pH Flexible LC
Gemini HPLC/UHPLC Columns
Tips for Care and Use
General Information
Each Gemini column manufactured by Phenomenex is individually prepared and tested. Every column is supplied with a
Certificate of Quality Assurance (CQA) which indicates testing conditions, operating parameters, and column details. The
column details, including specifications and performance test results should be entered into your information management
system for easy tracking and reference. Electronic copies of your column's quality documentation can also be acquired at:
www.phenomenex.com/mysupport.
Inspection
Upon receipt of column, please verify that the column you received is the one you ordered (i.e. dimensions, particle size, media).
Additionally, please check the column for any physical damage potentially caused during shipment. Test the column immediately to
verify performance and record the result of your test in your column information management system.
Column Characteristics
Gemini
Phases
Shipping Solvent
†
C18
Acetonitrile / Water (65:35)
NX-C18
Acetonitrile / Water (75:25)
C6-Phenyl
Acetonitrile / Water (65:35)
†
Depending on column dimensions, organic to aqueous solvent ratios may vary slightly.
* pH stability under gradient conditions. pH stability is 1.5-10 under isocratic conditions.
Typical Flow Rate, Backpressure, Temperature:
Here are some typical values for common dimensions of Gemini HPLC and UHPLC columns. These numbers are not absolute values
and can differ based on LC system, running parameters and sample analytes/matrix. The values below have been created using a
solvent system of Acetonitrile and Water.
Particle
Internal
Typical Flow
Size (µm)
Diameter (mm)
(mL/min)
3
0.3
0.005
3
0.5
0.015
3
1.0
0.05
3
2.0
0.2
3
3.0
0.5
3
4.6
1
5
2.0
0.2
5
3.0
0.5
5
4.6
1
5
10.0
5
5
21.2
20
5
30.0
40
10
4.6
1
10
10.0
5
10
21.2
20
10
30.0
40
10
50.0
50
** Column Length
Phenomenex products are available worldwide.
www.phenomenex.com/mysupport
Particle
Pore
Surface
Sizes
Size
Area
(µm)
(Å)
(m
3,5,10
110
375
3,5,10
110
375
3,5
110
375
Typical Pressure (PSI)
50 mm**
150 mm**
250 mm**
500
1000
NA
700
1400
NA
800
1000
NA
650
1550
NA
950
1900
2700
1000
1700
2600
400
650
850
550
1100
1550
1400
1500
1950
1400
1600
1800
500
800
1400
500
900
1200
NA
NA
1400
NA
NA
1400
NA
400
500
NA
400
600
NA
300
300
Carbon
Load
Reversed
/g)
(%)
pH Stability
Phase
2
14
1.0-12.0*
14
1.0-12.0*
12
1.0-12.0*
Typical Flow Rates (Independent of particle size):
i.
1.0 mL/min for 4.6 mm ID
ii.
0.2-0.6 mL/min for 2.1 mm ID
Maximum Backpressure:
iii.
> 5,000 psi (345 bar) may compromise column longevity.
Maximum Temperature:
iv.
Recommended max temperature for Gemini LC columns is
60°C, however temperature limits are dependent on your
running parameters. Running at a pH greater than 8 at 60°C will
compromise column lifetime.
v.
Continuous use of Gemini columns at the maximum
temperature limit may compromise column longevity.
100%
Normal
Aqueous
Phase
HILIC
Stable
QUALITY
MANAGEMENT SYSTEM
CERTIFIED BY DNV GL
9001:2008
Need help?
Do you have a question about the Gemini C18 and is the answer not in the manual?
Questions and answers
how to column regenrate
To regenerate a Phenomenex Gemini C18 column for reversed-phase use, follow these steps:
1. Apply the same gradient flush as used in the cleaning procedure.
2. Perform the flush overnight at a low flow rate.
3. Reverse flushing is acceptable.
This process helps restore column performance by removing contaminants.