5 CHECK, WARRANTY AND TROUBLESHOOTING

Annual checks and warranty

The product must always be repaired and maintained by qualified personnel. It is the legal
owners' duty to ensure that these checks are carried out on the product.
The following items is to be inspected during the annual check:
Seat tilt: Actuator runs both ways. Piston is lubricated. Actuator is mounted correct.
Locking rings are placed correct. O-ring is intact, lubricated and correct inserted.
Cables: Intact.
Backrest and clips: Intact and securely tightened.
Seat: Easy to remove and put back. PU surface is intact.
Arm rests and moving parts: Securely fastened, lubricated and moves easy up and
down. PU surface is intact. screws tightened (remember: Lock-Tight).
Footplates: Works up and down. Height adjustability works. Easy to remove.
Emergency stop: Works and releases again.
Lifting mechanism: Actuator runs both ways. Piston is lubricated. Actuator is
mounted correct. Locking rings are placed correct. O-ring is intact, lubricated and
correct inserted. Lifts from bottom to top with max SWL.
Hand Control: All functions must be checked.
Charging station: Intact. Lights works. charge function works.
Control Box: Correctly mounted. No visible defects.
Wheels: Wheels and brakes must be checked. Securely fastened.
Frame and Weldings: Must be checked for visual signs of cracks or deformations with
max. SWL.
If it is necessary to replace any parts due to wear or damage, Winncare Nordic ApS
recommends for safety reasons, to use original spare parts.
Warranty
There is a 2 years warranty on the chair, if the conditions of warranty are followed.