CARBON MONOXIDE MONITOR
O. M. 23301
DATE OF ISSUE:
REVISION:
Do not use this
equipment until you
have READ this
MANUAL and YOU
UNDERSTAND its
contents. *
These WARNINGS are
included for the health
and safety of the
operator and those in
the immediate vicinity.
*If you are using a Clemco Distributor
Maintenance and Parts Guide, refer to
the orange warnings insert preceding
the Index before continuing with the
enclosed instructions.
Electronic files include a Preface
containing the same important
information as the orange cover.
CMS 1
05/01
F, 09/17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CMS 1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Clemco CMS 1

  • Page 1 © 2017 CLEMCO INDUSTRIES CORP. One Cable Car Dr. *If you are using a Clemco Distributor Washington, MO 63090 Maintenance and Parts Guide, refer to Phone (636) 239-4300 Fax (800) 726-7559 the orange warnings insert preceding Email: info@clemcoindustries.com...
  • Page 2 Clemco manufactures. Users must contact each Abrasive Blasting. manufacturer and supplier of products used in the blast job for Clemco Industries Corp. One Cable Car Drive Washington MO 63090 warnings, information, training, and instruction relating to the Tel: 636 239-4300 — Fax: 800 726-7559 proper and safe use of their equipment.
  • Page 3 Numerous topics associated with the hazards of crystalline silica in silica blasting sand can be found on http:// osha.gov/. Clemco urges users of silica blasting sand to visit this website, and read and heed the information it contains.
  • Page 4 Exceeding the maximum working pressure. Clemco blast machines are built to ASME-code and carry a ‘U’ or ‘UM’ stamp, and National Board/serial number. Every machine is marked with its maximum working pressure. Never exceed the maximum working pressure limits of the blast machine.
  • Page 5 Always ground the machine to eliminate hazard of static shock. Install a blast machine screen to keep out foreign objects. Use a blast machine ©Clemco Industries Corp., Stock No. 20954P, 0692 Rev. F, 06/12 cover if left outdoors overnight. Never exceed the maximum working pressure of the...
  • Page 6: Table Of Contents

    CMS-1 monitor manufactured SUBJECT SECTION LOCATION Invertech Inc., for Clemco Industries Corp. The monitor INTRODUCTION ............1.0 is a potential life saving instrument. To assure its Scope of Manual ............1.1 performance, the monitor must be properly installed, Safety Alerts ............... 1.2 tested, calibrated, and maintained.
  • Page 7: Components

    120 V Power Cord Annunciator PPM CARBON MONOXIDE lights Run/Calibration Switch Test Gas Port Display Window Flow Meter Calibration UP Adjustment Zeroing Potentiometer Calibration DOWN Adjustment Arm/Disarm Switch (audible alarm) Figure 1 © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 8: Digital Display Window

    RUN position during the operating mode (all can cause death from the inhalation of carbon operations other than the actual calibration test and monoxide. calibration process). If the toggle is not in the RUN © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 9: Optional Accessories

    Refer to Section 2.3.2. The operating range. Pressure above 145 psi could damage monitor must be positioned upright and level to ensure the monitor. the flow meter responds correctly. Place the monitor on © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 10: Connecting To Respirator Air Supply

    2.3.2.5 Attach the urethane tubing between the air Electrical power may be rewired by a qualified electrician supply tube-adaptor fitting and monitor inlet-tube adaptor with conduit that conforms to applicable codes. fitting. Cut-off excess tubing. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 11: Ac To 12-Volt Dc Field Conversion

    Open the case cover and insert a straightened 2.4.2.3 Remove the four faceplate screws, and carefully paper clip into the "ARM" port in the faceplate. While remove the faceplate, being careful not to disconnect © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 12: Connecting An External Alarm And Shutdown Device

    Typical wiring for normally open alarm Figure 5 3.2.4 Make sure alarms function per Section 6.4. If the alarms function as described, proceed to place the instrument in operating mode, per Section 3.3. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 13: Operating Mode

    Failure to heed this warning 4.1.1 It is important to understand the difference can cause death from the inhalation of carbon between calibration testing and the calibration process. monoxide. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 14: Calibration Test

    Figure 6 Sections 4.2 and 4.5, and only if the digital display is other than 09, 10, or 11 (10 PPM plus or minus one). © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 15: Return Monitor To Operating Mode

    Calibration Connector ..... Stock No. 23011 to the operating mode. This test shows whether the Medium size screwdriver alarm condition is due to contaminated air, or a Miniature screwdriver malfunctioning monitor. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 16 Test gas is entering the unit when the flow meter ball rises. If the ball does not rise when the calibration valve is opened, the test gas cylinder is probably empty. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 17: General Maintenance

    Section 9.0. Contact 7.1.12 Apply sample air and power per Section 3.0. a Clemco distributor for authorization to return the Run the monitor in operating mode for at least four hours instrument for evaluation or service.
  • Page 18: Fuse Replacement

    8.2.4 Check internal instrumentation tubing for breaks, kinks, or disconnection. If tubing has come loose, reconnecting it will re-establish function and the monitor should operate correctly. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 19: No Annunciator Lights Or Alarms

    (calibration) position. This will cause intermittent alarm, and should register on the voltmeter each time the alarm activates. A monitor with faulty circuit board should be returned for service. © 2017 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 23301, Rev. F...
  • Page 20: Accessories And Replacement Parts

    ....29261 CMS-1, 220-volt CO monitor package Power supply, 120/220 AC to 12-v. DC ..28011 Special Order ..Contact Clemco Distributor Sensor (not shown) ........23015 Orifice, .004 purple restrictor (not shown) ... 24423 Orifice, .006 red restrictor (not shown) ..