For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series
should assure that it is used in such an environment.
a)
Test frequency
Band
(MHz)
(MHz)
385
380–390
450
430–470
710
704–787
745
780
810
870
800–960
930
1720
1700–
1845
1990
1970
2400–
2450
2570
5240
5100–
5500
5800
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME
SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,