Onicon FT-3100 Programming Manual

Inline electromagnetic flow meter

Advertisement

FT-3100 Transmitter
Inline Electromagnetic Flow Meter
Programming Guide
ONICON

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FT-3100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Onicon FT-3100

  • Page 1 FT-3100 Transmitter Inline Electromagnetic Flow Meter Programming Guide ONICON...
  • Page 2: Table Of Contents

    Partial Reverse Flow Totalizer Units [P- unit=Gal] ................15 Partial Reverse Flow Totalizer Decimal Point Position [P- D.P.=X] ..........16 Temperature Unit Type [Temp. unit= °F] ....................16 Mass Units Enable [Mass units=OFF] ......................16 Specific Gravity Coefficient [Sg=kg/dm X.XXXX] ................16 ONICON Incorporated 727.447.6140 Page 2 onicon.com...
  • Page 3 Enable Partial Totalizers [Part. tot.=OFF] ....................31 Enable Negative Totalizers [Neg. tot. = OFF] ..................31 Enable Net Totalizers [Net tot.=OFF] .....................32 Enable Quick Start Menu [Quick start=ON] ..................32 Default Display Page [Disp. fn = X] ......................33 ONICON Incorporated 727.447.6140 Page 3 onicon.com...
  • Page 4 Self Diagnostic Test [Self test] ........................35 Test Display [Display test] ...........................35 Simulate Flow Rate [Flow sim. = OFF] ....................35 Firmware Information [Firmware info] ....................35 Serial Number [S/N = XXXXXX] ........................35 Working Time [WT = XXXX:XX:XX:XX] ....................35 ONICON Incorporated 727.447.6140 Page 4 onicon.com...
  • Page 5: Introduction

    FT - 3100 TRANSMITTER PROGRAMMING GUIDE SECTION 1.0: INTRODUCTION The purpose of this guide is to provide programming information for the FT-3100 1.1 PURPOSE OF THIS GUIDE transmitter configured with the following options: • One (1) analog output • Two (2) digital outputs •...
  • Page 6: Accessing The Programming Menus

    ACCESS CODE ******* Code : 2. Enter the access code. 99999999 QUICK START E. P. Detect= R max=kohm 0220 3. Enter into the Quick Start menu. Pls2=Gal 1000.0 Tpls2=ms 0500.0 A1S=Gal/m 004500 Main Menu ONICON Incorporated 727.447.6140 Page 6 onicon.com...
  • Page 7: Menu - Quick Start

    (menu 2). In the same way, the list will show mass units, if mass units are enabled [Mass units = OFF] in menu 2. The tables below show the available units of measure. ONICON Incorporated 727.447.6140 Page 7 onicon.com...
  • Page 8: Output 2 Pulse Duration [Tpls2=Ms X.xxxx]

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 8 onicon.com...
  • Page 9: Main Menu

    Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 9 onicon.com...
  • Page 10: Liner Material [Lining=Xxxxxx]

    To enter a new coefficient, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 10 onicon.com...
  • Page 11: Empty Pipe Detector [E.p.detect=On]

    To change the existing setting, press enter to access the function and use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 11 onicon.com...
  • Page 12: Sensor Cable Length [S.cable=Ft Xxx]

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the interval. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 12 onicon.com...
  • Page 13: Menu 2 - Units

    To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 13 onicon.com...
  • Page 14: Pulse 1 Unit Type [Pl1 Unit=Imperial]

    To change the existing setting, press enter to access the function. Use the right arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 14 onicon.com...
  • Page 15: Partial Forward Flow Totalizer Units [P+ Unit=Gal]

    To change the existing setting, press enter to access the function. Use the up [P- unit = Gal] arrow key to change the selection. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 15 onicon.com...
  • Page 16: Partial Reverse Flow Totalizer Decimal Point Position [P- D.p.=X]

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the coefficient. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 16 onicon.com...
  • Page 17: Menu 3 - Scales

    Mega Gallon (US) Megaliter entry. Press enter to accept the Imperial Mega Gallon change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 17 onicon.com...
  • Page 18: Full Scale Flow Rate 2 [Fs2 = Gal/M Xxx.x]

    Mega Gallon (US) Press enter to accept the change and exit the function. Press escape to exit Imperial Mega Gallon the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 18 onicon.com...
  • Page 19: Output 1 Pulse Duration [Tpls1 = Ms X.xxxx]

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the entry. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 19 onicon.com...
  • Page 20: Output 2 Pulse Duration [Tpls2 = Ms X.xxxx]

    Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 20 onicon.com...
  • Page 21: Menu 4 - Measure

    Press escape to exit the program mode. Table of Contents Example wtih damping to 5,0sec. FILTERS >>>>> >>>>> Damping = from 0.2s to 1000s .FLOWRATE .MEASURE TIME (Sec) ONICON Incorporated 727.447.6140 Page 21 onicon.com...
  • Page 22: Low Flow Cut-Off [Cut-Off = % Xxx]

    To change the existing setting, press enter to access the function. Use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 22 onicon.com...
  • Page 23: Menu 5 - Alarms

    To enter a value, press enter to access the function. Use the up arrow and right arrow keys change the threshold value. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 23 onicon.com...
  • Page 24: Minimum Reverse Flow Alarm Threshold [Min - = Gal/M Off]

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 24 onicon.com...
  • Page 25: Frequency Output Value When In Alarm State [Hz V.alarm = %Xxx}

    To enter a new value, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 25 onicon.com...
  • Page 26: Menu 6 - Inputs

    To change the existing setting, press enter to access the function and use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 26 onicon.com...
  • Page 27: Activate Autozero Calibration Via Pulse Input [Calibration=Off]

    To change the existing setting, press enter to access the function and use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 27 onicon.com...
  • Page 28: Menu 7 - Outputs

    To change the setting, press enter to access the function. Use the up and right arrow keys to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 28 onicon.com...
  • Page 29: Ma Output 1 Function & Scaling [Out Ma1=4_20+/-]

    To change the setting, press enter to access the function. Use the right arrow key to move the cursor and the up arrow key to change the value. Press enter to accept changes and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 29 onicon.com...
  • Page 30: Menu 9 - Display

    To change the existing setting, press enter to access the function. Use the up arrow key to change the setting. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 30 onicon.com...
  • Page 31: Enable Partial Totalizers [Part. Tot.=Off]

    To change the existing setting, press enter to access the function and use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 31 onicon.com...
  • Page 32: Enable Net Totalizers [Net Tot.=Off]

    To change the existing setting, press enter to access the function and use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 32 onicon.com...
  • Page 33: Default Display Page [Disp. Fn = X]

    To change the existing setting, press enter to access the function and use the up arrow key to turn the function on or off. Press enter to accept the change and exit the function. Press escape to exit the program mode. Table of Contents ONICON Incorporated 727.447.6140 Page 33 onicon.com...
  • Page 34: Menu 11 - Functions

    This function loads the factory default operating parameters for the sensor body. Sensor Body Settings Note: Reloading these values will void the factory calibration. [Load Sens.f.def] Load Factory Default This function loads the factory default operating parameters for the sensor body. Transmitter Settings [Load Conv.f.def] ONICON Incorporated 727.447.6140 Page 34 onicon.com...
  • Page 35: Menu 12 - Diagnostic

    Displays a read only record of the current firmware and the date it was loaded. Firmware Information [Firmware info] Display a read only record of the electronics serial number Serial Number [S/N = XXXXXX] Working Time Display of the accumulated operating time for the meter. [WT = XXXX:XX:XX:XX] ONICON Incorporated 727.447.6140 Page 35 onicon.com...
  • Page 36 FT - 3100 TRANSMITTER PROGRAMMING GUIDE For questions regarding this manual, please contact the ONICON Technical Support Department at 727-447-6140. ONICON ONICON Incorporated 11451 Belcher Road South, Largo, FL 33773 USA April 2019 DOC-0000680 727-447-6140 • www.onicon.com • sales@onicon.com...