
3V/m
80 MHz to
Complies
2.5 GHz
Not
Not
applicable
applicable
Electromagnetic
environment - guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Field strengths from
fixed RF
transmitters, as
determined by an
electromagnetic site
survey, should be less
than 3V/m. Interference
may occur in the vicinity
of equipment marked
with the following
symbol:
Separation distance
calculation provided
below. If a known
transmitter is present,
the specific distance can
be calculated using the
equations.