WEIGHING INDICATOR
SERIE IPE 50
User manual 1/2
For calibration and basic function
See user manual 2/2 for special functioning mode and output RS232/485, this manual is available on our Web Site
http://www.lorenz-messtechnik.de/english/products/lc_weighing-electronics/ipe50.php
(valid from version 07.03)

Advertisement

Table of Contents
loading

Summary of Contents for LORENZ MESSTECHNIK IPE 50 Series

  • Page 1 WEIGHING INDICATOR SERIE IPE 50 User manual 1/2 For calibration and basic function See user manual 2/2 for special functioning mode and output RS232/485, this manual is available on our Web Site http://www.lorenz-messtechnik.de/english/products/lc_weighing-electronics/ipe50.php (valid from version 07.03)
  • Page 2 INDEX INTRODUCTION MAIN TECHNICAL SPECIFICATIONS DIMENSIONS POWER SUPPLY & START-UP FRONT PANEL KEYS AND INDICATORS BASIC FUNCTIONS SELECTABLE OPERATING MODES 7.5 Setpoint adjustment INSTRUMENT MESSAGES WHILE IN USE BLOCK DIAGRAM SETUP ADJUSTMENT ANALOG OUTPUT ERROR MESSAGES...
  • Page 3 1 INTRODUCTION The purpose of this manual is to help the user get to know the weight indicator’s various functioning modes, the keys’ functions and the display indications We advise to carefully follow the instructions for programming the weight indicator; by taking actions not indicated this manual, one could cause the scale to not work properly.
  • Page 4: Main Technical Specifications

    2. MAIN TECHNICAL SPECIFICATIONS POWER SUPPLY 12 to 24Vdc / 3.6 W max (110/220Vac for IPE50XLI) OPERATING TEMPERATURE From -10 to +40 °C (14 to 104 °F). DISPLAYED DIVISIONS 10000e, 2X3000e for legal for trade use expandable to 800.000 for internal use (with minimum signal coming from the 1.6mV/V cell).
  • Page 5 3. DIMENSIONS : IPE50 PANEL 1 and 2 Connection of power supply (12 to 24Vdc) connection of the 2 inputs, the contact of setpoints and RS232 and 485 connection of sensors connection of the 2 to 6 contact relays (6 relays in option)
  • Page 6 IPE 50 DIN 1 and 2 Connection of power supply (12 to 24Vdc) connection of the 2 inputs, the contact of setpoints and RS232 and 485 connection of 1 to 4 sensors...
  • Page 7 IPE50 XLI 1 : gland for power supply connection 110/220Vac 2 : gland for sensor connection...
  • Page 8 4. POWER SUPPLY & START UP Do not connect other equipment to the same socket as the one that the adapter is in. Do not step on or crush the power supply cable Power supply for IPE50 is from 12 to 25Vcc (except IPE50XLI) The display shows in sequence: 07.xx indicate the version of software if you press the key ->0<-...
  • Page 9 FONCTIONS Indicates that the weight detected on the weighing system is near zero, ->0<- within the interval of –1/4 to +1/4 of the division. Indicates that the weight is unstable. Indicates that the displayed weight is a net weight. Indique que l'IPE50P est dans un mode de fonctionnement particulier W1/ SP1 ou W1 or W2 show the functioning with 2 weighing ranges W2 / SP2...
  • Page 10: Basic Functions

    6 BASIC FUNCTIONS Tt will be necessary to go to different menus to modify some parameters. To enter into these menus , turn on the unit, press the ->0<- key when 07.xx is displayed, after 09.01, DGT and the version of the software are displayed, press ->T<- key during the countdown the display shows the first parameter “TyPE”.
  • Page 11 In case of MANUAL TARE or FROM DATABASE the net weight before unloading the scale must be of at least 2 stable divisions. To set the type of tare: - Select menu F.Mode. - Press ↵ to enter the menu. - Press ->0<- many times (to scroll ahead through the parameters) or ->T<- (to scroll backwards) until one finds the “tArE t”...
  • Page 12 tot S Vertical totalizer (please refer to user manual 2/2) Coun Counting (please refer to user manual 2/2) Confirm with ↵ if one has selected the inout, tot or, tot S or Coun mode, one will be asked to select another operating parameter;...
  • Page 13 1 GroS : setpoint on the gross display 2 nEt : setpoint on the net display 3 PCS : setpoint on the counting mode 4 Gro.O : the setpoint will be actionned when the display will be at 0 in gross value 5 nEt.O : the setpoint will be auctioned when the display will be at 0 in net value 6 Moti.
  • Page 14 The DISABLING SET POINT must be equal or less than the ENABLING one; If in the DISABLING SET POINT one enters and confirms a value greater than the ENABLING one, the instrument will automatically set 0 value and wait for a correct value. If in the ENABLING SET POINT one enters and confirms a value greater than the max capacity, the instrument will automatically set 0 value and wait for a correct value.
  • Page 15: Instrument Messages While In Use

    8 INSTRUMENT MESSAGES WHILE IN USE MESSAGE DESCRIPTION Message is displayed when the display is higher than the max capacity programmed in ¯ ¯ ¯ ¯ ¯ calibration mode Message displayed when the display is lower than the zero memorized in calibration mode - - - - - - In Alibi mode, when you switch on the unit, the unit detects a problem with the alibi card, AL.
  • Page 16: Block Diagram

    9 BLOCK DIAGRAM P 19 Only XLI version Only with option date/time or If DSD otpion Diagnosic mode Output relays anaolog Output...
  • Page 17 1.0, 1.00, 1.000, 100000 DECi Adjustment / kg, g, lb, t u.M. calibration rAnGE 1 rAnGE 2 Quick calibration of zero No operating Man.CAL Mod.Pnt WEiGht SErIAL You can download it on our WEB SITE Serial connection www.lorenz-messtechnik.de/english/products/lc_weighing-electronics/ipe50.php See user manual 2/2 for PC, printer….
  • Page 18 DESCRIPTION OF THE STEPS TYPE Ind.Ch. Standard using for 1 to 4 independent channels (selection of the channel displayed, press the key "MODE" 3 sec then select the channel with key "->T<-" and validation with key " ") ↵ DEP.Ch. Using for the connection of 2 to 4 identical sensors on a same system TrAnSM.
  • Page 19 instability” or “always”. ZEro passage of the net weight by zero inSt instability ALWAyS always (!) Zero L.int DISPLAY INTENSITY (IPE50 XLI only) Lint 1 (minimum) to Lint 4 (maximum) (!) Lint1 CLoCK DATE/TIME ADJUSTMENT (OPTIONAL) ↵ In this step one sets the date and time of the indicator; by pressing one is asked to enter, in this order, the day, month, year, hour and minute.
  • Page 20 FF. 400 can be used only in transmission mode via RS port Custom reserved for factory, don't used it The higher the filter value is, and greater is it intervention relative to the type of filter used. (!) FLt 3 Nota: the filter FF.200 is not suitabable in the mode DEP.Ch.
  • Page 21 See paragraph “ADJUSTMENT” section. (*) 0CALib. QUICK ZERO CALIBRATION / ADJUSTMENT See paragraph “CALIBRATION” section. NOTE: If the indicator is approved, the step is not displayed. SeriAL (SERIALS, PRINTOUTS, ETC…) Please refer to user manual 2/2 dSP.rF norm (keep this parameter) 1 h2, 2.5 h2, 5 h2, 10 h2, 20 h2 (!) norm dSALE Do not enter in this step...
  • Page 22 27 K.Zer : When you press the key "->0<-" the contact will be closed during approximatively 2 seconds 28 K.tAr : When you press the key "->T<-" the contact will be closed during approximatively 2 seconds 29 Err : the setpoint will be on if the display shows the message _ _ _ _ _ (overload) or ¯ ¯ ¯ ¯ ¯ (underload) rEL.1 to rEL.6 : selection of setpoint 1 or 6 for programmation of the parameters (6 relays in option for IPE50 Panel only) no/nC : no normaly openned or nC normaly closed...
  • Page 23 diV.int CALIBRATION INTERNAL DIVISIONS OF CALIBRATION By pressing the instrument shows the calibration internal divisions when the unit has been adjusted. Press ↵ “C” to exit this step. AdC.uU MICROVOLTS By pressing the instrument shows the output signal of the loadcell (in microvolts) relative to the weight on ↵...
  • Page 24 Furthermore the ASCII “TEST”<CRLF> string is continuously transmitted on both the serial lines. CtS.St TEST OF THE CTS STATUS By pressing one views the status/level of the CTS signal of the printer (on) connected to the PRT serial port. ↵ outPut TEST OF THE I/O EXPANSION BOARD RELAYS By pressing the instrument displays “rEL.1”...
  • Page 25 11 ADJUSTMENT Nota : The manufacter parameters of the IPE50 unit are : 2mV/V -> 10.000 per 0.001 Attention : the default calibration parameters will probably not match with your application requirements. You have to calibrate the system according to your needs. Depending on the application, one can carry out two types of calibration: the one for the simple scale or the one for the pallet truck.
  • Page 26 Type : TranSm idem that ind.Ch, except that the keys, the zero tracking and the autozero at the startup are disable and it is possible with one serial command, to receive the values of all the channels operating. Type : ind.Ch. (for DEP.Ch.
  • Page 27 16) If a calibration point has been set, once the weight acquisition has been made, the display shows for an instant the value of the internal divisions and then the “ntP” step. If there are various calibration points, repeat the operations for the “ddt2”, “tP2”, “ddt3”, “tP3” points.1 17) Once the calibration has been made of all the necessary points, press the C key various times until the indicator shows “SAVE? in the display: confirm with...
  • Page 28 Eq1 blinks and after the unit will pass at the next step automatically. 15) Eq2 is displayed to indicate the adjustment of the loadcell which is connected to the channel 2. Apply the load on loading point 2 (over loadcell 2) and press ↵...
  • Page 29: Quick Calibration Of Zero

    If no number after the coma has been selected, you could set value from -99999.9 to 99999.9 If 1number after the coma has been selected, you could set value from -9999.99 to 9999.99 If 2 numbers after the coma has been selected, you could set value from -999.999 to 999.999 If 3 numbers after the coma has been selected, you could set value from -99.9999 to 99.9999 Press ↵...
  • Page 30: Analog Outputs

    12 ANALOG OUTPUTS This IPE50 has an analog output (0/10V or/and 4/20mA) which is automatically calibrated for the higher value (10V or 20mA) of the max capacity selected in rAnGE 1.These outputs correspond to the displayed value. You must adjust this output through the step AoMAX and AoZEr (see step after) It works only on a positive display if the step SiGn = PoSit (if the display is negative, the output will be the minimum value set at Ao ZEr step), It works only on a negative display if the step SiGn = nEGAt (if the display is positive, the output will be the minimum value set Ao ZEr step).
  • Page 31: Error Messages

    13 ERROR MESSAGES While using the indicator, it is possible to incur in the following errors: MESSAGE DESCRIPTION Analog/digital convertor doesn't work. Check the wiring of the loadcells. Measure the signal ADC Err output of the loadcells between the pins + and - sig, the tension must not exceed 15mV. It is displayed if one tries to calibrate a point without first having confirmed the number of PREC.