Download Print this page

Radiant Innovation TMS Series Manual

Smart ir thermometer

Advertisement

Quick Links

TMS Smart IR Thermometer
Intended Use:
The Smart Infrared Thermometer can measure the infrared heat generated by central
forehead and surrounding tissues to reflect patient's body temperature accurately.
Intended operator: At least 11 years old (5 years intensive reading experience), no maximum.
Specifications
 Temperature measurement range: Forehead Mode 34~42.2°C (93.2~108°F)
 Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
 Storage temperature range: It should be stored at room temperature
between -20~+50°C, RH≦85%.
Transportation temperature shall be less than 70°C, RH≦95%
 Atmospheric pressure: 800~1013 hPa
 Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
 Accuracy: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F),
 Fever indicator, history and °C/°F switch function
 Battery:
TMS-51A: one lithium cell battery (CR1616*1pcs)
TMS-01A : Battery Free
 Battery life: around 3,000 continuous times. (TMS-51A)
 Expected Service Life: 4 years
This thermometer converts the forehead temperature to display its "oral
equivalent."
 Enclosure Rating: IP22
 Dimensions:
TMS-51A : 52.76*25.03*10.51mm; TMS-01A: 53.86*9*9.97mm
 Weight: 7.8g (TMS-51A, including one battery); 3.4g (TMS-01A)
 OS:
TMS-51A : Support iOS 8.0 / Android 4.1 or above
TMS-01A :Support iOS 7.0 or above
 Electrical Interface: 3.5mm Phone Jack Plug
 With the Specific APP (APP Name: Radiant Forehead)
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
INSTRUCTION FOR USE:
Warning: Choking from swallowing small parts by children or pets is possible, please keep small parts at places where children and pets can't reach.
1. Download "Radiant Forehead" APP from the App Store (TMS-51A, TMS-01A) / Google Play (TMS-51A). If the dialog box appears on the screen, please press "OK"/
"ACCEPT".
2. Start the APP.
3. Plug the device into the phone jack.
4. Press "I" for basic operation
o
5. Switch temperature unit
C or °F
6. Simply aim the device within 1 cm from the central forehead and press the "MEASURE"
button to get the forehead measurement. The time consuming for measurement might be 1sec.
7. Fever Indicator
If thermometer detects a body temperature <37.5°C, there will be "Green Circle"
if temperature ≧37.5°C and <38.0°C, there will be "Yellow Circle"
RED: Continually monitor the body temperature. If in doubt about the person's condition, must seek medical assistance.
YELLOW: Continually monitor the body temperature. If in doubt about the person's condition, suggest seeking medical assistance.
GREEN: Normal body temperature.
8. History: Press
for memories of body temperature. Each memory also records the measurement fever Indicator icon /date/time.
Press
to measurement mode, Press
9. Cal:
This mode is just for the laboratory test with cavity blackbody.
CLEANING AND STORAGE
The Probe Lens is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.
a. Use the 70% alcohol swab or the soft cotton moistened with the 70% alcohol to clean the probe lens.
b. Allow the probe to fully dry for at least 1 minute.
c. Keep the unit dry and away from any liquids and direct sunlight.
d. Storage temperature range: It should be stored at room temperature between -20~+50°C, RH≦85%
e. The Probe should not be submerged into liquids.
Important Notes:
1. If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device
should stay in the same room for at least 15 minutes before measurement.
2. Before the measurement, the subject should stay in a stable environment for 5mins and avoid the exercise, bath for 30mins.
3. Remember to keep the forehead clean and away from sweat, cosmetics and scar while taking temperature.
4. During the measurement, please keep away from the direct sunshine and wind.
±0.3°C (0.5°F) for other range.
(according to the result of the clinical evaluation to get the offset value)
and then select "Clear" to clear all records.
signal;
signal; if temperature ≧38.0°C, there will be "Red Circle"
signal.
** The thermometer is NOT water-proof. Please
keep the unit dry and away from any liquids.

Advertisement

loading

Summary of Contents for Radiant Innovation TMS Series

  • Page 1 TMS Smart IR Thermometer Intended Use: The Smart Infrared Thermometer can measure the infrared heat generated by central forehead and surrounding tissues to reflect patient's body temperature accurately. Intended operator: At least 11 years old (5 years intensive reading experience), no maximum. Specifications ...
  • Page 2 Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of temperature measurements, or meet any unexpected event, please contact the dealers or nearest service address. Warning: No modification of this equipment is allowed. Please read the instructions for use Radiant Innovation Inc. Medical Technology Promedt Consulting GmbH Http://www.radiantek.com.tw Add: Altenhofstrasse 80, D-66386 St. Ingbert, Add: 1F, No.3, Industrial East 9...
  • Page 3 Manufacturer’s declaration-electromagnetic emissions The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series should assure that it is used in such an environment. Electromagnetic environment – guidance...
  • Page 4 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TMS series is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TMS series should assure that it is used in such an environment.

This manual is also suitable for: